{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Maple Grove",
      "state": "MN",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "75243",
      "recalling_firm": "Vascular Solutions, Inc.",
      "address_1": "6464 Sycamore Ct N",
      "address_2": "N/A",
      "postal_code": "55369-6032",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution: US (nationwide) including Puerto Rico and outside the US to include:  Australia, Austria, Belgium, Brazil, Canada, Cyprus, Czech Republic,  Egypt, Finland,  France, Germany, Greece, Hong Kong, Hungary, Ireland, Israel, Italy, Netherlands, Norway, Poland, Saudi Arabia, South Africa, Spain, Sweden, Switzerland, Taiwan, Turkey, United Kingdom.",
      "recall_number": "Z-0055-2017",
      "product_description": "Vascular Solutions Twin-Pass .023\" Dual Access Catheter, model 5230.  Sterilized with Ethylene Oxide.    The Twin-Pass dual access catheter is a dual lumen catheter designed for use in the coronary and/or peripheral vasculature.  The Twin-Pass catheter (Models 5200 and 5230) consists of an over-the-wire (OTW) lumen that runs the length of the catheter and a rapid exchange (RX) delivery lumen on the distal segment.  The Twin-Pass catheter has a working length of 135cm and is compatible with various guidewire diameters.    Model 5200 also has hydrophilic coating on the distal 18cm of the catheter.    The Twin-Pass catheter has white positioning marks located at 95cm (single mark) and 105cm (double marks) from the distal tip, respectively.  The Twin-Pass catheter has a marker band located 1mm from the distal tip.  Model 5200 contains a second marker band located 11mm from the distal tip to identify the distal end of the OTW lumen.",
      "product_quantity": "11,670 units total",
      "reason_for_recall": "Vascular Solutions is recalling Twin-Pass dual access catheters  because there is a potential for excess manufacturing material to remain at the tip of the catheter or within the distal portion of the rapid exchange lumen.",
      "recall_initiation_date": "20160916",
      "center_classification_date": "20161020",
      "termination_date": "20180504",
      "report_date": "20161026",
      "code_info": "Model 5230, Lot numbers:    577279, 578997, 580613, 583021, 584463, 586310, 588542, 590739, 591262, 593696, 593985."
    }
  ]
}