  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Warsaw",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "65799",
      "recalling_firm": "Zimmer, Inc.",
      "address_1": "345 E Main St",
      "address_2": "N/A",
      "postal_code": "46580-2746",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution: US (nationwide) and countries of: Argentina, Australia, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Dominican Republic, Ecuador, El Salvador, Germany, Guatemala, Honduras, Hong Kong, India, Israel, Jamaica, Japan, Korea, Malaysia, Mexico, New Zealand, Nicaragua, Panama, Peru, Singapore, Taiwan, Thailand, and Venezula.",
      "recall_number": "Z-0069-2014",
      "product_description": "CPT¿ Femoral Stem\t(CPT 12/14 STEM SIZE 0 COCR; CPT 12/14 SIZE 0 COCR EXT; CPT 12/14 STEM SIZE 1 COCR; CPT 12/14 SIZE 1 COCR EXT  CPT 12/14 STEM SIZE 2 COCR; CPT 12/14 SIZE 2 COCR EXT; CPT 12/14 SIZE 2 COCR 180MM; CPT 12/14 SIZE2 COCR XEXT; CPT 12/14 STEM SIZE 3 COCR; CPT 12/14 SIZE 3 COCR EXT; CPT 12/14 SZ 3 COCR 180MM EXT; CPT 12/14 SIZE 3 COCR XEXT; CPT 12/14 STEM SIZE 4 COCR; CPT 12/14 SIZE 4 COCR EXT; CPT 12/14 SIZE 4 COCR EXT; CPT 12/14 SZ 4 COCR 230MM EXT; CPT 12/14 SZ 4 COCR 260MM EXT; CPT 12/14 SIZE 4 COCR XEXT; CPT 12/14 STEM SIZE 5 COCR; CPT 12/14 SIZE 5 COCR EXT; CPT 12/14 SIZE 5 COCR XEXT; CPT 12/14 STEM 2-180 VALGUSNCK; CPT 12/14 STEM 3-180 VALGUSNCK; CPT 12/14 STEM X-SML COCR; and CPT 12/14 STEM SML COCR).    Indicated for cemented use in patients suffering from severe hip pain.",
      "product_quantity": "24561",
      "reason_for_recall": "The low density polyethylene (LDPE) bag used to package implants adheres to the highly polished surface of the devices.",
      "recall_initiation_date": "20130830",
      "center_classification_date": "20131025",
      "termination_date": "20140925",
      "report_date": "20131106",
      "code_info": "Part Number Range: 00-8114-000/050-00/30"
    }
  ]
}