{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Aliso Viejo",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "86150",
      "recalling_firm": "Nuvasive Specialized Orthopedics Inc",
      "address_1": "101 Enterprise Ste 100",
      "address_2": "N/A",
      "postal_code": "92656-2604",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "Worldwide distribution.  US nationwide, Austria, Australia, Canada, Switzerland, Denmark, Spain, France, United Kingdom, Germany, UAE, Luxembourg, Israel, Netherlands.",
      "recall_number": "Z-0069-2021",
      "product_description": "NUVASIVE SPECIALIZED ORTHOPEDICS, INC. Precice Bone Transport Antegrade Femur Trochanter 10 Degree Bend    Component of the orthopedic Precice Bone Transport System of intramedullary nail, locking screws, reusable instruments, and hand-held External Remote Controller (ERC).",
      "product_quantity": "41 total units",
      "reason_for_recall": "During explantation, standard removal tools can generate sufficient forces to cause distal plug to disassociate from the bone transport system.",
      "recall_initiation_date": "20200724",
      "center_classification_date": "20201007",
      "report_date": "20201014",
      "code_info": "Item Number: PA1160-10D400-7  Description: ANTE FEMUR TROCH 10DEG11.5MM X 400MM 7/10CM ST  Lot/Serial Number:  9011815AAA  9021510AAA  9022225AAA  9072512AAA    Item Number: PA1161-10D400-7  Description: ANTE FEMUR TROCH 10DEG13MM X 400MM, 7/10CM ST  Lot/Serial Number:  9012506AAA  9061720AAA    Item Number: PA1160-10SJ360-7  Description: ANTE TIBIA 10DEG, 3-HOLE11.5MM X 360MM 7/10CM ST  Lot/Serial Number:  9013007AAA  9030706AAA  9031514AAA  9042208AAA  9050811AAA  9072218AAA    Item Number: PA1159-10SJ360-7  Description: ANTE TIBIA 10DEG, 3-HOLE10MM X 360MM, 7/10CM STK  Lot/Serial Number:  9020607AAA  9022815AAA  9050211AAA  9071604AAA    Item Number: PA1160-90SJ340-7  Description: ANTE TIBIA 10DEG, 3-HOLE11.5MM X 340MM, 7/9CM ST  Lot/Serial Number:  9020608AAA  9052110AAA  9052906AAA  9072217AAA    Item Number: PA1160-80SJ320-7  Description: ANTE TIBIA 10DEG, 3-HOLE11.5MM X 320MM, 7/8CM ST  Lot/Serial Number:  9021222AAA    Item Number: PA1159-80SJ320-7  Description: ANTE TIBIA 10DEG, 3-HOLE10MM X 320MM, 7/8CM STK  Lot/Serial Number:  9021223AAA    Item Number: PA1160-70SJ300-7  Description: ANTE TIBIA 10DEG, 3-HOLE11.5MM X 300MM, 7/7CM ST  Lot/Serial Number:  9021224AAA    Item Number: PA1160-10D380-7  Description: ANTE FEMUR TROCH 10DEG11.5MM X 380MM 7/10CM ST  Lot/Serial Number:  9021512AAA  9072513AAA  9072912AAA  9080106AAA    Item Number: PA1159-90SJ340-7  Description: ANTE TIBIA 10DEG, 3-HOLE10MM X 340MM, 7/9CM STK  Lot/Serial Number:  9020609AAA  9052208AAA  9071603AAA    Item Number: PA1161-10X380-7  Description: RETRO FEMUR STRAIGHT13MM X 380MM, 7/10CM ST  Lot/Serial Number:  9021913AAA    Item Number: PA1160-10X360-7  Description: RETRO FEMUR STRAIGHT11.5MM X 360MM, 7/10CM ST  Lot/Serial Number:  9022227AAA    Item Number: PA1161-10X360-7  Description: RETRO FEMUR STRAIGHT13MM X 360MM, 7/10CM ST  Lot/Serial Number:  9022226AAA    Item Number: PA1160-10SJ400-7  Description: ANTE TIBIA 10DEG, 3-HOLE11.5MM X 400MM 7/10CM ST  Lot/Serial Number:  9020809AAA    Item Number: PA1159-10SJ400-7  Description: ANTE TIBIA 10DEG, 3-HOLE10MM X 400MM, 7/10CM STK  Lot/Serial Number:  9020810AAA    Item Number: PA1160-10SJ380-7  Description: ANTE TIBIA 10DEG, 3-HOLE11.5MM X 380MM 7/10CM ST  Lot/Serial Number:  9022813AAA  9042506AAA    Item Number: PA1160-10B380-7  Description: ANTE FEMUR PIRIFOMIS11.5MM X 380MM 7/10CM ST  Lot/Serial Number:  9021513AAA    Item Number: PA1160-10B360-7  Description: ANTE FEMUR PIRIFOMIS11.5MM X 360MM 7/10CM ST  Lot/Serial Number:  9021514AAA    Item Number: PA1159-10SJ380-7  Description: ANTE TIBIA 10DEG, 3-HOLE10MM X 380MM, 7/10CM STK  Lot/Serial Number:  9031310AAA  9071605AAA    Item Number: PA1160-10D360-7  Description: ANTE FEMUR TROCH 10DEG11.5MM X 360MM 7/10CM ST  Lot/Serial Number:  9022816AAA    Item Number: PA1160-60SJ280-6  Description: ANTE TIBIA 10DEG, 3-HOLE11.5MM X 280MM, 6/6CM ST  Lot/Serial Number:  9021914AAA    Item Number: PA1159-70SJ300-7  Description: ANTE TIBIA 10DEG, 3-HOLE10MM X 300MM, 7/7CM STK  Lot/Serial Number:  9021915AAA    Item Number: PA1160-90X340-7  Description: RETRO FEMUR STRAIGHT11.5MM X 340MM, 7/9CM ST  Lot/Serial Number:  9041617AAA    Item Number: PA1160-10X380-7  Description: RETRO FEMUR STRAIGHT11.5MM X 380MM 7/10CM ST  Lot/Serial Number:  9041618AAA    Item Number: PA1159-10D380-7  Description: ANTE FEMUR TROCH 10DEG10MM X 380MM, 7/10CM STK  Lot/Serial Number:  9042211AAA    Item Number: PA1161-10B380-7  Description: ANTE FEMUR PIRIFOMIS13MM X 380MM, 7/10CM ST  Lot/Serial Number:  9052305AAA  9061009AAA    Item Number: PA1161-10B400-7  Description: ANTE FEMUR PIRIFOMIS13MM X 400MM, 7/10CM ST  Lot/Serial Number:  9052306AAA    Item Number: PA1160-10B400-7  Description: ANTE FEMUR PIRIFOMIS11.5MM X 400MM 7/10CM ST  Lot/Serial Number:  9053110AAA    Item Number: PA1160-10X400-7  Description: RETRO FEMUR STRAIGHT11.5MM X 400MM 7/10CM ST  Lot/Serial Number:  9061208AAA    Item Number: PA1161-10X400-7  Description: RETRO FEMUR STRAIGHT13MM X 400MM, 7/10CM ST  Lot/Serial Number:  9061209AAA    Item Number: PA1161-10B400-7  Description: ANTE FEMUR PIRIFOMIS13MM X 400MM, 7/10CM ST  Lot/Serial Number:  9061008AAA    Item Number: PA1159-80X320-7  Description: RETO FEMUR STRAIGHT10MM X 320MM, 7/8CM STK  Lot/Serial Number:  9062416AAA    Item Number: PA1161-10D380-7  Description: ANTE FEMUR TROCH 10DEG13MM X 380MM, 7/10CM ST  Lot/Serial Number:  9061721AAA  9071825AAA    Item Number: PA1159-10B400-7  Description: ANTE FEMUR PIRIFORMIS10MM X 400MM, 7/10CM STK  Lot/Serial Number:  9072219AAA    Item Number: PA1159-10D400-7  Description: ANTE FEMUR TROCH 10DEG10MM X 400MM, 7/10CM STK  Lot/Serial Number:  9072220AAA  9072911AAA    Item Number: PA1159-10B380-7  Description: ANTE FEMUR PIRIFORMIS10MM X 380MM, 7/10CM STK  Lot/Serial Number:  9072221AAA"
    }
  ]
}