{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Miami Lakes",
      "address_1": "14400 Nw 60th Ave",
      "reason_for_recall": "Certain HVAD System Controller units may not meet the labeled standard for protection against water or fluid ingress, specifically they may have the potential to develop hairline cracks at the power ports of the controller housing.",
      "address_2": "",
      "product_quantity": "17,275 units",
      "code_info": "Serial number range:  CON300175 - CON320540",
      "center_classification_date": "20181003",
      "distribution_pattern": "Worldwide Distribution",
      "state": "FL",
      "product_description": "Heart Ware(TM) HVAD(TM) System Controller/Controller Kits, Model Numbers:  1403, 1407, and 1420    Product Usage:  The HeartWare  HVAD  System is indicated for hemodynamic support in patients with advanced, refractory left ventricular heart failure; either as a Bridge to Cardiac Transplantation (BTT), myocardial recovery, or as Destination Therapy (DT) in patients for whom subsequent transplantation is not planned.",
      "report_date": "20181010",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Heartware, Inc.",
      "recall_number": "Z-0067-2019",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "80966",
      "termination_date": "20190405",
      "more_code_info": "",
      "recall_initiation_date": "20180828",
      "postal_code": "33014-2807",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}