      "status": "Terminated",
      "city": "Brea",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "77982",
      "recalling_firm": "Beckman Coulter Inc.",
      "address_1": "250 S Kraemer Blvd",
      "address_2": "N/A",
      "postal_code": "92821-6232",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Other",
      "distribution_pattern": "China  France  Italy  Kuwait  Spain  Turkey  United States",
      "recall_number": "Z-0046-2018",
      "product_description": "Power Processor 1K Stockyard.    The Power Processor performs all pre-analytical sample tube  preparation.",
      "product_quantity": "27 units",
      "reason_for_recall": "Beckman Coulter has identified that due to a PLC software nonconformity the 1K Stockyard  can initiate a retrieval of a sample tube during the rack loading process, which should not  occur. This issue and associated complaint were discovered and filed internally.",
      "recall_initiation_date": "20170803",
      "center_classification_date": "20171024",
      "termination_date": "20230927",
      "report_date": "20171101",
      "code_info": "Software version PLC V26R10"
    }
  ]
}