  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Tempe",
      "state": "AZ",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "78224",
      "recalling_firm": "Bard Peripheral Vascular Inc",
      "address_1": "1625 W 3rd St Bldg 1",
      "address_2": "N/A",
      "postal_code": "85281-2438",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US, Australia, Canada, Europe, Middle East, Africa, Latin America, Asia Pacific",
      "recall_number": "Z-0044-2018",
      "product_description": "BARD MAX-CORE Disposable Core Biopsy Instrument, Product Codes - MC1410, 14 g x 10 cm; MC1616, 16 g x 16 cm; MC1816, 18 g x 16 cm; MC1820, 18 g x 20 cm; MC1825, 18 g x 25 cm.    The core needle biopsy device is intended for use in obtaining biopsies from soft tissues.",
      "product_quantity": "110,550 units",
      "reason_for_recall": "During the review of May 2017 complaint data, a significant increase in MAX-CORE Disposable Core Biopsy Instrument complaints was observed for Failure to Prime, Failure to Fire, Failure to Obtain Sample, and Self-Activation.",
      "recall_initiation_date": "20170922",
      "center_classification_date": "20171020",
      "termination_date": "20180927",
      "report_date": "20171101",
      "code_info": "Lot Numbers: REBP0019, REBN0342, REBN2123, REBP1199, REBP1419, REBP1807, REBQ0084, REBQ0343, REBQ1012, REBP1420, REBP1809, REBP0869, REBP1266, REBP1267, REBP1421, REBP1422, REBP1810, REBQ0087, REBQ0088, REBQ0347, REBQ0811, REBQ1014, REBP0158, REBQ1904, REBQ1898, REBQ1978, REBR0468, REBQ2296, REBR0474"
    }
  ]
}