  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Warsaw",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "65799",
      "recalling_firm": "Zimmer, Inc.",
      "address_1": "345 E Main St",
      "address_2": "N/A",
      "postal_code": "46580-2746",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution: US (nationwide) and countries of: Argentina, Australia, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Dominican Republic, Ecuador, El Salvador, Germany, Guatemala, Honduras, Hong Kong, India, Israel, Jamaica, Japan, Korea, Malaysia, Mexico, New Zealand, Nicaragua, Panama, Peru, Singapore, Taiwan, Thailand, and Venezula.",
      "recall_number": "Z-0070-2014",
      "product_description": "Versys¿ Beaded Fullcoat Femoral Stem\t(POR FULL-CT FEM ST 11X160MM; VERSYS BEADED FC 11X200MM 8ST; POR FULLCT FEM ST 11 LOW EXT; POR FULLCT FEM ST 11 EXT; POR FLCT FEM STEM LM 11X160MM; POR FULLCT LM FEM ST 11 EXT  POR FULL-CT FEM ST 12X160MM  VERSYS BEADED FC 12X200MM 8 ST   POR FULLCT FEM ST 12 LOW EXT  POR FLCT FEM ST 12.0X190 STR  POR FULLCT FEM ST 12 EXT  POR FULL-CT FEM ST LM 12X160MM  POR FULLCT LM FEM ST 12 EXT  POR FULL-CT FEM ST 13X160MM  VERSYS HIP SYSTEM REVISION FEMORAL BEADED FULLCOAT PLUS COLL   POR FULLCT FEM ST 13 LOW EXT  POR FLCT FEM ST 13.5X190 STR  POR FULLCT FEM ST 13 EXT  POR FULL-CT FEM ST LM 13X160MM  POR FULLCT LM FEM ST 13 EXT  POR FULL-CT FEM ST 14X160MM  VERSYS BEADED FC 14X200MM 8 ST   POR FULLCT FEM ST 14 LOW EXT  POR FULLCT FEM ST 14 EXT  POR FULL-CT FEM ST LM 14X160MM  POR FULLCT LM FEM ST 14 EXT  POR FULL-CT FEM ST 15X160MM  VERSYS BEADED FC 15X200MM 8 ST   POR FULLCT FEM ST 15 LOW EXT  POR FLCT FEM ST 15.0X190 STR  POR FLCT FEM ST 15.0X250 STR  POR FULLCT FEM ST 15 EXT  POR FULL-CT FEM ST LM 15X160MM  POR FULLCT FEM ST 15 XEXT  POR FULLCT LM FEM ST 15 EXT  POR FULL-CT FEM ST 16X160MM  VERSYS BEADED FC 16X200MM 8 ST   POR FULLCT FEM ST 16 LOW EXT  POR FLCT FEM ST 16.5X190 STR  POR FLCT FEM ST 16.5X250 STR  POR FULLCT FEM ST 16 EXT  POR FULL-CT FEM ST LM 16X160MM  POR FULLCT FEM ST 16 XEXT  POR FULLCT LM FEM ST 16 EXT  POR FULL-CT FEM ST 17X160MM  VERSYS BEADED FC 17X200MM 8 ST   POR FULLCT FEM ST 17 EXT  POR FULL-CT FEM ST LM 17X160MM  POR FULLCT FEM ST 17 XEXT  POR FULLCT LM FEM ST 17 EXT  POR FULL-CT FEM ST 18X160MM  VERSYS BEADED FC 18X200MM 8 ST   POR FLCT FEM ST 18.0X190 STR  POR FLCT FEM ST 18.0X250 STR  POR FULLCT FEM ST 18 EXT  POR FULL-CT FEM ST LM 18X160MM  POR FULLCT FEM ST 18 XEXT  POR FULLCT LM FEM ST 18 EXT  POR FULL-CT FEM ST 19X160MM  POR FLCT FEM ST 19.5X190 STR  POR FLCT FEM ST 19.5X250 STR  POR FULLCT FEM ST 19 EXT  POR FLCT FEM STEM LM 19X160MM  POR FULLCT FEM ST 19 XEXT  POR FULLCT LM FEM ST 19 EXT  POR FULL-CT FEM ST 20X160MM  VERSYS BEADED FC 20X200MM 8 ST   POR FULLCT FEM ST 20 EXT  POR FLCT FEM STEM LM 20X160MM  POR FULLCT FEM ST 20 XEXT  POR FULLCT LM FEM ST 20 EXT  POR FULL-CT FEM ST 21X160MM  POR FLCT FEM ST 21.0X190 STR  POR FLCT FEM ST 21.0X250 STR  POR FULLCT FEM ST 21 EXT  POR FLCT FEM STEM LM 21X160MM  POR FULLCT FEM ST 21 XEXT  POR FULLCT LM FEM ST 21 EXT  POR FULLCT FEM STEM 22X160MM  POR FLCT FEM ST 22.5X190 STR  POR FLCT FEM ST 22.5X250 STR  POR FULLCT FEM ST 22 EXT  POR FLCT FEM STEM LM 22X160MM  POR FULLCT FEM ST 22 XEXT  POR FULLCT LM FEM ST 22 EXT.    Used in total hip replacement.",
      "product_quantity": "6380",
      "reason_for_recall": "The low density polyethylene (LDPE) bag used to package implants adheres to the highly polished surface of the devices.",
      "recall_initiation_date": "20130830",
      "center_classification_date": "20131025",
      "termination_date": "20140925",
      "report_date": "20131106",
      "code_info": "Part Number Range: 00-7843-011/022-06/56"
    }
  ]
}