  "results": [
    {
      "status": "Terminated",
      "city": "Brea",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "63191",
      "recalling_firm": "Beckman Coulter Inc.",
      "address_1": "250 S Kraemer Blvd",
      "address_2": "N/A",
      "postal_code": "92821-6232",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution and US (nationwide) and the countries of China, France, Italy and Taiwan.",
      "recall_number": "Z-0059-2013",
      "product_description": "Unicel Closed Tube Aliquotter (UCTA) for UniCel DxC 660i, UniCel DxC 680i, UniCel DxC 860i, UniCel DxC 880i Synchron Access Clinical Systems.  Part Number A45853.    Product Usage:  The UniCel DxC Synchron Clinical Systems are fully automated, computer-controlled clinical chemistry analyzers designed for the in-vitro determination of a variety of general chemistries, therapeutic drugs, and other chemistries. Analysis can be performed on serum, plasma, urine, or cerebrospinal fluid (CSF) and whole blood (sample type is chemistry dependent).",
      "product_quantity": "18 units",
      "reason_for_recall": "The recall was initiated because Beckman Coulter has confirmed that the current sample syringe pump drive assemblies installed on  DxC Integrated ystem with UniCel Closted Tube Aliquotter (UCTA) serial numbers between 547 and 574 may cause a hardware motion error (ID0x08000501).",
      "recall_initiation_date": "20120706",
      "center_classification_date": "20121012",
      "termination_date": "20121115",
      "report_date": "20121024",
      "code_info": "Serial Numbers:  547-574 (UCTA)"
    }
  ]
}