  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Motala",
      "state": "N/A",
      "country": "Sweden",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "86417",
      "recalling_firm": "AB ULAX",
      "address_1": "Merkuriusgatan 8",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide distribution and the countries of Sweden, Taiwan, Hong Kong, and Spain.",
      "recall_number": "Z-0059-2021",
      "product_description": "1500 Flex Heat and Moisture Exchanger, Catalog Reference 150 - Product Usage:  Heat Moisture Exchangers (HM Es) are designed to enhance humification of ventilator-dependent and non-ventilator dependent patients in critical care, home care and anesthesia.",
      "product_quantity": "1,484,100 units",
      "reason_for_recall": "Firm has received 6 complaints that the Heat Moisture Exchanger (HME) disconnected from the closed respiratory circuit during use. If the HME separates, the ventilator circuit may not provide adequate ventilation to the patient, potentially resulting in lack of oxygen leading to hypoxia or death.",
      "recall_initiation_date": "20200904",
      "center_classification_date": "20201006",
      "termination_date": "20230210",
      "report_date": "20201014",
      "code_info": "SAP 109381602, Lots: 2018-09-25  2018-11-20  2018-12-10  2018-12-20  2019-01-03  2019-02-20  2019-03-20  2019-04-15  2019-05-30  2019-06-25  2019-08- 15  2019-09-05  2019- 11-05  2019-11-25    ***    SAP 109381603, Lots 2019-12-20  2020-04-30  2020-05-15  2020-05-20  2020-06-15  2020-06-30"
    }
  ]
}