      "status": "Terminated",
      "city": "Maple Grove",
      "state": "MN",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "75243",
      "recalling_firm": "Vascular Solutions, Inc.",
      "address_1": "6464 Sycamore Ct N",
      "address_2": "N/A",
      "postal_code": "55369-6032",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution: US (nationwide) including Puerto Rico and outside the US to include:  Australia, Austria, Belgium, Brazil, Canada, Cyprus, Czech Republic,  Egypt, Finland,  France, Germany, Greece, Hong Kong, Hungary, Ireland, Israel, Italy, Netherlands, Norway, Poland, Saudi Arabia, South Africa, Spain, Sweden, Switzerland, Taiwan, Turkey, United Kingdom.",
      "recall_number": "Z-0054-2017",
      "product_description": "Vascular Solutions Twin-Pass RX Dual Access Catheter, model 5210.  Sterilized with Ethylene Oxide.    The Twin-Pass dual access catheter is a dual lumen catheter designed for use in the coronary and/or peripheral vasculature.  The Twin-Pass catheter (Models 5200 and 5230) consists of an over-the-wire (OTW) lumen that runs the length of the catheter and a rapid exchange (RX) delivery lumen on the distal segment.  The Twin-Pass catheter has a working length of 135cm and is compatible with various guidewire diameters.    Model 5200 also has hydrophilic coating on the distal 18cm of the catheter.    The Twin-Pass catheter has white positioning marks located at 95cm (single mark) and 105cm (double marks) from the distal tip, respectively.  The Twin-Pass catheter has a marker band located 1mm from the distal tip.  Model 5200 contains a second marker band located 11mm from the distal tip to identify the distal end of the OTW lumen.",
      "product_quantity": "11,670 units total",
      "reason_for_recall": "Vascular Solutions is recalling Twin-Pass dual access catheters  because there is a potential for excess manufacturing material to remain at the tip of the catheter or within the distal portion of the rapid exchange lumen.",
      "recall_initiation_date": "20160916",
      "center_classification_date": "20161020",
      "termination_date": "20180504",
      "report_date": "20161026",
      "code_info": "Model 5210, Lot number:  577762"
    }
  ]
}