      "status": "Terminated",
      "city": "Warsaw",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "78299",
      "recalling_firm": "Zimmer Biomet, Inc.",
      "address_1": "56 E Bell Dr",
      "address_2": "N/A",
      "postal_code": "46582-6989",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "This was a custom implant, there was only one (1) unit manufactured and distributed. The product has been implanted, so it is not subject to being returned.",
      "recall_number": "Z-0070-2018",
      "product_description": "Custom Ti MAK RS OSS Segmental Femoral Rotating Hinged Knee.     This Rotating Hinged Knee is intended to replace patient anatomy.",
      "product_quantity": "1",
      "reason_for_recall": "A custom implant manufactured entirely from titanium for a patient with a nickel allergy was requested by a surgeon. After the device was manufactured and used in the planned surgery, it was discovered internally that the implant assembly had a CoCrMo screw included as a component of the implant, which contains nickel.",
      "recall_initiation_date": "20170425",
      "center_classification_date": "20171108",
      "termination_date": "20180206",
      "report_date": "20171115",