  "results": [
    {
      "status": "Terminated",
      "city": "Portland",
      "state": "OR",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "86254",
      "recalling_firm": "Riverpoint Medical, LLC",
      "address_1": "825 NE 25th Ave",
      "address_2": "N/A",
      "postal_code": "97232-2304",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "US Nationwide distribution.",
      "recall_number": "Z-0040-2021",
      "product_description": "Velosorb Fast Braided Absorbable Suture:    Covidien Velosorb 4/0 Undyed 30\" P-12 Cutting\tSV426,\tUDI: 20884521151731   - Product Usage: indicated for use in soft tissue approximation of the skin and mucosa, where only 7-10 days of wound support is required, but not for use in ligation, ophthalmic, cardiovascular or neurological procedures.",
      "product_quantity": "1242",
      "reason_for_recall": "The surgical suture is non-sterile. Use could compromise the sterile field, increase risk of a surgical site infection, and increase break strength retention.",
      "recall_initiation_date": "20200803",
      "center_classification_date": "20201001",
      "termination_date": "20220908",
      "report_date": "20201007",
      "code_info": "Lot Number:    20060813FX"
    }
  ]
}