      "status": "Terminated",
      "city": "Malvern",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "63045",
      "recalling_firm": "Siemens Medical Solutions USA, Inc",
      "address_1": "51 Valley Stream Pkwy",
      "address_2": "N/A",
      "postal_code": "19355",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide Distribution including the states of MO, FL, MA, NY, WI, GA, CA and WA.",
      "recall_number": "Z-0048-2013",
      "product_description": "Siemens syngo.plaza radiological image processing system.",
      "product_quantity": "11",
      "reason_for_recall": "Siemens became aware of an unintended behavior when using syngo.plaza.  In syngo.plaza within the online \"Short Time Storage\" (STS), some studies are marked with an archived icon, however, the entry in the Archive Database referring to the physical file may be missing.",
      "recall_initiation_date": "20120606",
      "center_classification_date": "20121011",
      "termination_date": "20130413",
      "report_date": "20121017",
      "code_info": "Model number 10592457, serial numbers 100263, 100401, 100177, 100138, 100174, 100406, 100249, 100201, 100198, 100181, 100146."
    }
  ]
}