{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Sylmar",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "78093",
      "recalling_firm": "St Jude Medical Inc.",
      "address_1": "15900 Valley View Ct",
      "address_2": "",
      "postal_code": "91342-3577",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution",
      "recall_number": "Z-0038-2018",
      "product_description": "PM3544 QUADRA RELIEVE and PM3562 QUADRA ALLURE MP  These low voltage (LV) devices are implantable pacemaker pulse generators, intended to be permanently implanted in the body, that have a power supply and electronic circuits that produce a periodic electrical pulse to stimulate the heart. These devices are used as a substitute for the heart's intrinsic pacing system to correct both intermittent and continuous cardiac rhythm disorders",
      "product_quantity": "2726 units",
      "reason_for_recall": "New pacemaker firmware was developed to further mitigate the risk of unauthorized access to our pacemakers that utilize radio frequency (RF) communications.",
      "recall_initiation_date": "20170828",
      "center_classification_date": "20171019",
      "termination_date": "20200708",
      "report_date": "20171025",
      "code_info": "Serial Numbers:  4794381  4913465  4913480  4907866  4909681  4909682  4913495  4909675  4909680  4909677  4909674  4913482  4913489  4913487  4913472  4913458  4913133  4913137  4913477  4909678  4913469  4913459  4913498  4909347  4913492  4913139  4908522  4909679  4909350  4909345  4909348  4909190  4909189  4909188  4909186  4909349  4908662  4909182  4909184  4909183  4909346  4909344  4909185  4909352  4909351  4909187  4913155  4913493  4913844  4913143  4913148  4913166  4913153  4913167  4913163  4913158  4913140  4913466  4913164  4913150  4913125  4913131  4913159  4913162  4913160  4913461  4913146  4913500  4915969  4915951  4915954  4915947  4915958  4915968  4915961  4915949  4913468  4915957  4913842  4913471  4913475  4913481  4913499  4913147  4913496  4913479  4913843  4913132  4915955  4913476  4913152  4913501  4913470  4913462  4913463  4913138  4913134  4913460  4913124  4913473  4913145  4913478  4913464  4913467  4913491  4913483  4913484  4913485  4913486  4913474  4913142  4913127  4913130  4913136  4913129  4913156  4913161  4913488  4913157  4913497  4913141  4913490  4913151  4913154  4913149  4913144  4913168  4913457  4913165  4913128  4913126  4913135  4883225  4880664  4837327  4838100  4837325  4838488  4793649  4796562  4796798  4796555  4796792  4796575  4796564  4796567  4796566  4796795  4794382  4838101  4838105  4837315  4886740  4837316  4838496  4793656  4841520  4794374  4794390  4794370  4838112  4794378  4802372  4793654  4795227  4796785  4869214  4799303  4867932  4867933  4867927  4838091  4886419  4801890  4886421  4802322  4886420  4886729  7896805  4838097  4886751  4801893  4837306  4839292  4838393  4839310  4886746  4801904  4838494  4798072  4838498  4886743  4838116  4838445  4886416  4886747  4801897  4839303  4838076  4885056  4798962  4797671  4799291  4799609  4799296  4838497  4867937  4880259  4880270  4880674  4798963  4884428  4796790  4885053  4884427  7873203  4885058  4885754  4884439  4885048  4836568  4838072  4880268  7896803  7896795  4799302  4838378  4837354  4837314  4841537  4842197  4838083  4869225  4838383  7896802  4836205  4836564  4838102  7873202  4869217  4836204  4887988  7873209  4883223  4883242  4883252  4883251  4883246  4836546  7873201  7842069  4798952  7824442  4831719  4867298  4810905  4887870  4809771  7841319  4880285  4839297  7824469  4804313  4836561  4795221  4796807  4888066  4810913  4832721  4832727  7841517  4880648  4832888  4838092  4794375  4838117  4802319  4795672  4788631  4836572  4800596  4806724  4796799  4841527  4836560  4832718  4795214  4809767  4796576  4809774  4841522  4802312  4810892  4838084  4866969  4838103  4838374  4788532  4872150  4831716  4882459  4832738  4806093  4789586  4837351  4886739  4836566  4837320  4804408  4837318  4880612  4831689  4789284  4886742  4804915  4838492  4842512  4767628  4804308  4805158  4801888  4809753  4802309  4768183  4793648  4838085  4797668  4788533  4837322  4832707  4880564  7841225  4842507  7840617  4788617  4804917  4838450  4795212  4799300  4871271  4805163  4866628  7841547  4789279  4797661  4795226  4789267  4872156  4872151  4832720  4839283  4804404  4838446  4837341  4795206  4838075  4888036  4885045  4885757  4788526  4837353  4788628  7841235  4789590  4838444  4767717  4768192  4800823  4802379  4801914  4789275  4802378  4832728  4804910  4789594  4798070  7840582  4842508  4799614  7841512  7842364  4795671  4837331  4788614  4804405  4797660  4804411  4802310  4806723  4810916  4810910  4837345  4798054  7904615  4796572  4804914  4885047  4883237  4794384  4793653  4810940  4838485  4872159  4788625  4804403  4832889  4872143  7841925  4870923  4832704  7840615  4809775  4867150  4810936  4767719  7841322  4799299  4805162  4839295  7841527  4795205  4871255  4767571  4795218  4788525  4809754  4806310  7902604  4804400  4798057  4801891  4810931  4866960  7904609  4806725  4796800  7904820  4800595  4767578  4789599  4809755  4768182  4768181  4796808  4810907  4804412  4806884  7842359  4789606  4801905  4788634  4806891  4797666  4831705  4789273  4789271  4798059  4796568  4794359  4799617  4838118  4871208  4788630  4767570  7841388  4871217  7842424  7825756  4789290  7843413  4788517  7904606  4839294  4789582  4809749  4810938  4810906  7842403  4768186  7841230  7842375  7842394  7842066  4869211  4869212  4869216  7842068  4867939  7841904  7842091  4868252  7842391  7841915  7841391  7841231  7842361  7841234  7841381  7840667  7840605  7840610  7840663  7824463  7841911  4869219  7824450  7824479  7824449  7824459  7824446  7824426  7825758  7824438  7825757  7825766  7841529  4869220  7842379  4883239  7842084  7841909  7842374  7842376  7842085  7840580  7842075  7841908  7841926  4868579  7843400  4880657  7824440  7824476  4868582  7825772  7824439  7824467  7824454  7843425  7843411  7843407  7842429  4880294  7825774  4810911  4869218  4886076  4885758  4885747  48721266  4883106  4799289  4809777  4802313  4871222  4880636  4868580  7843418  4866954  4868587  4867929  4868254  4871266  4870915  4872161  4870921  7825773  4880675  7824480  7824464  7825769  7824462  4866981  4868584  4868585  4868251  4868577  4868253  4867930  4880655  4880295  4767720  4794373  4809766  4804312  4838120  4831723  4880254  4789268  4788624  7840674  4800603  4886741  4788609  4841883  4802367  4804413  4768184  4789282  4794377  7842086  4789583  4867286  4788619  4837336  4883100  4802366  7841301  4801887  4768189  4880653  7843404  4788534  4806882  4788531  4837308  4880611  7840664  4793650  4881761  4838486  4836558  4767572  4795665  4789595  4795230  4838442  4795229  4794357  4795653  4796565  4796802  4789584  7840584  4841880  4832726  4802308  4789285  7824483  4888064  4798048  4885051  7841907  7841537  4788620  7825762  7841305  4836549  4831713  4837343  4767724  4800601  4897523  7843420  4838086  4806881  4801903  4767625  4886413  4767626  7841540  4832729  4832705  4789283  4806309  4802376  4789280  4795224  4788623  4799301  4788622  4788615  7842071  4837321  4885760  4837317  4789272  7842367  4838104  4809770  7841919  7841551  4801898  4802369  4801912  4801911  7840672  7842060  7841906  4801902  7842369  7842405  7824452  4832716  7842083  7842417  7842416  7842386  4842193  4872163  7841226  7841232  4841543  4842191  4836556  7841559  4841882  7842389  7842385  7840585  4838081  7841538  4800598  7842372  7841578  4841534  4831715  7842090  4795664  4831703  4831688  4831722  4836203  4795669  4795222  4795213  4831709  4831704  4831694  4831695  4831692  4789287  4795668  4795228  7841573  7841326  4872135  7840683  7841579  7840671  7842384  4795656  7824465  7842368  7841525  7841382  4802371  4804410  4804409  4804406  4841525  7842414  7841917  7840606  7840607  7841311  7841546  7840665  7840668  7841329  7841302  7841390  7840684  7841398  7840681  7841558  7842076  7841568  4841524  4883228  7842418  7825770  7824457  7825767  7841513  7841910  7842081  7842087  7841323  7841321  7841383  7842398  7842089  4888047  4888003  4887886  7841317  7841324  7841233  4841521  4841532  7824425  7824448  7824470  7824468  7825759  7824451  7825760  7824484  7824435  7841307  7841518  7841542  4831693  7841239  4795663  7841315  7824455  7841320  4888045  7841554  7841567  7841532  7842370  7841903  4841538  7824436  7840609  7825763  7841577  7842093  7840669  7824466  7825764  7825765  7824482  4841533  7841310  7842094  7841224  7842392  7842354  7841920  7841929  7842078  7842410  4788629  7841516  7842423  7841561  7841924  7842408  7841384  7840618  7842358  7841570  4836552  7842393  7840613  7843403  7843402  4796558  4838107  7842426  4867300  7842412  7840601  7840619  4804309  7841569  4794362  7842355  7841905  4795674  4838080  4867944  4871567  4872134  4868581  4794389  4866968  7841572  7843422  4866631  4885744  4886059  4885052  7843426  7843409  7842420  7841921  7841514  4866633  4841884  4838087  4841547  4842196  4799616  4866646  4866640  4867926  4866638  7824461  4866632  4866653  4795661  4799611  4866635  4866652  7841555  4866651  4810899  4836550  4810900  4810912  4882457  4881762  4836532  7840677  4839298  4832732  4810939  4810932  4810933  4805161  7840588  4788618  7840616  4839284  7843558  4810915  4836537  4836547  4831717  4795225  4795219  4795216  4836545  7843415  4836548  4836542  4836544  7841522  7843429  4836554  4836539  7840608  4831718  7840611  4867152  4866958  4866959  4839308  7840666  4866978  7841237  7841511  4839301  7842396  7841300  4804923  4886728  4841871  7841385  4802316  4867149  7841535  7840587  7841574  7841556  7824474  4793658  4802320  4801896  4866964  4806734  4867291  4836534  4838376  4838499  4842515  4842514  4842195  4810928  4805164  7841918  4804919  4886737  7843405  7843557  7843556  7843427  7824478  4798968  7842425  4798050  7841220  4804913  4798056  4867934  4867925  4867301  4797656  7843551  4796786  4806311  4888049  4887977  4885746  4797670  4798960  4798068  4794365  4837309  7843410  4884431  4886066  4799293  4798066  4798065  4796809  4797667  4838094  4867949  7843428  7842373  7842427  7842061  4806314  4806722  4806315  4867292  4866980  4838073  4874231  4804314  4887893  4871259  4883249  4883247  7843417  4806727  7843559  4804920  7843406  4798953  4866957  4867947  4871564  4836573  7841916  4798060  4795647  7843550  4797662  4795651  4795655  4798956  4838119  4837329  4867297  4836571  4800832  4800831  7842088  4788626  4788518  7840602  4886724  7843554  4841528  4809769  4804307  7841901  4841887  4841886  4841885  4842211  4841874  4806094  4810895  4837349  4788616  4842506  4880625  4796791  4841530  4888004  4887879  4805160  4838379  4804918  4837344  4888005  4888013  4872152  4809744  4809747  4886738  4841872  4887982  4800592  4801913  4884426  4800824  4800602  4801909  4880622  4880286  4838106  4800597  4841542  4796557  4871261  4883089  4802374  4866630  4842504  4802368  4801906  4800827  4841881  4841870  4800833  4801908  4802315  4841545  4800828  4802311  4800599  4801916  4800826  4800594  4884423  4800834  4800829  4841879  4841873  4806730  4800593  7841389  7841222  4838500  4837328  4837350  4837347  7841325  7841544  4838390  7841386  4838391  7841530  7841560  7840676  7841223  7841553  7840614  7841565  7841236  7840603  4837324  7840589  7840604  7840661  7840679  7840678  7841557  4838388  4796788  7842062  7840675  7841392  7841387  4867948  4867938  4838380  4838384  7841539  7841524  4810904  4810903  4810893  4788527  7843408  7841923  7842366  7841327  7842371  7841550  7841536  7841393  7842406  4804922  4871272  7841566  4867942  7842356  7842098  7842064  4838079  7841545  7841303  7842099  7841552  7841309  7842387  4810898  4837348  7841227  7842070  7842082  7842096  7841519  4868578  7842378  7841229  7840673  7842402  4810902  4795231  4795666  4795232  4806099  4838089  4792795  4842214  7841902  7824432  7842428  7841318  7841571  7842421  7842353  7842409  7841914  4837330  4789589  4883253  4805154  4805159  4806098  7841238  7904608  4805155  4804306  4842189  4883227  7843423  7824460  7842072  7842407  4883244  7843552  4804311  7842383  4797663  4793657  7840583  4796561  4806102  4806100  4806097  4806092  4805168  4805167  7840581  7841510  4805165  4804315  7841526  4810897  4798069  7840662  4810921  7824447  7840670  7841306  4797673  4799298  4810923  4799297  4798071  4798955  7841396  4798958  4798966  4798961  4798055  4795650  4796805  7843419  7841563  4798957  4798052  4796574  4789289  7904607  4788636  4870918  4839300  4794361  4789276  4880584  4874415  4868575  4842210  4841546  4837312  4795662  4804909  4789281  4788621  4789591  7842380  4888038  4789588  4883232  4872138  7842399  4788633  4874255  4870914  4809735  4806316  4831691  7824475  4882452  4806726  4804401  4804302  4797658  4792799  4798964  4767721  7841575  4836540  4838088  4795209  4767573  4794386  4838375  4836565  4798058  4796806  4796793  7842077  4831720  4831725  4806091  4886060  7840660  4831724  4841539  7873204  7843555  4768193  4767715  4874413  4886725  4796559  4806720  4796573  4768188  4768190  4871264  7904612  7841316  4805156  4888059  7841328  7842390  4883103  4795667  4806890  4806888  7841531  7842365  4886068  4839293  4880631  4767723  7843424  4804912  4837323  7904823  4767575  7904821  4804921  4798951  4884436  7842401  4788516  4883231  4871275  7841399  4837313  4838487  4838115  4767716  4887990  4838502  4797659  4789604  4767725  4788523  4841877  4793661  4838448  4789278  4842509  4795208  7841534  4806101  7824472  7824473  4806893  4798049  4809745  4809742  4795217  7824481  4806892  7842400  4798051  7824437  4838489  4837310  4797665  4886415  4867943  7824441  4789600  4809733  4806320  4837335  4882456  4806090  4837337  4883102  4831721  4872142  7842360  7842074  4767718  4788612  4792796  4872139  7841523  4832714  7842095  4810927  7842413  4794358  4789291  7842415  4802317  4789587  4794379  4838443  4798965  7841562  4795657  4793651  4810929  7841313  7842352  7842395  4794360  7842079  7842422  4806096  4794380  4838392  7841533  4871213  4794372  7842363  7841308  7840680  4796563  7841927  4801886  7841576  4871263  4871569  7842362  7841515  4797657  4788530  7842419  7842351  7842397  4794371  4799612  4789585  7841549  4872146  4871274  4794367  4871260  4794388  4866645  4869222  7843553  4802373  7842411  7824428  4801899  4768191  4795207  4789596  4836569  4799610  7904618  7904611  4838114  4798063  4810937  4789597  4809739  4793665  4792797  4789593  4789603  4789605  4800822  4789598  4887880  4767577  4799619  4841531  4836535  4804316  4809764  4888053  4880279  4888012  4888014  4841540  4788613  4838113  7896798  4809780  4883234  4801889  4795211  4789602  4810926  4810909  4810917  4792798  4888031  4793662  4810919  4795220  4794385  4806318  4839286  4831687  4809760  4809778  4810901  4798967  4793667  4795654  4810920  4836541  4836530  4810924  4810930  4839289  4831696  4837307  4810908  4838111  4810918  4810922  4842218  4796571  4804916  4795675  4788522  4872149  7842065  7843416  7842063  7840682  7843421  4788519  7841380  7842382  4908027  4908026  4908529  4897581  4880587  4880586  4907532  4880666  4888046  4880589  4908663  4908667  4907868  4908029  4897585  4880638  4897586  4907862  4907865  4908661  4887867  4881775  4908666  4904183  4903873  4903869  4903864  4904182  4903866  4903876  4904184  4888044  4888042  4897871  4887898  4903875  4887984  4888067  4907867  4888028  4888052  4888040  4887897  4887980  4887877  4907540  4908028  4886732  4888039  4888022  4897869  4907538  4883097  4883091  4882454  4881760  4908668  4908030  4908024  4881767  4883240  4883094  4883098  4883224  4881757  4907870  4880665  4903867  4884424  4884425  4880604  4883241  4908032  4897868  4908524  4907539  4888018  4887868  4888010  4882443  4880639  4914832  4874240  7904616  7904605  4914812  4914815  4914818  4914809  4915173  4915175  4915179  4914817  4914826  4914828  4837346  4842192  4836543  4836536  4874417  4874244  4897588  4915174  4914824  4788608  4872157  4914808  4914810  4914822  4914833  4915172  4915178  7904825  4915169  4874239  4915180  4915181  4915176  4914814  4914820  4914821  4915170  4915177  4914816  4914830  4897883  4874224  4880656  4880613  4880552  4880647  4897515  4897590  4897582  4888011  4897887  4888006  4897521  4907863  4897520  4897519  4897516  4897524  4908526  4908664  4888063  4897587  4897870  4874419  4874416  4832703  4806733  4788627  4806313  7842404  4832724  4880553  4897872  4886722  4897874  4897888  4897879  4897525  4897878  4897877  4886735  4888037  4888025  4897873  4887996  4897885  4888015  4887993  4888033  4796801  4880634  4880642  4880677  4887881  4888050  4888071  4888051  4886071  4888016  4887974  4888007  4888043  4888017  4880635  4880651  4887999  4880292  4880602  4880571  4880284  4897529  4897876  4897589  4897579  4800825  4793664  4897880  4897882  4897527  4897884  4897889  4888034  4897886  4880551  4897514  4806887  4799294  4880283  4880550  4880659  4880661  4880670  4880570  4880280  4880627  4836570  4886070  4885050  4885060  4884434  4884432  4884430  4885057  4885742  4885063  4885061  4886077  4888041  4885761  4886072  7841564  7841543  7873208  7896791  7873206  7841397  7896792  7896807  7896808  4880676  4888029  7896804  4788528  4886062  4880630  4887874  4888068  4880291  4881765  4880588  4880641  4880548  4806885  4809763  4885743  4794364  4885752  4886726  4886744  4837326  4885759  4794376  4802370  4831710  4885046  4885755  4886067  4880623  4798959  4885062  4880261  4907869  4880271  4882448  4882450  4887890  4881759  4887887  4884440  4884429  4885059  4885055  4880289  4767574  4881758  4903863  4903861  4904188  4904185  4882449  4882446  4897522  7842073  4880262  4880609  4884441  4887998  4888023  4888027  4888056  4880290  4880282  4880255  4908025  4907537  4897528  4908031  4907864  4880567  4881771  4880549  4880257  4887888  4887994  4897530  4886073  4908530  4907535  4880263  4880264  4880287  4887987  4887992  4881769  4907536  4884437  4908665  4880652  4880603  4880565  4880258  4881773  4880658  4908527  4880590  4903872  4897526  4881774  4880601  4888069  4907531  4903857  4908523  4880668  4908660  4903858  4903860  4903859  4903874  4881772  4881776  4903868  4907534  4897881  4897875  4885751  4887989  4887894  4888020  4888062  4888030  4888009  4897583  4880297  4897517  4897518  4881763  4880660  4881770  4888026  4887979  4886733  4907533  4897580  4881764  4831701  4903865  4870917  4871254  4880626  4870911  4794366  4831686  4796552  4832735  4838074  4795215  4809734  4874254  4838490  4867289  7843412  4795210  4789592  4866966  4806317  4867305  4871212  4809736  4842213  4788610  4794387  4806307  4796797  4841523  4798061  4866975  4796569  4871220  7842080  7840600  7824427  7841395  7824445  7873581  7841541  7825755  7825768  4839307  7840586  7824453  7841228  7841548  7824444  7842067  7824431  7902608  7902610  7896793  7873583  7842381  7896800  7896799  7824458  7841900  7841304  7824434  7873207  7902612  7824433  7824429  7873210  4880267  4880615  4809759  4884435  4880614  4872137  4874232  4881766  4887872  4888070  4887975  4841529  4880583  4880585  7904604  4880628  4841544  4880293  4837342  4887983  4880600  4880616  4883235  4888032  4887896  7904610  4809757  4887878  4801895  4880645  4880572  4883095  4883104  4883088  4882462  4880607  4882458  7902605  4795652  4886731  4767714  4908525  4903871  7841928  4880672  4872162  4909676  4802377  4886418  4886412  4887985  4904186  4904187  4888001  4903862  4886414  4888060  4887884  4788632  4867302  4810896  4842503  4880260  4880605  4874256  4838386  4832706  4841535  4841541  4839285  4866629  4867046  4885750  7904601  7841314  4866970  4866974  4867148  4885756  4885049  4886750  4796556  4831698  4871265  4871209  4881768  4883105  4866976  4874243  4880663  4809772  4883099  4883236  4880298  4867295  4885748  4888035  7904600  4880667  4806095  4883087  4886061  4874250  4888024  4882455  4880671  4880669  4880569  4832702  4887873  4874227  4874421  4874418  4874234  4874246  4832737  4883238  4871267  4888000  4832723  4798067  4887891  4839291  4801907  4872155  4883233  7841922  4883226  4871269  4871216  4882445  4871258  4880252  4797664  7896796  4838090  4839311  4838121  4868250  4794383  4867940  4867935  4867945  4866955  7824430  4872140  4880281  4832712  4867290  4866627  4866956  4871257  4886745  4883229  4869213  4872160  4871273  4872136  4897584  4880643  4868583  4888055  4886727  4768187  4867294  7904602  4880256  4872145  4888061  4880606  4880649  4888058  4880610  4874252  4888002  4887885  4836555  4866952  4870919  4871211  4867151  4866972  4872164  4838093  7904826  7904827  7904603  4880650  4867293  4887892  4866963  4887997  4888019  4880253  4883254  4883256  4883101  4882451  4839296  4874228  4874229  4874242  4874226  4880566  4866626  4887995  4832713  4842215  7904829  4806731  4886074  4867931  4867936  4838077  4866636  4871566  4866647  4867306  4886075  4866953  4866637  4874248  7841912  4872148  4874241  4874249  4867047  4866951  4884422  4866967  4867304  4866977  4866965  4880640  4874235  4880568  4880278  4882460  4867045  4886736  4801910  7904613  4870920  4867928  4880629  4887876  4836533  4871219  4871276  4795670  4887991  4867946  4801894  4866648  4866642  4866644  4866649  4866654  4866634  4866643  4897513  4842516  4871207  4874414  4866641  4866639  4883250  4883255  4869224  4880633  4866650  4806729  4767629  4795648  7902613  4886417  7902606  4880269  4880644  4880632  7904619  4887976  4838095  4798062  4871256  4868576  4794369  4874253  4874233  4886078  4885749  4870922  4870916  4867303  4867299  4832710  4866973  4872153  4880637  4887986  4874230  4884438  4866962  4838449  4871268  4869223  4882453  4886730  4880554  7896806  4869221  4866961  4871210  4795659  4806088  7873205  4799613  4841875  4874238  4809765  4871270  4886069  4886723  7896809  4796553  4809741  4809756  4809740  4809737  4836557  4836551  4793668  4839305  4794368  4793655  7841221  4836563  4836562  7840612  4836528  4809762  4796789  4838491  4837311  4795658  4831714  4797669  4796570  4788515  4836559  4886748  4794392  4767722  4799295  4838109  4809751  4794391  4798950  4806895  4831708  4809779  4809732  4805166  4795660  4809776  4804911  4867288  4871218  4866979  4867296  4867048  4788635  4788611  4832708  4872147  4809773  4867941  4794393  4870913  4802314  4788520  4837339  4831711  4809761  4832725  4837334  4802321  4789274  4806089  4809768  4838098  4837319  4880662  4796804  4799615  4832709  4887895  4871570  4801900  4866971  4806883  4788607  4838096  4842216  4872141  4802318  4831697  4885745  4831690  4804303  4794395  4806312  4804310  4872158  4788529  7896790  7896801  7896794  7873582  4880265  4838082  4886749  4795223  4801901  4887882  4888008  4888021  4904189  4887869  4880266  4836574  4880288  7842097  4838377  4802380  4887978  4880654  7824471  4832890  4809746  4810925  4887889  7904614  4789288  4886734  4788521  4804305  7896797  4796560  4888057  4799618  4886065  4837333  4839302  4874420  4874225  4839299  4832736  4838387  4799290  4838099  4838385  4885044  4880296  4880646  4837338  4887883  4798064  4836529  7902611  7902609  7873584  4806728  4806721  4801915  4832711  4831699  4832700  4800830  4809752  4842212  4802375  4796554  4842513  4841878  4810891  7841520  7841913  4810934  4789270  4789601  4767624  4841526  4842505  4883096  7842388  4880582  4885753  7843401  4904181  4806308  4800600  4886063  4882447  4882444  4842511  7825771  4903870  4887875  4882461  4832701  4793659  7841394  4842190  4883248  4841876  4839290  4880673  4842194  4796787  4908528  4888048  4883090  4809738  4809758  4838108  4809743  7841521  4798954  4842217  4868588  4797655  4883230  4874245  4874251  4874237  4880608  4880617  4838495  7824477  4883243  4883222  4883092  7843414  4839306  4810914  4838501  4789269  4797672  4804414  7825761  4804407  4804402  4883245  4868586  4810894  7842092  4871565  4806732  4837332  4886064  4836538  4799304  4832733  4871221  4871214  4794363  4871215  4872144  4789277  4884433  4831706  4841536  4838382  4836531  4793666  4804304  4810935  4798053  4796794  4885054  4839287  4883093  7873585  4872154  4832730  4799292  4880624  7904822  4871568  4836567  4839288  4842510  4842502  7904828  7904824  4789286  4874236  4888054  4868574  4837305  4838447  4831712  4888065  4838110  4839309  4831700  4837352  4838078  4837340  4809750  4832731  4871262  4874247  4887981  4842501  4788524  4831702  7904617  4867287  4832734  4806319  4836553  4767576  4838389  4832722  4869215  7842350  7841528  4871223  4831707  4832699  4801892  4805157  4806886  4832715  4809748  4806889  4838381  4795203  7842357  4795649  4839304  4838493  4793660  4809781  4793652  4796796  4798969  4795676  4795204  4794394  7841312  4793663  4832719  4832717  4806894  4900602  4900604  4900610  4900609  4900603  4900608  4900605  4900569  4900621  4900629  4900628  4900623  4842851  4900606  4832885  4842852  4885431  4900626  4832881  4842850  4832880  4832883  4832886  4832878  4832882  4832884  4897906  4832877  4832879  4832876  4897905  4897904  4885430  4885429  4832887  4900625  4900622  4900624  4900620  4900627  4900607",
      "more_code_info": ""
    }
  ]
}