{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Austin",
      "state": "TX",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "86257",
      "recalling_firm": "Encore Medical, LP",
      "address_1": "9800 Metric Blvd",
      "address_2": "N/A",
      "postal_code": "78758-5445",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "U.S. Nationwide distribution including in the states of VA, TN, MN, TX, IL, IN, KY, OH, SC, RI, MD, PA, NY, CA, KS, OK, AR, LA, FL, AL, GA, AZ, NV, WA, MS, ME, NV and Puerto Rico.",
      "recall_number": "Z-0060-2021",
      "product_description": "DJO EMPOWR Tibial Knee Punch Handle Part No. 801-05-040 and 801-05-018 contained in Instrument Tray for EMPOWR Porous Knee  - Product Usage: The Tibial Punch Handle is used in conjunction with one of three appropriately sized Tibial Punch Guides to broach the tibial canal in preparation for the tibial baseplate keel.",
      "product_quantity": "190 units",
      "reason_for_recall": "There is a potential for the retaining pin of the tibial punch handle to become loose and fall out while impacting.  If  undetected, there is a potential of leaving debris in the joint space.",
      "recall_initiation_date": "20200803",
      "center_classification_date": "20201007",
      "report_date": "20201014",
      "code_info": "Model/Catalog no. 801-05-040; :Lot/serial No.s 281905L08, 281905L15, 283808L01, 305197L02, 318871L10  Model/Catalog no. 801-05-018, Lot/serial No.s: 215488L08, 236760L01, 228231L01, 241946L02, 259710L01, 314969L01"
    }
  ]
}