  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Ann Arbor",
      "state": "MI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "63079",
      "recalling_firm": "Terumo Cardiovascular Systems Corporation",
      "address_1": "6200 Jackson Road",
      "address_2": "N/A",
      "postal_code": "48103-9586",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution: USA including: AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MI, MN, MO, MS, NC, NE, NH, NJ, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, UT, VA, WA, WI, and WV. Puerto Rico and internationally to:    ALEXANDRIA, ARGENTINA, AUSTRALIA, AUSTRIA, BANGLADESH, BELGIUM, BRAZIL, CANADA, CENTRO AMERICA, CHILE, COLOMBIA, Dominican Republic, Ecuador, FRANCE, GERMANY, GREECE, Guatamala, HOLLAND, HONG KONG, India, ITALY, Japan, KL SEGUNDO, KOREA, LEICESTERSHIRE, MALAYSA, Malaysia, Mexico, Mexico, Mexico, NETHERLANDS, Panama, Peru, Philippines, RALEIGH, Republic of Georgia, SANTIAGO CHILE, Saudi Arabia, SEOUL KOREA, SHANGHAI, CHINA, Singapore, South Africa, SPAIN, SWITZERLAND, TAIWAN, Thailand, UNITED ARAB EMIRATES (UAE), Uruguay, USA, and Venezuela,",
      "recall_number": "Z-0049-2013",
      "product_description": "Battery, 100/115V (Includes charger and connecting cable)    The Sarns\" Centrifugal System is indicated for use in cardiopulmonary bypass procedures only.",
      "product_quantity": "1076",
      "reason_for_recall": "Terumo Cardiovascular Systems (Terumo CVS) has received multiple reports since 2004 of various  malfunctions of the battery backup system for the Sarns\" Centrifugal System.  The reports can be broadly categorized as follows:  \" Failure to charge  \" Failure to hold charge  \" Failure to operate    In the most severe case, a battery malfunction could result in an unexpected loss of arterial flow. In",
      "recall_initiation_date": "20120925",
      "center_classification_date": "20121011",
      "termination_date": "20140115",
      "report_date": "20121017",
      "code_info": "Catalog number: 9490; and lot numbers: 0101,1001-1025,1188-1192,1997, 2000-2017, 2019- 2082, 2086-2287,  2290-2446, 2452-2633, 2636-2938, and 2946-3063"
    }
  ]
}