{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Hoffman Estates",
      "state": "IL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "66544",
      "recalling_firm": "Siemens Medical Solutions USA, Inc.",
      "address_1": "2501 Barrington Rd",
      "address_2": "N/A",
      "postal_code": "60195-2061",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Other",
      "distribution_pattern": "Worldwide Distribution - US: Nationwide, including Washington D.C., and Puerto Rico. and the countries of: Poland, France, Austria, Czech Republic, Ireland, Germany, Italy, Japan, Canada, United Arab Emirates, Turkey, Netherlands, United Kingdom, Brazil, South Africa, Kuwait, Belgium, Russia, Taiwan, Australia, Bangladesh, Bulgaria, Brazil, Switzerland, Chile, China, Colombia, Denmark, Algeria, Ecuador, Estonia, Egypt, Spain, Finland, Hong Kong, Croatia, Hungary, Indonesia, Iran, Israel, India, South Korea, Kuwait, Kazakhstan, Lithuania, Latvia, Morocco, Mexico, Norway, New Zealand, Peru, Phillipines, Pakistan, Portugal, Qatar, Romania, Saudi Arabia, Sweden, Singapore, Slovenia, Slovakia, Thailand, Tunisia, Taiwan, Venezuela, and Vietnam.",
      "recall_number": "Z-0061-2017",
      "product_description": "Symbia S,    Product Usage 1) SPECT: To Detect or image the distribution of radionuclides in the body or organ, using the following techniques: planar imaging, whole body imaging, tomographic imaging for isotopes with energies up to 588 keV. 2) CT: The CT component is intended to produce corss-sectional images of the body by computer reconstruction of x-ray transmission data from either the same axial plane taken at different angles or spiral planes taken at different angles. 3) SPECT + CT: Perform CT scans and nuclear imaging studies with the same instrument. To obtain attenuation corrected images and to provide registration of anatomical and physiological images within the patient\u0019s anatomy.    .",
      "product_quantity": "732 units",
      "reason_for_recall": "The detectors on the Symbia S and Symbia T camera systems are supported and moved horizontally and vertically by four linear bearing cars. There is a potential for the linear bearing car to fail allowing the ball bearings to fall out.",
      "recall_initiation_date": "20130816",
      "center_classification_date": "20161007",
      "termination_date": "20161223",
      "report_date": "20161019",
      "code_info": "Material Number 8717741, Serial Numbers: 1001 ,  1002 ,  1003 ,  1005 ,  1006 ,  1007 ,  1008 ,  1009 ,  1010 ,  1011 ,  1012 ,  1013 ,  1014 ,  1015 ,  1016 ,  1017 ,  1018 ,  1019 ,  1020 ,  1021 ,  1022 ,  1023 ,  1024 ,  1025 ,  1027 ,  1028 ,  1029 ,  1030 ,  1031 ,  1032 ,  1033 ,  1034 ,  1035 ,  1036 ,  1037 ,  1038 ,  1039 ,  1040 ,  1041 ,  1042 ,  1043 ,  1044 ,  1045 ,  1046 ,  1047 ,  1048 ,  1049 ,  1050 ,  1051 ,  1052 ,  1053 ,  1054 ,  1055 ,  1056 ,  1057 ,  1058 ,  1059 ,  1060 ,  1061 ,  1062 ,  1063 ,  1064 ,  1065 ,  1066 ,  1067 ,  1068 ,  1069 ,  1070 ,  1071 ,  1072 ,  1073 ,  1074 ,  1075 ,  1076 ,  1077 ,  1078 ,  1079 ,  1080 ,  1081 ,  1082 ,  1083 ,  1084 ,  1085 ,  1086 ,  1087 ,  1088 ,  1089 ,  1090 ,  1091 ,  1092 ,  1093 ,  1094 ,  1095 ,  1096 ,  1097 ,  1098 ,  1099 ,  1100 ,  1101 ,  1102 ,  1103 ,  1104 ,  1105 ,  1106 ,  1107 ,  1108 ,  1109 ,  1110 ,  1111 ,  1112 ,  1113 ,  1114 ,  1115 ,  1116 ,  1117 ,  1118 ,  1119 ,  1120 ,  1121 ,  1122 ,  1123 ,  1124 ,  1125 ,  1126 ,  1127 ,  1128 ,  1129 ,  1130 ,  1131 ,  1132 ,  1133 ,  1134 ,  1135 ,  1136 ,  1137 ,  1138 ,  1139 ,  1140 ,  1141 ,  1142 ,  1143 ,  1144 ,  1145 ,  1146 ,  1147 ,  1148 ,  1149 ,  1150 ,  1151 ,  1152 ,  1153 ,  1154 ,  1155 ,  1156 ,  1157 ,  1158 ,  1159 ,  1160 ,  1161 ,  1162 ,  1163 ,  1164 ,  1165 ,  1166 ,  1167 ,  1168 ,  1169 ,  1170 ,  1171 ,  1172 ,  1173 ,  1174 ,  1175 ,  1176 ,  1177 ,  1178 ,  1179 ,  1180 ,  1181 ,  1182 ,  1183 ,  1184 ,  1185 ,  1186 ,  1187 ,  1188 ,  1189 ,  1190 ,  1191 ,  1192 ,  1193 ,  1194 ,  1195 ,  1196 ,  1197 ,  1198 ,  1199 ,  1200 ,  1201 ,  1202 ,  1203 ,  1204 ,  1205 ,  1206 ,  1207 ,  1208 ,  1209 ,  1210 ,  1211 ,  1212 ,  1213 ,  1214 ,  1215 ,  1216 ,  1217 ,  1218 ,  1219 ,  1220 ,  1221 ,  1222 ,  1223 ,  1224 ,  1225 ,  1226 ,  1227 ,  1228 ,  1229 ,  1230 ,  1231 ,  1232 ,  1233 ,  1234 ,  1235 ,  1236 ,  1237 ,  1238 ,  1239 ,  1240 ,  1241 ,  1242 ,  1243 ,  1244 ,  1245 ,  1246 ,  1247 ,  1248 ,  1249 ,  1250 ,  1251 ,  1252 ,  1253 ,  1254 ,  1255 ,  1256 ,  1257 ,  1258 ,  1259 ,  1260 ,  1261 ,  1262 ,  1263 ,  1264 ,  1265 ,  1266 ,  1267 ,  1268 ,  1269 ,  1270 ,  1271 ,  1272 ,  1273 ,  1274 ,  1275 ,  1276 ,  1277 ,  1278 ,  1279 ,  1280 ,  1281 ,  1282 ,  1283 ,  1284 ,  1285 ,  1286 ,  1287 ,  1288 ,  1289 ,  1290 ,  1291 ,  1292 ,  1293 ,  1294 ,  1295 ,  1296 ,  1297 ,  1298 ,  1299 ,  1300 ,  1301 ,  1302 ,  1303 ,  1304 ,  1305 ,  1306 ,  1307 ,  1308 ,  1309 ,  1310 ,  1311 ,  1312 ,  1313 ,  1314 ,  1315 ,  1316 ,  1317 ,  1318 ,  1319 ,  1320 ,  1321 ,  1322 ,  1323 ,  1324 ,  1325 ,  1326 ,  1327 ,  1328 ,  1329 ,  1330 ,  1331 ,  1332 ,  1333 ,  1334 ,  1335 ,  1336 ,  1337 ,  1338 ,  1339 ,  1340 ,  1341 ,  1342 ,  1343 ,  1344 ,  1345 ,  1346 ,  1347 ,  1348 ,  1349 ,  1350 ,  1351 ,  1352 ,  1353 ,  1354 ,  1355 ,  1356 ,  1357 ,  1358 ,  1359 ,  1360 ,  1363 ,  1364 ,  1365 ,  1366 ,  1367 ,  1368 ,  1369 ,  1370 ,  1371 ,  1372 ,  1373 ,  1374 ,  1375 ,  1376 ,  1377 ,  1378 ,  1379 ,  1380 ,  1381 ,  1382 ,  1383 ,  1384 ,  1385 ,  1386 ,  1387 ,  1388 ,  1389 ,  1390 ,  1391 ,  1392 ,  1393 ,  1394 ,  1395 ,  1396 ,  1397 ,  1398 ,  1399 ,  1400 ,  1401 ,  1402 ,  1403 ,  1404 ,  1405 ,  1406 ,  1407 ,  1408 ,  1409 ,  1410 ,  1411 ,  1412 ,  1413 ,  1414 ,  1415 ,  1416 ,  1417 ,  1418 ,  1419 ,  1420 ,  1421 ,  1422 ,  1423 ,  1424 ,  1425 ,  1426 ,  1427 ,  1428 ,  1429 ,  1430 ,  1431 ,  1432 ,  1433 ,  1434 ,  1435 ,  1436 ,  1437 ,  1438 ,  1439 ,  1440 ,  1441 ,  1442 ,  1443 ,  1444 ,  1445 ,  1446 ,  1447 ,  1448 ,  1449 ,  1450 ,  1451 ,  1452 ,  1453 ,  1454 ,  1455 ,  1456 ,  1457 ,  1458 ,  1459 ,  1460 ,  1461 ,  1462 ,  1463 ,  1464 ,  1465 ,  1466 ,  1467 ,  1468 ,  1469 ,  1470 ,  1471 ,  1472 ,  1473 ,  1474 ,  1475 ,  1476 ,  1477 ,  1478 ,  1479 ,  1480 ,  1481 ,  1482 ,  1483 ,  1484 ,  1485 ,  1486 ,  1487 ,  1488 ,  1489 ,  1490 ,  1491 ,  1492 ,  1493 ,  1494 ,  1495 ,  1496 ,  1497 ,  1498 ,  1499 ,  1500 ,  1501 ,  1502 ,  1503 ,  1504 ,  1505 ,  1506 ,  1507 ,  1508 ,  1509 ,  1510 ,  1511 ,  1512 ,  1513 ,  1514 ,  1515 ,  1516 ,  1517 ,  1518 ,  1519 ,  1520 ,  1521 ,  1522 ,  1523 ,  1524 ,  1525 ,  1526 ,  1527 ,  1528 ,  1529 ,  1530 ,  1531 ,  1532 ,  1533 ,  1534 ,  1535 ,  1536 ,  1537 ,  1538 ,  1539 ,  1540 ,  1541 ,  1542 ,  1543 ,  1544 ,  1545 ,  1546 ,  1547 ,  1548 ,  1549 ,  1550 ,  1551 ,  1552 ,  1553 ,  1554 ,  1555 ,  1556 ,  1557 ,  1558 ,  1559 ,  1560 ,  1561 ,  1562 ,  1563 ,  1564 ,  1565 ,  1566 ,  1567 ,  1568 ,  1569 ,  1570 ,  1571 ,  1572 ,  1573 ,  1574 ,  1575 ,  1576 ,  1577 ,  1578 ,  1579 ,  1580 ,  1581 ,  1582 ,  1583 ,  1584 ,  1585 ,  1586 ,  1588 ,  1589 ,  1590 ,  1591 ,  1592 ,  1593 ,  1594 ,  1595 ,  1596 ,  1597 ,  1598 ,  1599 ,  1600 ,  1601 ,  1602 ,  1603 ,  1604 ,  1605 ,  1606 ,  1607 ,  1608 ,  1609 ,  1610 ,  1611 ,  1612 ,  1613 ,  1614 ,  1615 ,  1616 ,  1617 ,  1618 ,  1619 ,  1620 ,  1621 ,  1622 ,  1623 ,  1624 ,  1625 ,  1626 ,  1627 ,  1628 ,  1629 ,  1630 ,  1631 ,  1632 ,  1633 ,  1634 ,  1635 ,  1636 ,  1637 ,  1638 ,  1639 ,  1640 ,  1641 ,  1642 ,  1643 ,  1644 ,  1645 ,  1646 ,  1647 ,  1648 ,  1649 ,  1650 ,  1651 ,  1652 ,  1653 ,  1654 ,  1655 ,  1656 ,  1657 ,  1658 ,  1659 ,  1660 ,  1661 ,  1662 ,  1663 ,  1664 ,  1665 ,  1666 ,  1667 ,  1668 ,  1669 ,  1670 ,  1671 ,  1672 ,  1673 ,  1674 ,  1675 ,  1676 ,  1677 ,  1678 ,  1679 ,  1680 ,  1681 ,  1682 ,  1683 ,  1684 ,  1685 ,  1686 ,  1687 ,  1688 ,  1689 ,  1690 ,  1691 ,  1692 ,  1693 ,  1694 ,  1695 ,  1696 ,  1697 ,  1698 ,  1699 ,  1700 ,  1701 ,  1702 ,  1703 ,  1704 ,  1705 ,  1706 ,  1707 ,  1708 ,  1709 ,  1710 ,  1711 ,  1712 ,  1713 ,  1714 ,  1715 ,  1716 ,  1717 ,  1718 ,  1719 ,  1720 ,  1721 ,  1722 ,  1723 ,  1724 ,  1725 ,  202703 ,  4439 ,  49161 ,  56173 ,  6162 ,  701158475 ,  7162 ,  71622 ,  73361 ,  7441 ,  7502 ,  D614 ."
    }
  ]
}