  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Dieren",
      "state": "N/A",
      "country": "Netherlands",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "83648",
      "recalling_firm": "Vital Scientific N.V.",
      "address_1": "24, Kanaalweg",
      "address_2": "Postbus 100",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "US Nationwide distribution including the state of New York.",
      "recall_number": "Z-0038-2020",
      "product_description": "V-Twin, Catalog # 6002-800, UDI: 03661540600180 - Product Usage: An in vitro diagnostic device, which is an automated chemistry analyzer, micro type, intended for clinical use in conjunction with certain materials to measure a variety of analytes, including applications in clinical chemistry, monitoring drugs of abuse and in therapeutic drug monitoring.",
      "product_quantity": "192 devices",
      "reason_for_recall": "The action is being initiated because the cooling indicator board was improperly  installed. This was first noticed by a service employee during assembly. If for some reason the cooling of the reagent rotor fails, the reagent cooling indicator light will not illuminate and the acoustic beeper will not go off. As a consequence the operator will not be warned and the reagent storage conditions may fall out of specification.",
      "recall_initiation_date": "20180907",
      "center_classification_date": "20191004",
      "termination_date": "20220629",
      "report_date": "20191016",
      "code_info": "The following serial numbers: 14-3301-14-3344;  15-3301-15-3326;  16-3301 -16-3353;  17-3301 -17-3346;  18-3301-18-3323"
    }
  ]
}