      "status": "Terminated",
      "city": "Sylmar",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "78093",
      "recalling_firm": "St Jude Medical Inc.",
      "address_1": "15900 Valley View Ct",
      "address_2": "",
      "postal_code": "91342-3577",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution",
      "recall_number": "Z-0037-2018",
      "product_description": "PM 1260 ASSURITY +  These low voltage (LV) devices are implantable pacemaker pulse generators, intended to be permanently implanted in the body, that have a power supply and electronic circuits that produce a periodic electrical pulse to stimulate the heart.  These devices are used as a substitute for the heart's intrinsic pacing system to correct both intermittent and continuous cardiac rhythm disorders.",
      "product_quantity": "6051 units",
      "reason_for_recall": "New pacemaker firmware was developed to further mitigate the risk of unauthorized access to our pacemakers that utilize radio frequency (RF) communications.",
      "recall_initiation_date": "20170828",
      "center_classification_date": "20171019",
      "termination_date": "20200708",
      "report_date": "20171025",
      "code_info": "Serial Numbers:  4471223  4435620  4471226  4478259  4473923  4465800  4465795  2960123  4471236  4465866  2960119  4480805  4456626  4456605  4456614  4456609  4455665  4470374  2992215  4470251  4478226  2992206  4435532  4436165  4442557  2888746  2992230  4470249  2992219  4481080  4477111  4486603  2992337  4478246  3028382  4470221  2992336  4470256  4470255  4477145  4477154  2992244  4479101  4470215  2817358  2897784  2897779  2897791  2878499  2912102  2912129  2912137  2912163  4427305  2992294  2992270  2992271  2870627  2888748  2912140  2912167  2912110  2912100  4419728  4428451  4495360  4428446  2992205  4479087  4485695  4486629  4486616  4486619  4486415  2992291  2992304  2992208  3028378  2992262  2836144  2992265  2992233  2992331  2992298  2992316  2992225  2992322  2992338  2992221  4480798  2992275  4583284  4507998  4505995  4508010  4540176  4495349  4501470  4510860  4508966  4495359  4535417  4507981  4427067  4508009  4552833  4501456  4505305  4505336  4505325  4505323  4506025  4506017  4507994  4505996  4486592  4568194  4510873  4589129  4548069  4513308  4510893  4450734  4495355  4599671  4538400  4507977  4552843  4505329  4506013  4505297  4505998  4545146  4495352  4505307  4495365  4444189  4420503  2880064  2992334  4486626  4486643  4428621  4486590  2866157  4490599  4510886  4444178  4595435  4435633  4481085  4450718  4450724  4450716  4450730  4478258  4478225  4452570  4444187  4535401  4450688  4508013  4508988  4508971  4508983  4510007  4510031  4510878  4508960  4510011  4522001  4505987  4535413  4535405  4536286  4536287  4538398  2912136  4538415  4510012  4536284  4509990  4510987  4441673  4445638  4445615  4445647  4444191  4444206  4444169  4441652  4441667  2958460  4436172  4436157  4436241  4435240  4435220  4435258  4428465  4428473  4428463  4442581  4427298  4429259  4417850  4419718  4419726  4428439  4428461  4428459  4428619  4429264  2992263  4436159  4502697  4502681  4502703  4502700  4502692  4508961  4513296  4428447  2992279  4436184  2992284  2992223  4417858  2992295  4418150  2992227  2992202  2992285  4462727  2992305  2992247  2992255  4513550  2992333  4445644  4513541  4456618  4455658  4462697  4462677  4462679  4445614  4442609  4473268  4445637  4490597  4445630  4445634  4445646  4442574  4442615  2961574  2961576  2961572  4445622  4480800  4513544  4505301  4505312  4521992  4522002  4490593  4476725  4513324  4479012  4477126  4490598  4490606  4490601  4490613  2959508  4476722  4473911  4475375  4478250  4615403  4506009  4552840  4552824  4508976  4634791  4540167  4621323  4634365  4540184  4538399  4508018  4597323  4589124  4615432  4601896  4638503  4634789  4545147  4513300  2929724  4615408  4507991  4634342  4601959  4552806  4589111  4621334  4552844  4601906  4597324  4552829  4540183  4536290  4615395  4601960  4604147  4604150  4634786  4634773  4604143  4558996  4595443  4552830  4615409  4589125  4615427  4548071  4638486  4552841  4604138  4641844  4599669  4552817  4597317  4632934  4615422  4634334  4641838  4452563  4464622  4621342  4634357  4615412  4574102  4604148  4595438  4634360  4604152  4583289  4548050  4450676  4634355  4601963  4548063  4601894  4583305  4582962  4601903  4582967  4601893  4450844  4621321  4464586  4465789  3028385  4455674  4455668  4462681  4462713  4464620  4464609  4450726  4464604  4450720  4464600  4464338  4450709  4465780  4465821  4465804  4465869  4464373  4464606  4464367  4452595  4452597  4634340  4548068  4452602  4452568  4452559  4452617  4479092  4464591  4462683  4462672  4464619  4464336  4464608  4450706  4450692  4450732  4450690  4452590  4457510  4465791  4455669  4455662  4462731  4464594  4464628  4464376  4456630  4456610  4464362  2915028  4464358  4462714  4464335  4464354  4464590  4464610  4464589  4464342  4464343  4465788  4465778  4464369  4465809  4465808  4464384  4465824  4457519  4452609  4456624  4457502  4464598  4464366  3028381  4456604  4484524  4457508  4457524  2958467  2920003  4457505  4604140  4548064  4595444  4582974  4582977  4582983  4599673  4638511  4634769  4621331  4638498  4601965  4589117  4589128  4545162  4552820  4540168  4574101  4574106  4634331  4601899  4548049  4615431  4638475  4636843  4634793  4558991  4599670  4548059  4638506  4595426  4638494  4548057  4589126  4595427  4595428  4595441  4638497  4638485  4621340  4548062  4480811  4574104  4583281  4621343  4615440  4558999  4558998  4545158  4538417  4583293  4601892  4540174  4480815  4477141  4479093  4479024  4479014  4456601  4452589  2956555  4552813  2958462  4604139  4621324  4510852  4548052  4621344  4545163  4545148  4552839  4582979  2874087  4604145  2915049  4615407  4615405  4589115  4621346  4626457  4634368  4634782  2874100  4597325  4548065  4634752  4638482  4548047  4548051  4548048  4548053  4597322  4601957  4582968  4548067  4583287  4565686  4583288  4615398  4583299  4540173  4540198  4595442  4634754  4634776  4597331  4595431  4548055  4601898  4513331  4552827  4634338  4552825  4604144  4634777  4548066  4634367  4601895  4615425  4552846  4595437  4601900  4597329  4597330  4615415  4582989  4548058  4615420  4601904  4604137  4615417  4559443  4604151  4604146  4638501  4574103  4597318  4601897  4583298  4552821  4599676  4589118  4634364  4634358  4552815  4595430  4552834  4545153  4615430  4552808  4552838  4540190  4536301  4538406  4641847  4536295  4583316  4634351  4615441  4615433  4552831  4548070  4589121  4589119  4582981  4582965  4540196  2952259  4486608  4479086  4484246  3028377  4484513  4487206  4486642  4487214  4487215  4480820  4486599  4483699  4486413  4485704  4486620  4486591  4486411  4485707  4485706  4486596  4483976  4486604  2872076  2912171  2912124  2912158  2912104  2912141  2912105  2912148  2912112  4483979  2872103  4485697  2872090  2846347  2846361  4480825  3025482  4480834  4483957  3025483  4484511  2874092  4483977  4483962  4484512  4484533  4484528  4484523  4483980  4484514  4483954  4485694  4483964  4484535  4483978  4483975  4483692  4483956  4483974  4483970  4477140  4478233  4483700  4486600  2915732  4485702  4486611  4486598  4486410  4486625  4486615  4486621  4486635  4483967  4485698  4484526  4486637  4484534  4484521  4484525  4484517  4484245  4483968  4483969  4485693  4485705  2920032  2927668  2923624  2919965  2919994  2919991  2919975  2920018  2920024  2915700  2919959  2927656  2920034  2920023  2919970  2920028  2920025  2918067  2918056  2918040  2912099  2920033  2929718  2956552  2952261  2952263  2952255  4456602  2929712  2929720  2929717  2929715  2927669  2929713  2927660  2929721  2929723  2929726  2929725  2927666  2927663  2927662  2927661  2915698  2929714  2920030  2874090  2874089  2874088  2870619  2832948  4480832  2919955  2923627  2918039  2923628  2912156  2919978  2920007  2920011  2920015  2919972  2918065  2919958  2915710  4479036  2923625  2915013  2915742  2915695  2854998  2915040  2915029  2915018  2915017  2915015  2915725  2912165  2915026  2912115  2915011  2915730  2915039  2915044  2915718  2912117  2912133  2912135  2915032  2915706  4464363  4481056  4465790  4464584  4465792  4481076  4465799  4465816  4464359  4477153  4464371  3028383  4464368  4464370  4462701  4464372  4464345  4464337  4462700  4462691  4464380  4464364  4479032  4479022  4478997  4479018  4479043  4479052  4479047  4479031  4480818  4465863  3026689  4470211  4479105  4479056  4471215  4470205  4486605  4479002  4479040  3026691  4476731  3026690  4456622  4456625  4455666  4455655  4456600  4455682  4455678  4455677  4455671  4452598  4456628  4456615  4455660  4455684  4462666  4462725  4462716  4452608  4452569  2872102  4450678  4456617  2959504  4480826  4483693  4455672  4457509  4456631  4455667  4457497  4457491  4456620  4455664  4462726  4455680  4480831  4457517  4456629  4455676  4456632  4455656  4456606  4455657  3028379  4462669  4478239  4475378  4477125  4479025  4479050  4478243  4478248  4476730  4478224  4471261  4478244  4479028  4477122  4477108  4477138  4477144  4477143  4477132  4477151  4476743  4479100  4478222  4480833  4478232  4479051  4480819  4480790  4480794  4481083  4481067  4480816  4480830  4479006  4480796  4479099  4481062  4480813  4479088  4479023  4479027  4479096  4479055  4479053  4471248  4480806  4464605  3025479  2958463  2958468  3026695  2956551  4456607  4464630  4464374  4476742  4464621  4471225  4464615  4456603  4484243  4484520  4484537  4484518  4483698  4483955  2952258  4464378  4470373  4462662  4471199  4471245  4471211  4471242  4470226  4471235  4471230  4471209  2952260  4470203  4471216  4470375  4471214  4471219  4471231  3025480  4486630  4486595  4477112  4479011  4470200  2956559  4452565  4479049  3025481  4479008  4465861  2958469  2959506  2959509  2959497  2959498  2959496  2959495  2959505  2959503  4470257  2956550  2958465  2958464  2958466  2958461  2956553  2956556  2956558  2952257  2952254  2959500  2960120  4479098  4470243  4484509  4486607  4486602  4484248  4484250  4483953  2960117  2961573  2960118  2960109  2960111  2960115  2960105  2960114  2960112  2960106  2960110  4471210  2960113  2952256  2878500  2878496  2878510  4428475  2853271  2853272  2893460  2894444  2894436  2894461  2894458  4428452  2897780  4428441  2897762  4428464  4428438  4428627  4428460  4427053  4428440  4428458  2897792  4435229  2829689  2829481  2820568  2822826  2822828  2820564  4436244  4435260  2846366  4435631  4419700  4435239  4435237  4435207  4428623  4429300  4429287  4429260  4428442  4435538  2912160  2897764  2900976  2902522  2902539  2902520  2902534  2902548  2902544  4427292  4419711  2897778  2912121  2912170  2878471  2903440  2903435  2903439  2903442  2903444  4420025  2897760  2903443  4420041  4420023  4420026  4420526  4420519  2897763  2897759  2897793  2897787  2897767  4419682  4419686  2894439  4419710  2829684  2897772  4436176  2897773  4427070  2897774  2870623  4478236  4488969  2894467  2894460  2893456  2893430  2893425  2893426  2894459  2894457  4450710  2870622  4450713  2893459  4488974  4488984  4488982  4488968  4488985  4488967  4488975  2862244  2865768  4429266  4488965  2894466  2893432  2893433  2893438  2893458  4486414  2897783  4490595  2893446  2893440  2893445  2894473  2894468  2870615  2893451  2893462  2894464  2894465  2894454  4450694  4450685  4450680  4450675  4450851  4450698  2893447  4441633  4488972  4444196  4441655  4441657  4442571  4442569  4442556  4442594  4442602  4442560  4444159  4441676  4444164  4441644  4435621  4436147  2992303  4435533  2846363  4435624  4435528  4436153  4436156  4436183  4441677  2897785  2872075  2872078  2888735  2872080  2872086  2870616  4450852  4452558  4452578  4473916  4444198  4473896  4435628  2870613  2888753  2894435  2894452  2894437  4446206  4442613  4445660  4445656  4442575  4444186  4473881  2992231  4489636  4489642  4490589  2897781  4470212  2897771  2897766  2888731  2888724  2888733  2845619  4484529  4473266  4473917  4473921  4473256  4470208  4473259  4473260  4473261  4473284  4471212  4473241  4473257  4473901  4435238  4435234  4435636  4435622  4435626  4435215  4435216  4435213  4435223  4435554  4470247  2894462  4473235  4487204  4487203  4487207  2894445  2886257  4473887  4473884  4473240  4473269  4473889  4470365  4479097  2894453  2865764  2862258  4479007  4473892  4464592  4465862  4464382  2897775  4465776  2888728  4470372  4470228  4470245  4464340  4465876  4465820  4465815  4465823  4470361  4470364  2897788  2888726  2992280  2829688  2829687  2829685  2829482  2822827  2817365  2817352  2817371  4455661  2992228  2894447  4486628  4479046  4436186  2992326  2992276  2992257  2992256  2992307  2992259  2992216  2992218  4470207  2992232  2897777  4478256  4493139  4490619  4464601  4464593  4493136  4493138  4493152  4490587  2817674  2894455  2829483  2894456  2817355  2870636  2880080  2870634  2870618  2870629  4464597  2894446  2894463  2845604  2817369  4462661  4479004  4479003  4479021  4479020  4479034  4479030  4479026  4464351  2872098  2872101  2894474  4462705  2870621  4462675  4479019  2866152  2870624  4489632  4489631  4489644  4489649  4489635  4489640  2872091  2845629  2880143  4510990  4505980  4583319  4538408  4508970  4509993  4509998  4508969  4508982  4486638  4510868  4508990  4510866  2846358  2845627  2845603  2880135  2865782  2862262  2858076  2862264  2858093  2858084  2831549  2865792  4508989  4507996  2862267  2862247  2862248  4510859  2865763  4510017  2865781  4508963  4508987  4508980  4508978  4510870  4502687  2873282  4508015  4495369  4507974  4508007  4508974  4508967  4510024  4507973  4508022  4508017  4508016  2845611  4507993  2872104  2832936  2880084  2880068  2894441  2894449  2894451  2897789  2897768  2897782  2886261  2886259  2880148  2866155  2880071  2872106  2872083  2872077  4510015  4510030  4508968  4508991  4510027  4510001  4510862  4510876  4510342  2886269  4436175  2880069  2880134  2880144  2817709  2865788  2866160  2865783  2880079  2865785  2870641  2880081  2880087  2846389  2865784  2880138  2880086  2880076  2846377  4489637  2845609  4489648  4489647  4510980  4510992  4510877  4508965  4489650  4486639  4477101  4479013  4487202  2858082  2880063  2817359  2880065  2846362  2811909  2866159  2865778  2865773  2811912  2858079  2855027  2855028  2893424  2893427  2893428  2894470  2893453  2888752  2845617  2845606  2862259  2820561  2811925  4488989  2880073  2853269  2853263  2853262  4484247  4485691  4488973  4485700  4488971  4488966  4488963  4488962  2880152  2888751  4479091  4488976  4488983  4490592  4490591  4479089  4479094  4486644  4479095  4479039  4479029  2858118  4480792  4485696  4666232  4419703  2878506  2893431  2846335  2893461  2862260  2865777  2862245  2865774  2865772  2865767  4419709  4666250  4419699  4666231  4666236  4666241  4666251  4666261  4666233  4666243  4589123  4568196  4568202  4568203  4552845  4666258  2858115  2880151  4476733  4480810  2878502  2878501  4480786  2878514  2858121  2893444  4480812  4419684  4486624  4486614  2903437  2811910  2811916  2858114  4419706  2888747  4486416  4510881  4510993  4509992  2903436  4486640  4486601  4486622  2825188  2820562  2992248  2992327  2992299  2992236  4540200  4489645  2827250  2845625  2845623  2836145  2836155  2836148  2992302  4511624  4552849  4505333  4510000  4508984  4507997  4540187  4538412  2825194  4509995  2825196  2845616  4511619  2845614  2836161  2836162  2846350  4508001  4505320  4540197  4513323  4510871  4510985  3116267  2846375  4442598  4486594  2912169  2912108  2912123  2912149  2912153  2912145  2912114  2912130  2912125  2912122  4429286  4419692  4452572  4457498  2902537  2902540  2927670  2831548  2888737  4428466  2858095  4420048  3028376  4452555  2832939  2992210  4479016  2992281  4510034  4508011  2915038  2915739  4495372  4495371  2832949  4507990  4457493  4510889  4450681  4420508  4420515  4455673  2992320  4506002  4435245  4489641  2897776  2897770  2920002  4420501  4427073  2858128  4429257  2927657  4435545  4435634  4435251  4435244  4435556  4452564  4426784  2874102  2992214  2992289  4450855  2858101  2862266  2865766  2888723  2874097  2992286  2888727  2888729  2884440  2897786  4452567  2918041  2874111  4429274  4427270  4427060  4427065  4427066  4427048  4429290  2878527  4429292  4429277  2858111  2866153  2870631  4428456  4429269  4427295  4427061  2831562  2831551  2831552  4420513  4505335  2919966  2920016  2920008  2992268  2992212  4470235  4470230  4495363  4495357  4495368  4505331  4502708  4508981  4508014  4510857  4475368  4419712  4476727  2846355  2870642  2846365  2845621  4471213  2831554  2831565  4495351  2831543  4417856  2915723  4418141  2878518  2880149  2915033  2915042  2915719  2915738  2915016  4484527  2919983  4486632  4418151  4418140  2927664  4428445  4435264  4435228  4442552  4476721  4442583  2915705  4418154  4471217  4495386  4508002  4508975  4510028  2915037  4418157  4477146  4436180  4445642  4476734  4475379  4477104  4477105  2960107  4479090  4485703  2919967  4475380  2920022  2918057  2918049  2918042  2918064  2920017  2920009  2919997  2920000  2918060  4417854  4417857  4429294  2920014  2915729  2902524  2915699  4427074  4427050  4429293  4429298  4418145  2915743  2915741  2915727  2915733  2915744  2915722  2915702  4429299  2902528  4429283  4428477  4428453  4428448  4428618  4428471  4420033  4420509  4428462  4420510  4429265  4419695  2900977  4420497  4420523  4420518  4420032  4420044  4420494  4420504  4418146  4418156  4418152  4417849  4429279  4429263  4428626  4428622  2880150  4428468  4428620  4418142  4429255  4429270  4429272  4428444  4428443  4428457  4418158  2915740  442965  2888745  2880072  2880083  2880074  2880075  4452581  4452623  4473237  4452576  4493154  2865786  2866161  2865776  2865780  4452588  4452603  4452616  4493147  2900969  4473245  4473253  4473251  4489643  4493157  4493145  4484531  4493153  4493150  4493155  4489639  4493151  4493137  4493142  4452580  4493146  4462729  4452600  4452560  4462676  4462684  4471237  4473243  4450729  4462728  4452545  4462696  4462688  4462724  4462732  4462692  2878516  2878524  4473281  4478242  4473262  4452548  4452624  4452619  4452582  2878519  4452575  2878493  4452614  4477109  4478247  4455683  4462709  4445613  4452592  4452566  2878525  4473890  4484510  4477142  4476748  4473277  4473914  4473899  4473244  4473895  2854981  4473915  4473250  4473255  4473282  4462710  4471240  4473906  2894438  2902535  2902521  2900974  2900979  4429253  2900978  2846346  2897769  2854972  2894448  2893449  2893437  2893435  2880145  2855014  4470204  4420027  4475371  4471208  2880070  2874112  2854983  2878508  2880136  2880078  2874098  2880085  2888756  4476735  4477133  4477127  4476729  4473270  4473279  2878512  4470202  2902543  4471197  4471196  4471238  4471254  4471243  4428455  4471203  4470258  4470222  4470377  4470359  4470219  4471249  2886273  2886271  4470363  4505989  4506007  4506010  4506000  4505994  4506019  4506008  4506012  4505981  4506020  2870628  4505985  4502704  4505306  4502705  4505988  4505992  4505339  4538413  2854979  2854985  2854991  4501455  4506022  4502679  4501463  4501464  4501498  4501485  4502671  4502717  4501459  4502672  4502680  4502701  4506003  4502684  4506024  4502706  4502676  4502699  4502715  4502696  4502711  4502685  4501499  4501462  4505343  4501477  4428472  2846373  2846360  2846336  2846371  2846383  2845615  2845601  2846344  2886260  4615396  2886266  2880139  2886277  4417851  2886270  2886265  4535411  4615413  4615410  4507985  4508000  4506011  2886267  2874099  2872092  4535414  4538397  4615435  4615414  2854968  2878498  2878509  2878521  2874108  2846388  2874101  2846342  2878491  2880146  4538395  4538407  4536305  2880147  2880142  2880137  2880140  4502710  2874103  4471207  2832950  4435529  4462715  4429302  4502719  4501450  4470227  4470234  4435625  2888739  4436168  2888725  2888732  2888722  2858122  4435630  4450719  2858125  2832955  2832954  2884441  4436163  4435231  4435218  4435224  4435225  4436195  4436181  4436185  4436173  4435221  4436162  2832947  4435261  4436155  4435551  4435253  4436167  4436191  4436179  4435263  4435549  4436192  4501452  4475385  2858123  2836141  2880077  2854994  2886274  2886275  2886278  4540178  4540181  2832946  4501493  4476720  2855032  2855020  2855021  2855010  2855015  2854989  2854990  2855009  2855031  2886272  2886262  4540185  4476747  2854975  4462689  4476736  4477106  4477117  4477136  4477150  4477114  4477113  4477129  4477135  4476738  4476737  4476740  4476723  4470236  2846367  2855013  2855022  2854984  2854996  2846359  2854973  2855030  2854967  2858078  2846357  4478260  2858100  2846348  2846380  2817370  2992249  2992266  2992209  2992238  4490614  4490603  2858105  4471193  4420029  4477124  4420527  4490610  4480797  4471204  4480808  4444182  4452551  4444208  4441654  4442589  4436233  2858091  4419702  2846387  2817708  2858108  2992297  2858099  2858112  2854970  2855025  2854992  2855016  2855019  2845608  4436148  4513309  4545154  4558997  4615436  4552826  2846354  2817710  4450845  2817707  4535400  4521995  4513332  2854993  4632928  4510869  4513549  4634337  2992222  4583297  2836152  4506018  4656218  4502695  4502707  2846369  2854965  4452574  4508979  2836142  2836147  4471257  4470362  4442582  4457522  4490586  4436240  2846370  2846381  2845618  2846337  2846351  4475384  2855029  2855017  2858120  2854971  2846376  4490585  2811924  4444199  2845626  4490608  4450704  4490584  3096799  3096810  3101791  3071063  3071054  3071051  3071050  3096796  3071041  2866158  4508973  2858085  2858124  2858102  4436235  3071053  3086947  2836150  3096804  2865790  3099854  3100111  3099153  3097816  3071042  3097830  3101810  3099150  3097812  3097808  3097826  3071040  3071064  3071046  3071044  3071052  4501473  3102330  2811927  3071056  3101806  3101803  3071045  3071059  3071058  3071065  3071061  3071048  3071047  3071057  3071062  3071066  2811906  4511625  4535416  2822831  2822824  2853268  2915024  2853270  2854980  4419705  2853266  2811923  4538405  4508008  4538414  4510022  2811917  2811902  2811913  2811914  2811900  2811911  2873273  2872073  2854982  2854963  2854997  2858103  4538411  3101789  2846349  2872089  2846382  2845600  2873269  2873276  4501484  4495377  4508024  4508021  4495350  4508006  4510882  4507976  4510851  4507979  4513330  4507986  4513548  4513543  4513320  4509999  4505326  4505302  4505341  4510864  2870635  4507971  3096809  3079373  3079372  3079734  2956554  3148604  3097827  2845605  3099158  2817374  3097831  3097834  3100108  3099849  3099852  3099851  3099170  3071060  3096802  3099146  3097814  3099169  3099151  3099141  3097832  3100092  3099163  2915708  2912138  2912144  2912120  2912164  2912166  2912157  2912106  2836160  2912177  2912152  2912139  2897790  2915714  2919974  2918054  2846379  2920005  2919989  2918063  2918053  2915724  2918051  2915704  2915005  2915006  2915014  2915003  2915703  2912174  2918052  4475388  2912168  2811919  2992224  2992235  2992203  2858117  4475372  2874093  4475366  4465803  4465868  2865789  4475389  4473880  2992273  2874104  2912173  2888743  2878494  2878513  2878529  2811922  2878522  2912175  2825186  2865770  2873277  2873272  2873275  2873274  2878526  4442549  4436229  4436151  4436238  2811908  4420045  4427307  4435226  4435537  4435635  4435547  4436182  4435235  4435623  2919960  4445643  3153673  3153672  3153668  3071049  4442573  4442586  4445629  4442563  4444160  4444188  4444170  4444218  4444201  4435546  4446193  4420506  4436236  4420021  4420024  4419721  4419722  4418148  4420496  4420525  4420495  4419720  4420489  4418139  4419690  4418153  4420490  4427310  4465782  4435262  4435617  4429261  4427063  4420499  4427291  4436177  4427311  4427075  4427071  4427051  4420521  4420035  4427294  2858097  4456634  2836149  2822825  2854987  4475397  4475373  4476741  2870640  2872079  2855035  4490594  4490590  2873281  4479005  4481082  4478253  2872097  4478237  2870620  2870625  4450703  4480803  4478227  4480788  4481077  4450856  2858110  2858092  2855008  4450731  3071043  4488988  4476732  2870639  2862253  2836164  2854995  2862256  2853260  2870630  2992243  2845622  2870617  2858086  2853273  2855023  4486618  4450728  4486407  4486631  4486627  4486412  2811905  2855018  2855037  4486641  4486633  4486606  4485699  2858096  2853259  2855024  4457503  2817706  2870632  2865779  2858087  2862249  2870637  4462733  2854964  2855006  4493149  4456616  4456611  2858075  4450841  4462671  2872074  2817354  4471198  4471200  4471194  4471253  2870638  2858077  4462670  2872105  2872096  4470369  4470371  4470242  4470366  4470201  2870626  2820569  4452605  2873279  4484519  2866154  4452587  2872082  4452571  2858116  4450854  2855011  4450853  4450849  4465875  4479017  2874110  4477137  4479041  4478261  4473236  4473278  4452607  4441637  4450842  4452557  4452615  2874091  2873280  4452606  2832941  2872072  3071055  2992311  3026694  3026692  3026693  3026698  3026697  4897827  4897791  4897792  4897793  4897817  4897828  4897830  4897836  4897849  4897842  4897856  4897786  4897782  4897840  4897848  4897778  4897816  4897795  4897814  4897818  4897780  4897789  4897807  4897802  4897825  4897851  4897794  4897784  4897812  4902280  4897838  4897783  4897788  4897846  4897841  4897801  4897799  4902278  4902287  4902293  4902286  4902283  4897806  4902271  4897833  4902268  4902269  4902267  4902264  4902261  4902277  4897796  4897790  4897779  4897798  4897821  4897823  4902262  4897847  4897819  4897839  4897854  4897850  4897826  4897781  4897815  4897811  4897853  4897832  4897845  4897803  3157023  3139865  3157043  4897835  4897824  4897855  4897797  4897822  4897852  4897808  4897787  4897805  4897820  4897831  4897829  4897844  4490617  2817351  4490618  4479033  4478999  4479000  2858126  4457499  4455675  4455663  4490620  2858083  4490612  2811915  4481081  4481078  4481079  2858098  2836156  2855003  2836153  2854969  2817349  4480824  4457520  4456613  2846378  4490588  2854988  2855007  2874106  4427296  2858074  2858104  2832945  4452612  4481073  2832953  2858127  2836146  2817362  2836143  2853267  4452621  4452549  2855001  4452622  2866156  4452573  4452610  4452594  2836158  2874113  2811932  2836163  2836151  4479035  4487212  4452593  2811918  4495364  4505986  4510872  4513555  4495374  4502702  4502689  4502678  4502683  2811903  4621339  4457507  2811898  2832935  2832956  2846345  2811907  2811931  2817360  2832944  2858113  2845624  4510879  4479042  2862271  2817361  2854978  4441630  2811901  2811921  2811926  2836159  4535415  4495383  4511623  4495367  4509991  4501486  4508962  4510002  4508959  4502670  4495387  4495353  4495385  2836165  4513539  4457494  2874094  2920010  4436187  2811920  2811929  2811904  2811899  2832933  2853264  2855004  4427046  2865775  4427293  2888749  2893439  2912161  2912147  2912176  2912101  2915707  2915736  2915720  2915012  2862261  4420031  2915043  2915709  2918047  2919998  2920013  2920020  2919986  2919954  2919973  4436231  4420028  2922005  4420502  4420498  4420512  4420034  4420038  4419704  4420505  4420511  4417852  4420528  2918062  4446210  2919963  4442591  4444172  4444165  4441624  4444173  4444217  4444184  4444162  4441674  4446202  4441669  4445619  4446208  4446194  4446196  4445649  4445657  4445618  4471192  2858081  4457514  4446199  4436164  2922006  4419701  4420522  4420520  4420507  4420039  4426781  4427263  4427285  4441642  4435530  4436150  4436243  4436228  4442588  4441650  4441635  4441623  4441643  4441646  4441658  4441653  4427049  4489633  4470220  4471201  4473888  4473891  4473912  2854977  2853265  4471227  4479044  2858089  4481084  2992246  2992315  2992272  2992201  2992306  2992308  2992229  4420043  2831558  2855000  4480821  2854976  4464361  4436161  2915027  4464365  4464347  4464352  4464353  4456608  4464350  4473886  4479001  2858094  2854966  2853261  2858107  2811930  2836157  4470368  4470254  4471218  4471247  4471250  4479015  2902530  2893457  2888736  2900971  2900972  2900982  2900970  2902526  2902527  2902529  2893436  2902542  2902523  2902538  2878490  2878483  2878484  2912142  2858109  2992204  2912150  2912162  2915047  2902536  2865769  2862269  2862273  2862268  2902545  2893452  2862272  2862250  2862252  2862255  2862265  2865765  2874105  2888742  2862246  2888744  2865771  2888738  2858106  2873271  2888754  2888755  2888741  4565687  4536306  4510892  4429280  4535418  4535409  4535410  4535407  4535412  4538403  4535402  4535419  4536298  4538396  4538418  4538404  4538401  4540165  4641824  4536289  4565688  4558992  4601958  4601961  4656229  4522009  4597327  4513542  4565695  4548060  4565699  4552814  4656237  4583295  4583282  4583292  4595434  4513305  4589114  4634763  4565698  4473876  4540180  4540179  4540169  4540162  4540182  4540192  4540189  4538409  4545159  4548056  4540175  4442618  4465786  4583308  4583311  4656243  4552842  4552809  4552811  4545156  4510885  4510021  4508964  4510339  4510349  4510005  4508993  4510016  4510350  4510359  4510038  4510357  4510037  4510029  4510986  4510853  4510875  4510865  4508012  4510982  4513317  4510340  4510351  4505979  4506001  4505993  4505983  4505999  4506027  4507980  4510020  4510352  4510354  4510345  4510362  4510035  4510344  4510013  4509994  4510004  4510019  4510355  4505309  4510356  4513315  4511618  4507992  4595445  4595432  4536299  4513540  4513556  4505311  4513321  4513554  4508003  4510883  4522005  4521998  4522010  4522008  4522013  4522011  4522014  4522853  4522006  4510891  4513546  4513328  4513312  4513313  4513301  4513302  4513306  4513329  4508004  4510032  4522850  4513303  4513314  4583312  4513318  4513333  4513307  4513552  4513297  4506016  4513545  4450723  4450725  4450717  4450687  2817353  2855036  4445607  2817357  4450707  4488961  4450682  4450684  4450693  4441675  4452561  2827251  4462703  4450733  4488978  4478255  4478223  4478230  4450691  4450859  4450843  4450695  4427286  4450683  2820563  4488970  4488964  4488990  4488981  4488977  4488980  4464625  4450715  4419689  4464595  4487205  2878504  2817368  2817373  2817356  2820570  4452554  4462665  4452585  4465781  4462730  4471241  4473905  4473883  4506014  4452604  2817367  4452562  2893455  4464583  4464585  2831561  2831563  2832943  2832937  4487213  2832913  4464612  4452586  4452611  4452613  4452579  4452550  4452547  4452546  2832938  4464346  4435555  4559444  4632935  4559440  4510867  4565693  2912159  4656222  4418149  4540201  4464618  4462686  2915731  2831544  2831547  4476739  4478257  2919977  4510026  4508025  4522851  4506021  4522852  4540199  4634758  4510358  4559442  4540171  2918055  4536292  4513304  4508992  4507978  4502716  4621326  4597335  4495382  4464603  4442608  4450689  4450696  4450840  4450702  4450848  4450714  4465818  4450686  4450712  4464349  4462704  4465798  4465779  2820566  2822830  4589113  4450858  4457516  2918066  4476724  4473907  2817372  4457513  4457492  4441648  4450722  4457512  2902546  2820567  4441647  4465873  4465864  4465807  4450679  4450699  4445612  4477116  2874107  2874109  4456633  4452591  2873270  2873278  2870614  2886276  2858119  2858080  2878507  2880153  2872085  2831541  3028384  2874096  2886279  2874085  2886263  4427052  4429288  4429297  4429295  2884442  2884444  4471260  4502674  2846384  4510363  4505340  4505294  4574105  4502667  4502677  4502686  4502673  4589116  4502675  4510353  4470199  4477110  2846339  4477147  4470224  4502668  4470237  4470370  2918068  4471251  4471220  4470229  2825185  4604141  4470376  2832934  4589120  4568198  4595439  4513319  4565692  4521990  4559441  4471221  4473878  4505324  4427279  4589112  4632933  4521999  4545145  4510003  4501492  4522000  4471224  4568193  4501495  4495384  4495354  4495356  4495378  4495375  4495381  4495361  4522007  4495362  4615393  4510360  4589130  4510036  4597319  4565689  4552807  4513326  4510887  4574107  4508020  4568209  4535403  4507975  4552823  4565694  4495376  4545161  4505984  4495379  4505321  4505338  4505342  4505313  4505314  4505319  4505303  4501444  4502712  4506015  4506004  4506005  4505982  4505997  4505990  4506028  4505298  4501467  4568199  4495358  4501468  4501479  4501441  4501460  4501465  4505300  4501480  4495380  4501472  4501490  4502666  4502669  4501475  4502688  4502709  4501442  4505316  4540172  4565685  4509997  4583313  4510854  4495370  4510033  4574108  4505318  4502720  4505330  4505295  4505296  4495366  4545155  4568208  4595433  4505308  4501453  4506026  4442565  4501458  4495348  4501466  4501483  4501449  4568201  4501454  4568207  4501447  4502693  4502690  4501476  4502694  4501471  4508972  4501478  4582971  4513325  4513298  4545149  4552837  4582987  4582975  4582963  4540177  4540195  4626455  4583306  4583307  4583294  4583283  4583318  4582973  4470206  4582970  4597334  4501489  4540194  4505317  4505322  4505315  4501482  4597326  4513299  4597333  4510849  4506006  4568200  4583315  4513553  4634761  4552836  4634750  4597332  4436178  4428625  4435541  4435542  4436149  4436169  4436158  4435209  4436194  4435618  2825191  2825195  2825192  2825183  2825184  2827",
      "more_code_info": "253  4436232  4435250  2915728  4427265  4429281  4429275  4429291  4429284  4435212  4435247  4445609  4435548  4435266  4435254  4435259  4435256  4435632  4435629  2923620  2915697  2915711  2918046  2918044  2919999  2920001  2832951  2920019  2915712  2922003  2922004  2922002  2923621  2923619  2923626  2920012  2893442  4429267  2878530  2878528  2878505  2880066  2888734  2915696  2893441  2915010  2902525  2912127  2912128  2912111  2915035  2915046  2845620  2893463  2919981  4429276  2920027  2919990  2919980  2920036  2919982  2920035  2919961  2920029  2919985  2920021  2920031  2919996  4473264  4420030  2919957  2919962  2915025  2915008  2915713  2915734  2915701  2915721  2920037  2919953  4419679  2919964  2919956  2919968  2920004  2919969  2919979  2918059  4427300  4426783  4427306  4427047  4427278  4427266  4427271  4420046  4427077  4419723  4427281  4427062  4428476  4428454  4427056  4429273  4427272  4419696  4420019  4419687  4419708  4419717  4419725  4419698  4419685  4419714  4420036  4419681  4419680  4418143  4419694  4419697  4419715  4420040  4419713  4444180  2923623  4444209  4444210  4444192  4444193  4444167  4441651  4442604  4442584  4444211  4444202  4444205  4444216  4444166  4444181  4444183  4442596  4441622  4441664  4442603  4441634  4441641  4442553  4444214  4441626  4442572  4441671  4442597  4441668  4441640  4442617  4441659  4435627  4470367  4445627  4445640  4445633  4445662  4445648  2992283  4442611  4478240  4442568  4471228  4471232  4481068  2992264  4505332  4470210  4465825  4445628  4442600  4442570  4442614  4445620  4445610  4445621  4445617  4445639  4445631  4445624  4445645  4445650  4445608  4445659  4445663  4442593  4445636  4427273  4417855  4426782  4427058  4427059  4427284  4426779  4428469  4427267  4427301  4427302  4427274  4427269  4427280  4427277  4441656  4427282  2846352  4420022  4420020  4419727  4419707  4419693  4419683  4427308  2855012  4427309  2915716  2886264  4427275  4427287  4427069  4427057  4428624  4420047  4436196  4435208  4435257  4452577  4436190  4436154  4436239  4428467  4436171  4435619  4435552  3079369  4441649  4441662  4441678  2927659  4436237  4435540  4427289  4428474  4429289  4429268  4429282  4435550  4435217  4435536  4435222  4435531  4435249  4435265  4435255  4435232  4435534  4441670  4435543  4473273  4471233  4473875  4473898  4473238  4475391  4475363  4473893  4470213  4465819  4475361  4464616  4464599  4462674  4464602  4464614  4475398  2878517  4490600  4471256  4471244  4471259  4471234  4471258  4473267  4473904  4464357  2992261  4465806  4465805  2992335  4477134  4465865  4475367  2992237  3079371  4471239  4450708  2825189  4476746  4477102  4464631  4465794  4475390  4477103  4478251  4465814  4465813  4465810  4465822  2831559  3028380  4470225  4473280  4465874  4475376  4470232  4475396  4435219  4470253  4462719  2915000  2992325  4473913  4473263  4477123  4473900  4464627  4473894  4452601  4471252  4470209  4473254  4462718  4473275  4473239  4473258  3026696  4473265  4473908  4462685  4462702  4457504  2992199  2992300  4470252  4473909  4473248  4473902  2915023  4473249  4470246  4473922  4473283  4470231  4457506  4452618  2831550  4464632  4464613  4464375  4464611  4473924  4442592  2846356  2846368  4456623  4455681  4457501  4457518  4476728  4457500  4435230  4478241  4471229  4464587  2845613  2846353  2831556  4457496  4462698  4475370  4457490  4457515  4457521  4452584  4457523  2845628  4464383  4475399  2817364  2831560  2831564  2846338  4456612  4429258  4457511  4457495  2878511  2855005  2855002  2854974  2855033  2878492  2878503  2894442  2880067  2836154  2888730  2893450  2893429  2893448  2831545  2817703  2880141  2846374  2822832  2825197  2827252  2827254  2817363  2831555  2846364  2845607  2846385  2846386  2845612  2846343  2846340  2845610  2845602  2894443  2832942  2915715  2912143  2912132  2915002  2915030  2915031  2915041  2894440  2915036  2912119  2915717  4475364  4456621  2915737  2915022  2915735  2915045  2900973  2902531  2902532  2902547  2902533  2900983  2900975  2912134  2900981  2912155  2912109  2912116  2912118  2912146  2912154  2912126  2893434  2900980  4475383  4452596  4473920  4470248  4477152  4473274  4475394  4470218  4477131  4471255  4475369  4477118  4477121  4477107  4477148  4478228  4478231  4476749  4473242  4490581  4478249  4477120  4475381  2893443  4442587  4470233  4473918  4473879  4473247  4473910  4473903  4473276  4473272  4473246  4464341  2992287  4484522  2992313  2992314  2992317  4470250  4473885  4473882  2992290  4485708  3025486  2992239  3025484  2992274  4473919  2992226  4462667  2992293  4471668  4478245  4488986  2992240  4488987  4490615  2992267  2992251  2992319  2992288  2992292  2992258  2992260  2992269  2992312  2992242  3096405  4435544  4455670  2915009  3101802  3096801  3102326  3102318  3102323  4481065  4481075  4481072  4490604  4462695  2878523  4480804  4481060  4462660  4462711  4462708  4462663  4464381  4487209  4462717  4465817  4480829  4450697  4489634  4487211  4490605  4465811  4465777  4464629  4450701  4465867  4481070  4450705  4450850  4450847  4450727  4450677  4450711  4477130  4462687  2992253  4490583  4476726  4479045  4493135  4489638  4490616  4490602  4490596  4493156  2992282  4473897  2992323  4475393  2992328  2992234  2992213  2992321  2992330  2992277  2992207  2992211  4470378  4502691  2992200  4480835  4470239  4470214  4478998  4475386  4470223  4470238  4478238  4475382  4450846  4479009  4475387  4478234  4470360  4465870  4465787  4465871  2872100  4476745  4471205  4475365  2992310  4475392  4475400  4478252  4510888  4521994  4505991  4507989  4507999  4507982  4507984  4507988  4507972  4508023  4507995  4502714  4510855  4501443  4510348  4510014  4510025  4510983  4510850  4510008  4511627  4510856  4521991  4501446  4508986  4501474  4510343  4510006  4510009  4510010  4513547  4510347  4501481  4568197  4493159  4501457  4502698  4501488  4522003  4501469  4501451  4501448  4502682  4501445  4501461  4495373  4436234  4501491  2992245  4501496  4486636  4521993  4481066  4480814  4480801  4480789  2886280  4493141  4493144  4493143  4493140  4568204  4486634  4565691  4485709  4486617  4485692  4486612  4490609  4490582  4490611  4444163  4445626  2872099  4493158  4538410  4522004  4521996  4535404  4535406  4536307  4536304  4536288  4536285  4536291  4536294  4480823  4536296  4476744  4540163  4536302  4552816  4552822  4552812  4536303  4558994  4565690  4501440  4540164  4536297  4436188  4441665  4436152  4435553  4435211  4436174  4436227  4436189  4436160  4435527  4436245  4435227  4436170  4435233  4441632  4441621  4441660  4441663  4442566  4442607  4442590  4442559  4442564  4442558  4436246  4444200  4429254  4442562  4442595  4442601  4444203  4442577  4441627  4441619  4441645  4441666  4436242  4442576  4441628  2862251  2878520  4420037  4420514  4426780  4427068  4428449  4429262  4435539  4435248  4441661  4446192  4442580  4445658  4445655  4445652  4445632  4446205  4446198  4445635  4445654  4445665  4445606  4446191  4445641  4446209  4445623  4441672  4442561  4444185  4446204  4445611  4442605  2992278  2992254  4446207  4445664  4441629  4441625  4441620  4442606  4441639  4442578  4442585  4442610  4442550  4444197  4445616  4442567  4441638  4445653  4442612  4446203  4446195  4446201  4445661  4436193  4435535  4435236  4445651  4442551  2915007  4420516  2912172  2912098  2912107  2912113  2912151  2915004  2915001  2915021  2915034  2902549  2915020  2903438  2918043  2918058  2918061  2918050  2918048  2919988  2927667  4419688  4419691  4442579  2915019  2872094  2992296  2992220  2992217  4464377  2992324  4462694  2862257  2865791  2862254  2872081  2894472  2870633  4420500  2872088  2872087  2865787  2874086  2874095  2874084  2886281  2886258  2902541  2903441  2872095  4441636  4419724  4435246  4442554  4441631  4444161  4444213  4444215  4444194  4444171  4444212  4429285  4442555  4429256  4444190  4444195  4444168  4444175  4444176  4444204  4444207  4444177  4444174  4444179  4442599  4427045  4420524  4420492  4420493  2929716  4420042  4420491  4420517  4419719  4427283  4427054  4513322  4427055  2992252  4427072  4427264  4427304  4427290  4427276  4427076  4427044  4427297  4427303  4427299  4427288  4436230  4513311  4510979  4565697  4508019  4510361  4510890  4509989  4480787  2872093  4568195  4502713  4510341  4510984  4480795  4510861  2884443  4510991  4511622  4511626  4511621  4513316  4513335  4634330  4513310  4513327  4513334  4513557  4513558  4513551  4510981  4565696  2886268  3149125  3143261  3147601  3148274  3147599  3153667  3139851  3157047  3149121  3157036  3157015  3141738  3146350  3146612  3147598  3139834  3139869  3157044  3157022  3156432  3149120  3157048  3139845  3157028  3139867  3139881  3156428  3147597  3156429  3141708  3124389  3154504  3154502  3157030  3154503  3157031  3157054  3141728  3153670  3148620  3157032  3146614  3131160  3157016  3157050  3157026  3157042  3146336  3157020  3154509  3154507  3157017  3148289  3154499  3148277  3149126  3157038  4583309  3139827  3104691  3148617  3157037  3157052  3157033  3141740  3143259  3157041  3143492  3148616  3157018  3157025  3157040  3154510  3154506  3154500  3143267  3157045  3150097  3154498  3116261  3089718  3150096  3148272  3146339  3148276  3143260  3148601  3149119  3149127  3157051  3149111  3139828  3139873  3157034  3157024  3136853  4583317  4621318  4666259  4583302  4634771  4668077  4638505  4621317  4668076  4668067  4638491  4638481  4621319  4638478  4641821  4621327  4636846  4659170  4548054  4583291  4636850  4638508  4589122  4583301  4615397  4582976  4638510  4621325  4661234  4634344  4668071  4666242  4641841  4638499  4615424  4615411  4668064  4634796  4621341  4661239  4632930  4638509  4641848  4621336  4668081  4621348  4615429  4615404  4661237  4552848  4583303  4582985  4583285  4615401  4601962  4540170  4599667  4641818  4615437  4501487  4634762  4501497  4595440  4641827  4638479  4536293  4634349  4506023  4641819  4510989  4621337  4656221  4582990  4582986  4621338  4638516  4641851  4641832  4634756  4636841  4552819  4634764  4636844  4668080  4558993  4599674  4599672  4634775  4583286  4661232  4656233  4621332  4621345  4661225  4638514  4659167  4641825  4615439  4668075  4615416  4615423  4545152  4583304  4601905  4638484  4659161  4668087  4656236  4621335  4615428  4595429  3105869  4634770  4634794  4636852  3141726  4615399  3141741  4661230  4632921  4632923  4632922  4632931  4632938  4634755  4659174  4540186  4634333  4634759  4661240  4626456  4450700  4545160  4636847  3089902  3099152  3100101  3146615  3086958  3120305  4634779  4552818  4634766  4661226  4545144  4634781  4634354  4659175  4621320  4668074  4638487  4641842  4659168  4656231  4656228  4552828  4659165  4632925  4668069  4634361  4668082  4668084  4510023  4604149  4666253  4634366  4666257  4636853  4634350  4638474  4659169  4522012  4666239  4634788  4659173  4666834  4601901  4501494  4641850  4659171  4634792  4597316  4666252  4636845  4540188  4666835  4666838  4638489  4621322  4661229  4656235  4641839  4634784  4656219  4511620  4615394  4668072  4634778  4538416  4656224  4656226  4638488  4661231  4634790  4632927  4634363  4634359  4634346  4666839  4636848  4666240  4666246  4583300  4661228  4595436  4641831  4641828  4668086  4558995  4545150  4582964  4666256  4552832  4666247  4656241  4597320  4601902  4668078  4510018  4582972  4559000  4540193  4636851  4634336  4582961  4634757  4638507  4666833  4638504  3120785  4462723  3139864  3146352  3146613  3136836  3105809  3157035  4465785  3154508  3147616  3090073  3130804  4455679  3118367  3090067  3102333  4638483  4465793  4465872  3086955  3086954  3146617  3124407  3146616  4465783  3139856  4456627  3089900  4465801  3131173  2960116  4621330  4656234  4656215  4638515  4668083  4632936  2919976  4634780  4666249  4638512  4632939  4632929  4668070  4641846  4510858  4634348  3104678  4464607  3099157  3101800  3104688  3104699  3120779  4634753  3086968  4666255  4464360  4464355  2956557  4636849  4601964  3120299  3128369  3105862  4465784  3131159  3101799  3131158  3101804  3089408  3116274  3141709  3086978  4464626  3111178  3120295  3096398  3097824  3105858  3116271  3096411  3101787  3139843  3146351  4481071  3102319  3099856  3090069  3116247  3141730  3111180  3141739  3141694  3141743  3118371  3139857  3104692  3116264  3139836  3139833  3125615  3139872  3105868  3131161  3147600  4462721  4462690  3115077  4464356  4464588  3089402  3115086  4634787  3099156  3089416  3126482  3100090  3105863  3102313  3148271  3124391  3115084  3141736  3096419  3086966  3148280  3113558  3125624  4632940  4638500  4638502  4656240  4656216  4656220  4638496  4634343  4582982  4508005  4601966  4634774  4582969  4634765  4659172  4656238  4661235  4552835  4599668  4641817  4641837  4641820  4599675  4666238  4510880  4656230  4641826  4666832  4666237  4621329  4659164  4621328  3141712  4641843  4552847  3118377  3096800  3104689  4634798  3104700  4582984  4634332  4615392  4641822  4604142  4668065  4615421  3089405  4548061  4582988  4638517  4668068  4615406  4666260  4641835  4641830  4638493  4638495  4582966  4510863  4582978  4597321  4634751  4666245  4634795  4638477  4659163  4636842  4615426  4666837  4634353  4666244  4641823  4661233  4634356  4668079  4659166  4582980  4583314  4589127  4668073  4661227  4632937  4597328  4668066  4661241  4661238  4615434  4641840  4632924  4632932  4634767  4615438  4641833  4634329  4634772  4634335  4538402  4656225  4638480  4536308  4545151  4666248  4634749  4634797  4634347  4666836  4641834  4615419  4634760  4638476  4659162  4638490  4666254  4661236  4668088  4583296  4634341  4641829  4641845  4634345  4666235  4634362  4641836  4638513  4545157  4615402  4634339  4656242  4641849  3139829  3126484  3141735  3086944  3120310  3104693  3116255  3139854  3139838  3118388  3118372  3128452  3141742  3141686  3141731  3139862  3116262  3148287  3148283  3089411  3101790  3131163  3128445  3120306  3089719  3097797  3119353  3119359  3096807  3097828  3120778  3097802  3141687  3125619  3124412  3104686  3150102  3143266  3141688  3115074  3102335  3131165  3128366  3089908  3141733  4638492  4656227  3131178  3124403  3148602  3122112  3139852  3115087  3147609  3130796  3101797  3090051  3090047  3124410  3146349  3131174  3116275  3116272  3099162  3102320  3096412  3089710  3102329  3113562  3146609  3111171  3101795  3096400  3118380  3090057  3089918  3089905  3101792  3089714  3090090  3126481  4668085  4634768  4656244  3147604  4615400  4656232  3102336  4634352  3089400  4632926  3105866  3096402  3097833  3090080  3111179  4634785  3111170  3116269  3115080  3120777  3122117  3136844  3097821  3157021  3116270  3115072  3147620  3141737  3090072  3090048  3141725  3113564  3090082  3143262  4483972  3101811  3136850  3089399  3118379  3090049  3096414  3101793  3157046  3141732  3118385  3118382  3139842  3116263  3118363  3090044  3118390  4483958  2960121  2960108  2959507  2959502  2952262  3116260  3099161  2929722  2929719  3143493  2919992  3154501  3128450  3148613  3104690  2912103  3116248  2920006  3086961  3143269  3148606  3141705  3148275  3105870  3089903  3086965  3089917  3089909  3128444  3118364  3115078  3141706  3150100  3118369  3118375  3090050  3090071  3100089  3089723  3148290  3153671  3089404  3143490  3099142  4464623  4479103  3102327  3141707  3141744  3116253  3116251  3097801  4484532  3097800  3090083  3086981  3086957  3105859  3086983  3090068  3090052  3104696  3146342  3100109  3086956  3148605  3086945  3136845  3086971  3116258  3124409  3131170  3146353  3115069  3113560  3099159  3086980  3124406  4486593  4486610  4480791  3136842  3131179  3100098  3153669  3089414  3136851  3116266  3139860  4477128  3090077  3090079  3116259  3100094  3115085  4465802  3115079  3118378  3141713  4484530  3089912  3104683  3099148  3102332  3096417  3146354  3139875  3089914  3097817  3090061  3096422  3124404  3118381  3089910  3090066  3149123  3115091  3102328  3143489  4477139  3099160  3090053  3154494  3101798  3089709  3104685  3101808  3139878  3097806  3147606  3147608  3131177  3115089  3096808  3100097  3102316  3101809  3086959  3122113  3147610  3113566  3105864  3115081  3143273  3131169  3118370  3132253  3097805  3086963  3147607  3147617  3115075  3116257  3147612  3113565  3115071  3097796  3118384  3143268  3100107  3099848  3086969  3118386  3139840  3131157  3099145  3097822  3097825  3097818  3099167  3136846  3146347  3097813  3122115  3113555  3113556  3100110  3136847  3089711  3147602  3113567  3118387  3089403  3100099  3118391  3118374  3130802  3097819  3150098  3157027  3111174  3116268  3141696  3086976  3102321  3100093  3102315  3089904  3089901  3090089  4484536  3147611  3148286  3125616  3115070  3096404  3118376  3099165  3100095  3148611  3096408  3090070  3118365  3089707  3122110  3143263  3115083  3118373  3025487  3148608  3120301  3099166  3124392  3101796  3139841  3125623  3146346  3090062  3115076  3136852  3119356  3097799  3089398  3118366  4485701  3087757  3119357  3141704  3146341  3139849  3139879  3124411  3120308  3139877  3146610  3119360  3139846  3100096  3086977  3104695  3116252  3139871  3099853  3111182  3136848  3105860  3099144  3099154  3115082  3096407  3120783  3139858  3128449  3141729  3141699  3141689  4483696  3096413  3157049  3089713  3139850  3148612  3119355  3148285  3130797  3105810  3148610  3148619  3120311  3086979  3086970  3125622  3086975  3090088  3143270  3146340  3148614  3147614  3141734  4484244  4471246  3154497  3086948  4480807  3149122  3113559  3149117  4483959  3079374  3139839  3089721  3086946  3096401  3111173  3139874  3105808  3128451  3120782  3089899  3097807  3141685  3116250  4483966  3141700  3139861  3086962  3139844  3104682  3096427  3101805  3025485  3100087  3148284  3149128  4462707  3097811  3104679  3136840  3136837  3105865  3086953  3090042  3090043  3120303  3097809  3099155  3149113  3101801  3148288  3111177  4456619  3104680  3148618  3139847  3089708  3089715  3115088  3089406  3149114  3126486  3104697  3099850  3089705  3146343  3089409  3141701  3086960  3089410  3089913  3089717  3086967  3146345  3089401  3090054  3097804  3115073  3105853  3104687  3096797  3131164  3090085  3087758  3111176  3099149  3099147  3086973  3087755  3086982  4483960  3100102  3086951  3105852  3090055  3118383  3102334  3099143  4480817  3102322  3116276  3154512  3105806  3139837  3139853  3097823  4486609  3125620  3105854  3139848  3139830  3124408  3149124  3097820  3105851  3139832  3119362  3096421  3141698  3096410  3089915  3119354  3105807  4484538  3154495  4481074  4477149  4483982  3146348  4480827  4483961  4481059  4484242  3146618  4486408  4486623  4484241  3096805  4483697  4478235  3089720  3099847  3089706  2912131  4487210  4480822  3090045  3099164  4478254  4478229  4479054  4481069  4481058  4481064  3090064  3104698  3089716  3141697  3105856  3104694  3146337  3090058  4464339  3090081  3104702  4464344  3089916  3131176  3150099  3139831  4464596  3116256  4486417  3130803  3146344  3089415  2960122  3096426  3141703  3089911  3089412  3096409  3090091  3089407  3090059  4481061  4483971  4480799  3089704  3101807  3089906  4487208  3086949  4481057  4470244  3100105  3131168  3105855  4480802  4483694  3141692  3141702  3096406  3130801  3118362  3131166  3090076  3105857  3089417  3105861  3111183  3141711  3100100  3130798  3130800  3130799  3139868  4471222  3097803  3099168  3149109  3139826  3147618  3146611  3147619  3131172  3120302  3125617  3120786  3119358  3120780  3148603  3118368  3126483  3150101  3124390  3141695  3141714  3115090  3119361  3147621  3143272  3148609  3143271  3100103  3149118  3139866  3120313  3147605  3143265  3122111  3147615  3143491  3096416  3102314  3116273  3139876  3128448  3089413  3122114  3105867  3131156  3126485  3120298  3120307  3141693  3139859  3141690  3148279  3148273  3148278  3139870  3139880  3087759  3147603  3149110  3149115  3131175  3124388  3128367  3128368  3147613  3102325  3154496  3136843  3090074  3090063  3096425  3090084  3102317  3101794  3086974  3096803  3125625  3128453  3139863  3090078  3096798  3096806  3097829  3086950  3089722  3102324  3090046  3096424  3096418  3090086  3090065  3111175  3097798  3090087  3104681  3124405  3148607  3096403  3086964  3087756  3120784  3089712  3139825  3111172  3125621  3118389  3141691  3148615  3090060  3097810  3120297  3090075  3096423  3096420  3096415  3101788  3157029  3079370  3136839  3100106  3096399  3154511  3157053  3146335  3146338  3148281  3149116  3111181  3122119  3143264  3141710  3097835  3113563  3154505  3136838  3128447  3139855  3139835  3113561  3125614  3131167  3113557  3136849  3100104  3131162  3131171  3116254  3102312  3148282  3136841  3141727  3116265  3122116  3104684  3104701  3116249  3122118  3102331  4479010  4540166  4470216  4473271  2829480  2888740  4462668  4462706  4462699  4462734  4462682  4462680  4568206  4480828  2880082  4535408  4508977  4508985  4505299  4505304  4505337  4495388  4615418  4480809  2894469  2832940  4479102  4429271  4479104  4538419  2854999  2992332  2992309  2992301  4493148  4486409  2992329  2919995  4475377  4475362  4897800  4902295  3086952  4902276  4902272  4902273  4897834  4902274  4902288  4902292  4902290  4902275  4902270  4902294  4902289  4902297  4897857  4902266  4902285  4902281  4902282  4902284  4902279  4902265  4902263  3157039  3157019  4902296  4897843  4902291  4897810  4897809  4897804  4897785"
    }
  ]
}