      "status": "Terminated",
      "city": "Hoffman Estates",
      "state": "IL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "66544",
      "recalling_firm": "Siemens Medical Solutions USA, Inc.",
      "address_1": "2501 Barrington Rd",
      "address_2": "N/A",
      "postal_code": "60195-2061",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Other",
      "distribution_pattern": "Worldwide Distribution - US: Nationwide, including Washington D.C., and Puerto Rico. and the countries of: Poland, France, Austria, Czech Republic, Ireland, Germany, Italy, Japan, Canada, United Arab Emirates, Turkey, Netherlands, United Kingdom, Brazil, South Africa, Kuwait, Belgium, Russia, Taiwan, Australia, Bangladesh, Bulgaria, Brazil, Switzerland, Chile, China, Colombia, Denmark, Algeria, Ecuador, Estonia, Egypt, Spain, Finland, Hong Kong, Croatia, Hungary, Indonesia, Iran, Israel, India, South Korea, Kuwait, Kazakhstan, Lithuania, Latvia, Morocco, Mexico, Norway, New Zealand, Peru, Phillipines, Pakistan, Portugal, Qatar, Romania, Saudi Arabia, Sweden, Singapore, Slovenia, Slovakia, Thailand, Tunisia, Taiwan, Venezuela, and Vietnam.",
      "recall_number": "Z-0062-2017",
      "product_description": "Symbia T Series SPECT + CT     Product Usage 1) SPECT: To Detect or image the distribution of radionuclides in the body or organ, using the following techniques: planar imaging, whole body imaging, tomographic imaging for isotopes with energies up to 588 keV. 2) CT: The CT component is intended to produce corss-sectional images of the body by computer reconstruction of x-ray transmission data from either the same axial plane taken at different angles or spiral planes taken at different angles. 3) SPECT + CT: Perform CT scans and nuclear imaging studies with the same instrument. To obtain attenuation corrected images and to provide registration of anatomical and physiological images within the patient\u0019s anatomy.",
      "product_quantity": "1109 units",
      "reason_for_recall": "The detectors on the Symbia S and Symbia T camera systems are supported and moved horizontally and vertically by four linear bearing cars. There is a potential for the linear bearing car to fail allowing the ball bearings to fall out.",
      "recall_initiation_date": "20130816",
      "center_classification_date": "20161007",
      "termination_date": "20161223",
      "report_date": "20161019",
      "code_info": "Material Number 10275007, Serial Numbers 1001 ,  1006 ,  1007 ,  1032 ,  1053 ,  1104 ,  1134 ,  1012 ,  1226 ,  1014 ,  1015 ,  1284 ,  1024 ,  1034 ,  1055 ,  1060 ,  1064 ,  1038 ,  1039 ,  1065 ,  1041 ,  1071 ,  1043 ,  1044 ,  1076 ,  1098 ,  1108 ,  1109 ,  1110 ,  1050 ,  1051 ,  1111 ,  1112 ,  1113 ,  1117 ,  1056 ,  1057 ,  1119 ,  1131 ,  1137 ,  1145 ,  1146 ,  1147 ,  1160 ,  1170 ,  1066 ,  1180 ,  1181 ,  1183 ,  1189 ,  1191 ,  1199 ,  1204 ,  1074 ,  1213 ,  1215 ,  1231 ,  1232 ,  1249 ,  1252 ,  1269 ,  1287 ,  1292 ,  1318 ,  1324 ,  1086 ,  1087 ,  1325 ,  1326 ,  1090 ,  1091 ,  1327 ,  1099 ,  1101 ,  1102 ,  1116 ,  1130 ,  1042 ,  1096 ,  1097 ,  1115 ,  1173 ,  1162 ,  1235 ,  1259 ,  1309 ,  1171 ,  1175 ,  1176 ,  1178 ,  1179 ,  1185 ,  1186 ,  1036 ,  1200 ,  1202 ,  1206 ,  1209 ,  1210 ,  1211 ,  1018 ,  1026 ,  1045 ,  1216 ,  1217 ,  1058 ,  1114 ,  1128 ,  1192 ,  1193 ,  1194 ,  1195 ,  1246 ,  1257 ,  1258 ,  1291 ,  1229 ,  1307 ,  1315 ,  1234 ,  1236 ,  1242 ,  1253 ,  1262 ,  1266 ,  1267 ,  1273 ,  1276 ,  1278 ,  1282 ,  1290 ,  1295 ,  1296 ,  1310 ,  1317 ,  1319 ,  1321 ,  1322 ,  1100 ,  1208 ,  1281 ,  1062 ,  1078 ,  1009 ,  1122 ,  1132 ,  1011 ,  1017 ,  1019 ,  1077 ,  1094 ,  1106 ,  1149 ,  1168 ,  1172 ,  1221 ,  1225 ,  1251 ,  1268 ,  1305 ,  1142 ,  1254 ,  1320 ,  1247 ,  1264 ,  1285 ,  1299 ,  1125 ,  1143 ,  1190 ,  1277 ,  1005 ,  1008 ,  1021 ,  1047 ,  1049 ,  1070 ,  1073 ,  1075 ,  1105 ,  1107 ,  1124 ,  1144 ,  1150 ,  1158 ,  1163 ,  1164 ,  1165 ,  1177 ,  1182 ,  1184 ,  1188 ,  1196 ,  1197 ,  1203 ,  1214 ,  1218 ,  1219 ,  1239 ,  1244 ,  1250 ,  1263 ,  1265 ,  1271 ,  1279 ,  1288 ,  1303 ,  1306 ,  1312 ,  1313 ,  1029 ,  1030 ,  1035 ,  1046 ,  1079 ,  1085 ,  1156 ,  1159 ,  1207 ,  1224 ,  1261 ,  1272 ,  1283 ,  1151 ,  1152 ,  1153 ,  1154 ,  1169 ,  1227 ,  1311 ,  1088 ,  1135 ,  1166 ,  1238 ,  1297 ,  1010 ,  1031 ,  1120 ,  1121 ,  1123 ,  1013 ,  1061 ,  1069 ,  1083 ,  1103 ,  1167 ,  1187 ,  1223 ,  1294 ,  1016 ,  1020 ,  1025 ,  1027 ,  1054 ,  1067 ,  1068 ,  1080 ,  1081 ,  1082 ,  1089 ,  1129 ,  1161 ,  1201 ,  1205 ,  1228 ,  1230 ,  1233 ,  1240 ,  1245 ,  1286 ,  1298 ,  1300 ,  1302 ,  1304 ,  1157 ,  1040 ,  1275 ,  1023 ,  1033 ,  1092 ,  1118 ,  1139 ,  1140 ,  1141 ,  1155 ,  1212 ,  1220 ,  1222 ,  1243 ,  1255 ,  1256 ,  1270 ,  1308 ,  1260 ,  1003 ,  1028 ,  1095 ,  1241 ,  1293 ,  1314 ,  1048 ,  1063 ,  1136 ,  1289 ,  1138 ,  1198 ,  1248 ,  1148 ,  1072 ,  1127 ,  1022 ,  1323 ,  1052 ,  1084 ,  1126 ,  1301 ,  1037 ,  1059 ,  1093 ,  1133 ,  1174 ,  1237 ,  1274 ,  1316 ,  46114 ,  6053 ,  68282 ,  74401 ,  74404 ,  8473047245.  Material Number 10275008, Serial Numbers 1071 ,  1077 ,  1051 ,  1063 ,  1124 ,  1054 ,  1172 ,  1016 ,  1032 ,  1033 ,  1060 ,  1093 ,  1171 ,  1045 ,  1049 ,  1056 ,  1082 ,  1092 ,  1095 ,  1107 ,  1111 ,  1121 ,  1123 ,  1131 ,  1138 ,  1144 ,  1145 ,  1176 ,  1182 ,  1098 ,  1002 ,  1068 ,  1166 ,  1009 ,  1035 ,  1011 ,  1037 ,  1044 ,  1059 ,  1074 ,  1083 ,  1102 ,  1103 ,  1118 ,  1141 ,  1142 ,  1149 ,  1023 ,  1156 ,  1161 ,  1170 ,  1028 ,  1029 ,  1173 ,  1178 ,  1181 ,  1186 ,  1038 ,  1040 ,  1058 ,  1090 ,  1091 ,  1094 ,  1053 ,  1097 ,  1062 ,  1099 ,  1100 ,  1104 ,  1108 ,  1113 ,  1114 ,  1117 ,  1139 ,  1150 ,  1177 ,  1165 ,  1167 ,  1168 ,  1169 ,  1179 ,  1180 ,  1184 ,  1185 ,  1566 ,  1039 ,  1042 ,  1055 ,  1065 ,  1066 ,  1096 ,  1109 ,  1127 ,  1128 ,  1153 ,  1175 ,  1003 ,  1057 ,  1080 ,  1001 ,  1004 ,  1006 ,  1007 ,  1012 ,  1013 ,  1021 ,  1024 ,  1026 ,  1036 ,  1048 ,  1052 ,  1061 ,  1070 ,  1084 ,  1087 ,  1101 ,  1105 ,  1106 ,  1135 ,  1158 ,  1010 ,  1030 ,  1031 ,  1078 ,  1143 ,  1018 ,  1020 ,  1183 ,  1120 ,  1140 ,  1017 ,  1025 ,  1064 ,  1089 ,  1188 ,  1022 ,  1047 ,  1067 ,  1075 ,  1086 ,  1116 ,  1125 ,  1126 ,  1132 ,  1136 ,  1137 ,  1151 ,  1041 ,  1085 ,  1147 ,  1008 ,  1069 ,  1115 ,  1133 ,  1160 ,  1162 ,  1164 ,  1110 ,  1073 ,  1159 ,  1088 ,  1154 ,  1076 ,  1014 ,  1015 ,  1134 ,  1072 ,  1119 ,  1122 ,  1130 ,  1146 ,  1155 ,  1174 ,  1050 ,  1079 ,  1148 ,  1081 ,  1005 ,  1157 ,  1046 ,  1019 ,  1034 ,  1129 ,  1152 ,  17283 ,  17440 ,  62322 ,  7283 ,  72836 ,  72837 ,  72862 ,  7440 ,  74409 ,  7444 ,  8473047245.  Material Number 10275009, Serial Numbers 1140 ,  1006 ,  1044 ,  1046 ,  1089 ,  1153 ,  1195 ,  1009 ,  1014 ,  1072 ,  1090 ,  1100 ,  1132 ,  1155 ,  1178 ,  1182 ,  1041 ,  1124 ,  1024 ,  1026 ,  1069 ,  1076 ,  1091 ,  1104 ,  1120 ,  1127 ,  1145 ,  1159 ,  1123 ,  1128 ,  1033 ,  1102 ,  1106 ,  1110 ,  1156 ,  1165 ,  1187 ,  1193 ,  1194 ,  1114 ,  1086 ,  1051 ,  1080 ,  1092 ,  1149 ,  1066 ,  1001 ,  1007 ,  1008 ,  1047 ,  1060 ,  1011 ,  1013 ,  1015 ,  1017 ,  1019 ,  1022 ,  1025 ,  1027 ,  1028 ,  1029 ,  1030 ,  1034 ,  1040 ,  1093 ,  1131 ,  1133 ,  1052 ,  1056 ,  1057 ,  1064 ,  1071 ,  1077 ,  1078 ,  1082 ,  1083 ,  1084 ,  1088 ,  1094 ,  1095 ,  1098 ,  1099 ,  1101 ,  1105 ,  1108 ,  1111 ,  1036 ,  1129 ,  1079 ,  1085 ,  1103 ,  1112 ,  1134 ,  1135 ,  1115 ,  1126 ,  1139 ,  1166 ,  1168 ,  1142 ,  1170 ,  1144 ,  1173 ,  1146 ,  1179 ,  1151 ,  1152 ,  1157 ,  1158 ,  1160 ,  1161 ,  1162 ,  1164 ,  1174 ,  1180 ,  1183 ,  1184 ,  1186 ,  1190 ,  1196 ,  1012 ,  1050 ,  1068 ,  1070 ,  1109 ,  1143 ,  1004 ,  1075 ,  1031 ,  1038 ,  1065 ,  1119 ,  1167 ,  1192 ,  1113 ,  1010 ,  1016 ,  1018 ,  1020 ,  1021 ,  1023 ,  1039 ,  1042 ,  1048 ,  1063 ,  1074 ,  1081 ,  1097 ,  1117 ,  1125 ,  1148 ,  1154 ,  1163 ,  1175 ,  1191 ,  1058 ,  1141 ,  1130 ,  1172 ,  1087 ,  1107 ,  1121 ,  1122 ,  1037 ,  1053 ,  1062 ,  1137 ,  1005 ,  1043 ,  1045 ,  1061 ,  1067 ,  1138 ,  1176 ,  1032 ,  1059 ,  1150 ,  1054 ,  1116 ,  1118 ,  1147 ,  1171 ,  1185 ,  1002 ,  1055 ,  1049 ,  1177 ,  1073 ,  1181 ,  1003 ,  1035 ,  1096 ,  1169 ,  1197 ,  62323 ,  7440 ,  7502.  Material Number 10275010, Serial Numbers 1161 ,  1163 ,  1059 ,  1072 ,  1006 ,  1053 ,  1061 ,  1076 ,  1078 ,  1096 ,  1108 ,  1109 ,  1113 ,  1116 ,  1117 ,  1118 ,  1119 ,  1141 ,  1142 ,  1153 ,  1160 ,  1166 ,  1173 ,  1030 ,  1049 ,  1145 ,  1150 ,  1057 ,  1058 ,  1156 ,  1012 ,  1052 ,  1149 ,  1151 ,  1005 ,  1014 ,  1079 ,  1122 ,  1168 ,  1037 ,  1045 ,  1056 ,  1066 ,  1088 ,  1121 ,  1136 ,  1155 ,  1167 ,  1125 ,  1019 ,  1022 ,  1034 ,  1044 ,  1048 ,  1050 ,  1093 ,  1101 ,  1107 ,  1112 ,  1135 ,  1064 ,  1111 ,  1120 ,  1172 ,  1103 ,  1140 ,  1152 ,  1171 ,  1001 ,  1003 ,  1004 ,  1008 ,  1009 ,  1011 ,  1013 ,  1015 ,  1016 ,  1017 ,  1018 ,  1025 ,  1027 ,  1028 ,  1031 ,  1032 ,  1038 ,  1039 ,  1042 ,  1043 ,  1051 ,  1054 ,  1055 ,  1070 ,  1071 ,  1075 ,  1002 ,  1021 ,  1026 ,  1047 ,  1085 ,  1083 ,  1084 ,  1095 ,  1086 ,  1154 ,  1089 ,  1090 ,  1094 ,  1097 ,  1098 ,  1100 ,  1102 ,  1104 ,  1105 ,  1110 ,  1115 ,  1130 ,  1137 ,  1138 ,  1139 ,  1143 ,  1164 ,  1158 ,  1170 ,  1007 ,  1024 ,  1040 ,  1065 ,  1074 ,  1082 ,  1087 ,  1091 ,  1092 ,  1133 ,  1144 ,  1147 ,  1148 ,  1162 ,  1165 ,  1174 ,  1146 ,  1035 ,  1077 ,  1134 ,  1029 ,  1036 ,  1041 ,  1063 ,  1069 ,  1080 ,  1081 ,  1114 ,  1159 ,  1010 ,  1067 ,  1124 ,  1033 ,  1129 ,  1062 ,  1073 ,  1106 ,  1126 ,  1127 ,  1128 ,  1131 ,  1132 ,  1046 ,  1020 ,  1023 ,  1157 ,  1068 ,  1060 ,  1099 ,  1123 ,  1644 ,  268045 ,  72838 ,  74401 .  Material Number 8717733, Serial Numbers 1040 ,  1088 ,  1103 ,  1157 ,  1158 ,  1159 ,  1163 ,  1041 ,  1042 ,  1043 ,  1071 ,  1072 ,  1073 ,  1074 ,  1083 ,  1090 ,  1107 ,  1147 ,  1171 ,  1173 ,  1178 ,  1196 ,  1197 ,  1009 ,  1038 ,  1066 ,  1068 ,  1070 ,  1076 ,  1079 ,  1087 ,  1128 ,  1143 ,  1144 ,  1150 ,  1153 ,  1177 ,  1188 ,  1189 ,  1193 ,  1199 ,  1108 ,  1194 ,  1000 ,  1005 ,  1057 ,  1061 ,  1094 ,  1116 ,  1141 ,  1151 ,  1183 ,  1205 ,  1025 ,  1101 ,  1024 ,  1029 ,  1044 ,  1048 ,  1053 ,  1062 ,  1081 ,  1086 ,  1106 ,  1117 ,  1135 ,  1181 ,  1209 ,  701001 ,  1058 ,  1082 ,  1105 ,  1162 ,  1064 ,  1010 ,  1015 ,  1032 ,  1050 ,  1119 ,  1013 ,  1014 ,  1016 ,  1017 ,  1027 ,  1035 ,  1036 ,  1039 ,  1051 ,  1060 ,  1078 ,  1084 ,  1085 ,  1099 ,  1109 ,  1110 ,  1111 ,  1112 ,  1122 ,  1129 ,  1132 ,  1136 ,  1140 ,  1145 ,  1155 ,  1156 ,  1164 ,  1165 ,  1166 ,  1172 ,  1186 ,  1192 ,  1195 ,  1201 ,  1120 ,  1137 ,  1146 ,  1148 ,  1149 ,  1176 ,  1185 ,  1045 ,  1095 ,  1097 ,  1004 ,  1007 ,  1008 ,  1011 ,  1019 ,  1020 ,  1021 ,  1022 ,  1023 ,  1026 ,  1028 ,  1056 ,  1059 ,  1098 ,  1034 ,  1121 ,  1200 ,  1037 ,  1052 ,  1054 ,  1055 ,  1065 ,  1067 ,  1075 ,  1077 ,  1080 ,  1091 ,  1093 ,  1113 ,  1115 ,  1118 ,  1125 ,  1126 ,  1133 ,  1139 ,  1142 ,  1089 ,  1124 ,  1130 ,  1134 ,  1160 ,  1152 ,  1154 ,  1161 ,  1167 ,  1168 ,  1169 ,  1170 ,  1175 ,  1180 ,  1182 ,  1187 ,  1191 ,  1198 ,  1202 ,  1203 ,  1204 ,  1206 ,  1208 ,  1092 ,  1012 ,  1049 ,  1096 ,  1063 ,  1033 ,  1100 ,  1174 ,  1030 ,  1127 ,  1138 ,  1184 ,  1003 ,  1104 ,  1179 ,  1190 ,  1046 ,  1001 ,  1002 ,  1006 ,  1018 ,  1031 ,  1102 ,  1114 ,  1131 ,  1210 ,  733612 ,  D613 ."
    }
  ]
}