  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Mansfield",
      "address_1": "15 Hampshire St",
      "reason_for_recall": "The firm identified the potential for a device to have an incorrect tissue gap. Use of a device with an incorrect tissue gap may result in incomplete staple formation and/or the inability to remove the device from tissue following application potentially leading to bleeding, anastomotic leak or tissue trauma. There have been no reports of serious injury related to this issue.",
      "address_2": "",
      "product_quantity": "1,902",
      "code_info": "P8A0976X  P8A0977X  P8A0978X  P8A0979X  P8A0980X  P8A0981X  P8B1538X  P8B1539X  P8B1540X  P8B1541X  P8B1542X  P8C0032X  P8C0033X  P8C1177X  P8C1178X  P8C1179X  P8C1180X  P8C1181X  P8C1183X  P8D1405X  P8D1406X  P8D1407X  P8D1408X  P8D1409X  P8D1410X  P8D1411X  P8E1211X  P8E1212X  P8E1213X  P8E1214X  P8E1215X  P8E1216X  P8E1217X  P8E1304X  P8E1305X",
      "center_classification_date": "20181004",
      "distribution_pattern": "Worldwide Distribution - US Nationwide in the states of CA, FL, IL, NC, NY, UT. and the countries of Austria, Belgium, Denmark, Finland, France, Germany, Iceland, Italy, Latvia, Netherlands, Norway, Poland, Portugal, Reunion, Spain, Sweden, Switzerland, and United Kingdom.",
      "state": "MA",
      "product_description": "EEA Circular Stapler with Tri-Staple Technology, Part No. TRIEEA28MT\tTRIEEA28XT\tTRIEEA31MT\tTRIEEA31XT    Product Usage:  The EEA Circular Stapler with Tri-Staple Technology places a circular, triple staggered row of titanium staples and resects the excess tissue, creating a circular anastomosis.",
      "report_date": "20181010",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "COVIDIEN LLC",
      "recall_number": "Z-0070-2019",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "81043",
      "termination_date": "20200410",
      "more_code_info": "",
      "recall_initiation_date": "20180817",
      "postal_code": "02048-1113",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}