  "results": [
    {
      "status": "Terminated",
      "city": "Sylmar",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "78093",
      "recalling_firm": "St Jude Medical Inc.",
      "address_1": "15900 Valley View Ct",
      "address_2": "",
      "postal_code": "91342-3577",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution",
      "recall_number": "Z-0034-2018",
      "product_description": "PM1214 NUANCE SR RF; PM1230 and PM2230 NUANCE MRI; PM2214 NUANCE DR RF  These low voltage (LV) devices are implantable pacemaker pulse generators, intended to be permanently implanted in the body, that have a power supply and electronic circuits that produce a periodic electrical pulse to stimulate the heart.  These devices are used as a substitute for the heart's intrinsic pacing system to correct both intermittent and continuous cardiac rhythm disorders.",
      "product_quantity": "2741 units",
      "reason_for_recall": "New pacemaker firmware was developed to further mitigate the risk of unauthorized access to our pacemakers that utilize radio frequency (RF) communications.",
      "recall_initiation_date": "20170828",
      "center_classification_date": "20171019",
      "termination_date": "20200708",
      "report_date": "20171025",
      "code_info": "Serial Numbers:  2678856  2678867  2678859  2678857  2678877  2678872  2678870  2678869  2683929  2678863  2678858  2678866  2678878  2678868  2678860  2678874  2678879  2678864  2683921  2673458  2683918  2683917  2683914  2683928  2683927  2678862  2683922  2678880  2683920  2683930  2683911  2683916  2675733  2678861  2678871  2683923  2673452  2678876  2675748  2675747  2675742  2675737  2675731  2675743  2673467  2675739  2673449  2673448  2673457  2673459  2673469  2673468  2673466  2673465  2675727  2683924  2678865  2675734  2675728  2675725  2675736  2675745  2675740  2678875  2675732  2675730  2675749  2675746  2675738  2675729  2675735  2675726  2692492  2689903  2691326  2691327  2691328  2691334  2691330  2683919  2689905  2691332  2692495  2692497  2692483  2692488  2692489  2692490  2691329  2689901  2678873  2689899  2689908  2689906  2689907  2689897  2691325  2689900  2691333  2689898  2689909  2689904  2689902  2689896  2691331  2692481  2689910  2686717  2692491  2686730  2686728  2686722  2686721  2686726  2686712  2686718  2686723  2686720  2686715  2683913  2683915  2683926  2683912  2683925  2686725  2692485  2675741  2692482  2692499  2692500  2692493  2692494  2686714  2692484  2692487  2686729  2686719  2686716  2686711  2686713  2686727  2686724  2692496  2697416  2697406  2697407  2697411  2697415  2697401  2697403  2697408  2697405  2697413  2697414  2697412  2697409  2697410  2697419  2697400  2613924  2697404  2613921  2613928  2617417  2613927  2613937  2613925  2613943  2697417  2613926  2699066  2613920  2613919  2673464  2613917  2675744  2692486  2692498  2613923  2699063  2697402  2700386  2700385  2699072  2700388  2699069  2700394  2699065  2700389  2699060  2699059  2699058  2699055  2697418  2699057  2699061  2699073  2700382  2699062  2699056  2699067  2699071  2699068  2699074  2700392  2699070  2613918  2700384  2700387  2700393  2700390  2700381  2700380  2700383  2699064  2646730  2613916  2646718  2613930  2646736  2646735  2646734  2645827  2646731  2646728  2646727  2646726  2646725  2646723  2646722  2646721  2646732  2673454  2673463  2673462  2673461  2673460  2673456  2673447  2646719  2673455  2646733  2673453  2673451  2673450  2673445  2646724  2646729  2646737  2673446  2613938  2613929  2613944  2617416  2613932  2613945  2613941  2645808  2613939  2613936  2645826  2613935  2646720  2613933  2613922  2613931  2613940  2645822  2613934  2645809  2645825  2645823  2645821  2645820  2645819  2645818  2645817  2645812  2645810  2645824  2645811  2645816  2645814  2645815  2645813  3004783  3004782  3004780  3004779  3003113  3004781  3003111  3057612  3061676  3003112  3003110  3003109  3057616  3065294  3065289  3061677  3065287  3061675  3061678  3057614  3065295  2847869  3057615  3057613  3061674  2847868  2971148  2958000  3065291  2971146  2971149  2996824  2976064  3096830  3096828  3096826  3096827  2847872  2847875  3096829  2847870  2971153  2847874  3085187  3084071  3084069  3085185  3084070  3090092  3090093  3084068  2971151  3084072  3083191  2958004  2957999  2958002  2957997  2957995  2958003  3095477  3095475  3095479  3095476  3095478  2957998  3083189  2957996  3083192  3083193  3065292  3065290  3065286  3065293  3085183  3085186  3090497  2971150  2971145  3090496  2941096  2941098  2941088  2941086  2941087  2941091  2941092  2941093  2941099  2941097  2941090  2941100  2958001  2941095  3085184  2941089  2943232  2943233  2943234  2943235  2943231  2948921  2948922  2948923  2948919  2948920  2941094  2997964  2835430  2835422  2994720  2994721  2997963  2994718  2835428  2835425  2997967  2997966  2996828  2835424  2835429  2994717  2994719  2720440  2716838  2996826  2276862  2971154  3002165  3002164  3002162  3002169  2997965  2716833  2977854  2977856  2725067  2977857  2741101  2741098  2741104  2741099  2741102  2741097  2741095  2835427  2716831  2716835  2716836  2741096  2835426  3001457  2977853  2977855  2725064  2725071  2834077  2835423  2996827  2741103  2983123  3083190  2966233  2966234  2966232  2966236  2966235  2725069  2725066  2725068  2725073  2725072  2716834  2983125  2716830  2983121  2983122  2976062  2976066  2976065  2716837  2976063  2999362  2999363  2999364  2999361  2983124  3002166  2720442  2964408  2964409  2964407  2964410  2964411  2996825  3083194  3083188  3001456  2720444  3002161  2999365  3001455  3001458  3001453  3001454  3001452  3002160  3001451  3002163  3001450  2725070  2847876  2716832  3001449  2836952  2837612  4522087  4522221  4522088  4510898  2836968  4508646  2836962  2837504  4522100  2837579  4510896  4522219  2837571  2837495  4522226  2861264  2836927  2861260  4522222  4522105  4508411  4522090  2836933  4522220  4511236  2836930  4522102  4522208  4510903  4522213  4522107  2861257  2837559  4522212  4522225  4522215  2836931  4511230  4510913  2858287  4462963  4510907  4508406  4510895  2837488  4508648  4511239  2858285  4484561  2837610  2845407  2837465  4496338  4508428  4511235  2837570  4508405  4508417  4508403  2837455  2837596  4522218  4510908  2837514  2837473  4508425  4508414  4495976  2858283  2837597  2837578  2858288  4508409  4495974  4508426  4511243  4508412  4495966  4522205  4492542  4522206  2837547  4511244  4511237  4496339  4496332  2837588  2837587  4496327  4522227  4522103  4522098  2837536  2837512  4511240  4522104  4510920  4496326  4496323  2836926  4495968  2836937  4510919  2837592  4496328  4496159  4522210  4522106  2837440  4496325  4510925  2837532  4522094  4522096  4510900  4496157  4496333  4522207  4490307  4522204  4522095  2836936  2837628  2837546  4522108  4492031  2837494  4492043  2836942  2836954  2837464  2837468  2837479  2837591  4491940  4492520  4479310  4491927  2837601  2837540  4522089  2837616  2837433  4510917  2837503  2837541  4522099  2837530  2836929  4522224  2837623  2836965  2837449  2836932  4492045  4492498  4491924  2837584  2836963  4496334  2836958  2837641  2837627  4505397  4472648  2837566  4505400  2837534  2878387  2836935  2837550  2837622  2837625  2837595  2836934  2837524  2837492  2836948  4505396  2837533  2837619  4491936  4508640  2837452  4517463  2837500  4505417  4508642  4478902  2837618  2836949  2836956  4508415  4508643  2836946  2837520  2878386  2837630  4508418  2837477  2849540  2849543  4505406  2878384  2878385  2849529  2849536  4508407  2837466  4508641  4482440  2851868  2851863  2849544  4511245  4519381  2849528  4508647  4478908  2837493  2837544  2837632  2837472  2878388  2837478  2837480  2849534  2836940  2849537  2837526  2837565  2849527  2849535  2837548  2836967  2837448  2837444  2837557  2837450  2849538  2837511  2837436  2841531  2841585  4508416  4508402  2841521  4482438  2837521  2837491  2841526  2841586  2841557  2841554  2841549  2841574  2837457  4482421  4508400  4508423  4482443  4508408  2837432  2836947  2837575  2837604  2837638  4508413  4482437  4508419  4508421  4508399  2837499  2837589  4517456  2837542  2837498  4519387  4522091  4482441  2836970  4482436  4482445  2836953  2837569  2837516  4508420  4468387  2836959  4468393  4468374  2847146  2836944  2837439  4522216  4522223  2837568  2837581  2837490  4505395  4482428  2837522  2837607  2837551  2841545  2841592  2841566  4511249  2837582  2837475  4508424  4482435  4508427  4510901  2837486  2837434  2841578  4482434  2841539  4512791  2837633  2837634  4512815  4513021  4513023  4512827  4512833  4512818  2837617  4512809  4512798  4513024  4513028  4512797  2837580  2837553  2854150  2847156  2837447  2837576  4515199  2837483  2837640  2837563  2837620  2837599  2837598  2837510  2837611  2837626  2837572  4512806  4512828  2837585  2837460  2837602  4512810  4512814  2837519  4487864  2837637  4511247  4511241  2841563  2841594  4511229  4510899  4495970  2861266  4510902  4510906  4510923  2841561  2841540  2861259  2841583  2841542  2841533  4495969  2837639  2854147  2851856  2836950  2837527  2837459  2837609  2854144  4495973  2837615  2847160  4473448  2854148  2854149  4472688  4472667  4472665  4470315  4511250  2854145  4517458  4517459  4519375  4518031  4518028  4517467  4519374  4519385  4519369  4517466  4515188  4522888  4517243  4517244  4517450  4517461  4517452  4518032  4505408  4462953  4517468  2837496  4471057  4514958  4515206  4518026  4518025  4517250  4519370  4518029  4514948  4517247  4517249  4517457  4517465  4519372  4519380  4519377  4517453  4518027  4505418  4517449  4472687  2854146  4484556  4484094  2847161  4484081  4471046  4505398  4484562  2837513  2837577  2837586  2837624  2837636  4505412  4505411  2837456  2858284  4512808  4517455  4517451  4517248  4518033  4518034  4517462  4518030  4472646  4517246  4487865  4519371  4519373  4519384  4519378  4519388  4519379  4519382  4519376  4515187  4484546  2847168  2847155  2847172  2837538  2837518  2837446  4511248  2837543  4496321  2847164  4471186  4484579  4492531  4496337  2837470  2837539  2837560  4510921  4511238  4496340  4492515  4510904  2837603  4510926  4492525  4492026  4492496  2847170  4492534  2858418  4510927  4511251  2847149  2847143  2837621  2837528  4510909  4484077  4492049  4484290  2847147  4484288  4484293  4484082  4484574  4496324  4511242  4511232  2837458  4510928  4484552  4510915  2845405  4511231  4522097  4522211  2836928  4495965  4510916  4519383  2858421  2847152  2847144  2837501  4492524  4496329  4484080  4505409  4492506  4492523  4511246  4496160  4484095  4484091  4484573  4496322  4505405  4510910  4484542  2841523  4492038  2841576  2841546  4471181  4471185  4471183  2837605  2841519  2847148  4510894  2841517  4470314  4471178  2841559  2841589  2841580  2841569  2841525  2841556  2841584  4496330  4510924  2837590  2841520  2841575  2841544  2861261  2841562  2861256  2841534  2841535  2841538  4508401  4484086  2841590  2841587  2841564  2841524  2861254  4492512  4492527  4492039  2841572  2856136  2856137  2841530  4510914  2841555  4496331  4492505  4492047  4492029  4492027  4510911  4492522  4492533  4492511  4508645  4492540  2841573  4492030  2841536  2841541  4492516  4492519  4492494  4491918  2841522  4492526  2841567  4492510  2841591  4491698  4492530  4492536  4492539  4492048  4510922  2841595  2841547  2896556  4471055  4480063  4471188  4472681  2896567  2896564  4472666  2896568  4472682  4472671  4471053  4472684  4472672  2896569  2896566  2904436  2904444  4472656  4480059  2904433  2904443  2904441  2898930  4480055  2896561  4592297  4592462  4480050  4480052  4480058  2902713  4462968  4462946  4462950  4472664  4472645  4472651  4472675  4462947  4462980  4462973  4462972  2904434  2905739  4462971  4462966  4472674  4462957  4480047  4480051  4480049  4462942  4462949  4462956  4462961  4462983  4462979  4462982  4462969  4462975  4462985  4462981  4480057  4462987  4484283  2905724  2902714  2905738  2908587  2905748  4479301  2905718  4484083  2905732  4484287  2905742  2905723  2905707  2905717  2905747  4484554  2908583  4480064  4462964  4472662  4471177  4470350  4472640  2905710  2905745  2905727  2905741  2905731  2905744  2908588  2908585  2908584  4592299  4484281  2905743  2905721  2905720  4484285  4484555  4484577  2905737  4484543  2905712  2904439  2904438  2904447  2904435  2904446  2904445  4484085  2905728  2905729  2905736  2905716  2905715  2905714  2905713  2905734  2905733  2904437  2905746  2905749  2905726  2905740  2905705  2905709  2904442  2967780  4492036  2967775  2967783  2967778  2967777  2845409  2967785  2967776  2967786  2967784  2967779  2967782  4492041  4512790  2967781  2845393  4480053  4505413  4505394  4468386  4468397  4468394  2967773  4468389  2967772  4468382  4468384  4468379  4468376  4468385  4478897  4468398  4482427  4512801  4480061  4480054  4470334  4470351  4470347  4480056  4482431  4471051  4482424  4482419  4482425  4592885  4472655  4472650  4482423  4470318  4505410  4478911  4478910  4480065  4470327  4479303  4480062  4468383  4478900  4470316  4470328  4470322  4470337  2969195  4480046  4470357  4491691  4478899  4484548  4484549  4484580  2889061  4505402  4491939  4471189  4491699  4491693  4491925  4491696  4491937  4491697  4478896  4462965  4462977  4462960  4462958  4462978  4462974  4462941  4478912  4462954  4472686  4462945  4462940  4462970  4471048  4471191  4491921  4462939  4512811  4491920  4491700  4491931  4492037  2898935  2898939  4491934  2858416  4491935  2896557  2898928  2841579  2898927  2967774  4505401  2898938  4491914  4462986  4491913  4491941  4491929  4491928  4491915  4491926  4491919  4491922  4491923  4491938  4491695  4491933  4491932  4491692  4491694  2908615  2908612  2908589  2908632  2908643  2908642  2908617  2908619  2908620  2908590  4530897  4530896  4530892  4530885  4530890  2908639  2908638  4517460  2908622  2908614  2908582  2908586  2908646  2908640  2908641  2908637  2908616  2908618  2908631  2908624  2908623  4530901  2908635  4515194  4530908  4530893  4530891  4530895  4519386  4522093  4530886  4522214  4530888  4522092  4522889  4522217  4522209  4505414  4471049  4514947  4530900  2908644  4530912  4530910  4530907  4530905  4530902  4530894  4530913  4530899  4530906  4530904  4530903  4530898  4530887  4530889  4530909  2928878  2934404  2928450  2928882  2928454  2928453  2908628  2928452  2934403  2928875  2928451  2928880  2928876  2928884  2928881  2928883  2934408  2937055  2937073  2937070  2937061  2937075  2937078  2934405  2937049  2934407  2934402  2934401  2934400  2934399  2934406  2924632  2928879  2905719  2928877  2908633  2908630  2908625  2908627  2905706  2902712  2905722  2898934  2908626  2905725  2908591  2902711  2902710  2908613  2905708  2924627  4513027  2924628  2924635  2924634  2924631  2924630  2908634  2924633  2924636  2898926  2908629  2904440  2908621  2908645  2905711  2924629  4592461  4592473  4468395  4462948  4462984  4468396  4592478  4592454  4592482  4468375  4468381  2849532  2849550  2849548  2849547  4592472  4470354  4522101  4470348  4470355  4592476  4592293  4470323  4592468  4592467  4592459  4592471  4592464  4592451  4592291  4592300  4592465  4592296  4470340  2849554  4470332  4470321  4470329  4470325  4470319  4470312  4470341  4470313  4470339  4470349  4470356  2969193  2969197  2969194  4470310  4471179  4592474  4468388  4592483  4592887  4470353  4470333  4470335  4470343  2841532  4472647  4472642  4472661  4472660  4472659  4470344  4470320  4470324  4471044  4472669  4472668  4471190  4472641  4470342  4472678  4512805  4472676  4471180  4592455  4592477  2969192  2969199  4472663  4492518  4496158  4462951  4462935  4462943  4462952  4462944  4471052  4462989  4471054  4492529  4592298  4472644  4468390  4468378  2896563  4462937  4462955  2969196  4470331  4592479  4592475  4592457  4470346  4592470  4470311  4470336  4470317  4470338  4470358  4470352  4592452  4592292  4470330  4462938  2969201  2896555  2896559  2896558  2898937  2898931  4470326  2898929  2969198  4462967  4462988  4468392  4468391  4468380  4468377  2898932  4592882  4592881  4592481  4592884  4592294  4592480  2889059  4495972  4592484  4492042  2896565  4491930  4492500  2847162  2837573  4592883  4511233  2847171  4512812  4492532  4492535  4492537  4492495  2861268  4513014  4592886  2889057  2889065  2889058  4492521  4492497  2841552  4513019  2847157  2841527  2896560  2841543  2858417  2858420  2858419  2841593  2858414  2841560  2847167  2847166  4512816  2847153  2847163  4462959  2858415  2845395  2889064  2841570  2841529  2841581  2898933  2845406  2858422  2845401  2841548  2845394  2845399  4513015  2841577  2841568  4512820  2845397  4514944  4515203  4515197  4514934  4512789  4484545  4512803  4478909  4484570  4514939  2861255  2861263  2861265  2861258  2858286  4478913  4484098  4484087  4496335  4515204  4515180  4512831  4515183  4514959  4484088  4484567  4484292  4514935  4514949  4514950  4484544  4515172  4484289  4512800  2861267  2964229  2964235  2964230  2964234  2964232  2964238  4513018  2964239  4512804  4478898  4478915  4512825  4513026  4592466  2964240  4512807  2847169  4512826  4478903  4478907  4478901  4478914  2964233  4512832  4515174  2964228  2964227  2964231  2964237  2964236  2964241  4512794  4472652  2836951  2837631  2837531  2836961  4472673  2837556  4472658  2837462  2837469  2837567  2841553  2847150  2841596  2861262  4472653  2837525  2847159  4492499  2845402  2845392  2845408  2845396  4484093  2837481  2836941  2841537  2841558  2845410  2841528  2841518  2851860  4492032  2851869  2851865  2849541  2849539  2849530  2837529  2851866  2837593  2851867  2837606  2851859  2851858  2851861  2851857  2837489  2837437  2851864  4484078  2841571  2847151  2837562  2837608  2837442  4472643  2837445  4472677  2851855  4482444  4505407  4505416  4482426  2837482  2836957  4472689  2837545  2837506  2837502  2837461  2837629  2837561  2837435  2837451  2837454  2836964  2836955  2837613  2837552  2847158  2847145  2837438  2837537  2837471  2837453  2837443  2837467  2837549  2837509  2836966  2837474  2837441  2837507  2837564  2837600  2837497  2836945  2836943  2837485  2845398  2849546  2837484  4505399  4471045  4492502  4505415  2837505  2837583  2847154  2845403  2847165  2837554  2845404  2845390  2845400  2841582  4487863  4478905  2849549  2849525  2841588  2836939  2837476  2837487  4505403  2837574  2841565  2841550  2837594  2837463  2849542  2841551  4505404  2893345  4492503  4490311  4492040  4492504  4484551  4490516  4496336  4490309  2893351  4508404  4510918  4512817  4512793  2966241  4492513  4490505  4484100  4490514  4479298  4492501  4491917  4490510  4492507  4490312  4492025  4492538  4492044  4490511  4490501  4490503  2966239  4490519  4482422  2966248  2969200  4512799  4512802  4510905  2893350  2947162  4513022  4513025  4480048  2893346  2893355  4490507  4490506  4482429  4513016  2966244  4490520  2966237  2966247  4512795  4512822  2966251  4512823  2966245  2966250  2966238  2966243  4512830  4512829  4462962  2937052  2966246  4484576  4472654  4478906  4479297  4479296  2898940  4490513  2893344  4592460  4484291  4479306  4479305  4484547  4484079  4484286  2896562  4484539  4472680  4471050  4471047  4473449  4484096  4484575  2851862  4484280  4592458  4484282  4484563  4592485  4592469  2854143  2962095  4484569  4492514  4479311  4510897  4517464  4517454  4511234  4492543  4492033  4479299  4479315  4492034  4490313  4490512  4492024  4492046  4490502  4492028  4479302  2893357  4479313  4479312  4479308  4484572  4484571  4490508  4484084  4479314  4479300  4479309  4479307  2898936  2889066  4517245  4484076  4484097  4484092  4484284  4484566  4478904  4471056  4482433  4484565  4484089  4484099  4484559  4484560  4484540  4492509  4490310  4471058  4513020  4515193  4515179  4515177  4512813  4515200  4515176  4484541  4514932  4484564  4512796  4512824  4512821  4471182  4484090  4495967  4514943  4515171  4484557  4515182  4514957  4514956  4515173  4514951  4515208  4515195  4514952  4514937  4513017  4495971  4490306  4484558  4471184  4515169  4492035  4515205  4492528  4492541  4484550  4484553  4484578  4514946  4492508  4480060  4492517  4512819  4515198  4514942  4514945  4512792  4495975  4508422  2849533  2893358  2893348  4490315  4490515  4490509  4482420  4508644  4482430  4472670  4470345  4470309  2889060  4515170  4514961  4508410  4490314  4484568  2889063  4490308  4490504  4482442  2893349  4479304  2893354  2849531  2893347  2893356  4482439  4490517  2889062  2893353  2893352  4514954  4515178  2849551  4514941  4515189  4472649  4515184  4515175  4515192  4514940  4515190  4515185  4515181  4515186  4515201  4490518  4515202  4514933  2849526  2849545  4472657  4472679  4514953  4472683  2849553  4515196  4514938  4491942  4514936  4514960  2849552  4515207  4515191  4471187  2951012  2953504  2951017  2951011  2944950  2944952  2944951  2953505  2951014  2953506  2951010  2944937  2944938  2944934  2951013  2951015  2944953  2953515  2947163  2953507  2953508  2953503  2953501  2953500  2951004  2951000  2944936  2951003  2951008  2951007  2951006  2951005  2953511  2951009  2947158  2947169  2947184  2947156  2947143  2946937  2946936  2951018  2947155  2947001  2947151  2946930  2947139  2946995  2947148  2947161  2947189  2947195  2953496  2951016  2951002  2944935  2944942  2944946  2947191  2944943  2947150  2947160  2947185  2947206  2947168  2942504  2946942  2944948  2944944  2960763  2953512  2962089  2962084  2960760  2960768  2960762  2962091  2960765  2962087  2960767  2960766  2960759  2956632  2956636  2957992  2960761  2962094  2836969  2937063  2947205  4514955  4530911  2960764  2962090  2962096  2950999  2962093  2962092  2962088  2962086  2962082  2962085  2962083  2956630  2956649  2956635  2956640  2956633  2956646  2956643  2956642  2956634  2957993  2953502  2953498  2953513  2953509  2951001  2953497  2956637  2956639  2957985  2956644  2956641  2957990  2957989  2957991  2957987  2956631  2957988  2956648  2953514  2956645  2953510  2953499  2956647  2956638  2957994  2942491  2946933  2942501  2940223  2942493  2942503  2942494  2942496  2942492  2942505  2942495  2942500  2905735  2940221  2940231  2940222  2942498  2937072  2942508  2946932  2946997  2946999  2946994  2946940  2942502  2947006  2940240  2942489  2942506  2942497  2942490  2942499  2942507  2947002  2937077  2937076  2937068  2937067  2937058  2937071  2937050  2940230  2937053  2937069  2937060  2937062  2947167  2957986  2937057  2937074  2937051  2940233  2940229  2940227  2940237  2940232  2940239  2940224  2937066  2940238  2937059  2940225  2940235  2940226  2937054  2937065  2937064  2940228  2940236  2947200  2947144  2946998  2947149  2947182  2947188  2947187  2947164  2947177  2947180  2947142  2947005  2947171  2947153  2946943  2947157  2947178  2947186  2947199  2947166  2947198  2947197  2940234  2946934  2947141  2947174  2947203  2947165  2947172  2947152  2947190  2947008  2947207  2946996  2947179  2946929  2947208  2947147  2944947  2944939  2944949  2947003  2947000  2947192  2946935  2944941  2947146  2944940  2946939  2944945  2947004  2947007  2946938  2946931  2947204  2947145  2947194  2947202  2947176  2947183  2946941  2947173  2947154  2947201  2947196  2947159  2947193  2947170  2947140  2957303  4466273  2818717  2818708  4482413  4466271  2818702  4491000  4466272  4472024  4466274  4466277  4466275  4485060  4487860  2965405  4466276  2965402  2965399  2965400  2965406  2965404  2965398  4482414  2818718  2965397  2818714  2818703  2818711  2818715  2818698  2818700  4490995  2965403  2963299  4431574  4431570  4431571  4431577  4431580  4490994  2963297  4431573  4501434  4508639  4490998  4501437  2818707  4501436  2963306  2957300  4485059  4487861  4490997  4501438  4501435  2818697  4431578  2957301  4431581  2957302  4431575  4431576  4431579  4431572  4490999  2957304  2847200  2868533  2868534  2818716  2818699  2868535  4501439  2818704  2868537  2847202  2847201  2847211  2847199  4508638  2818706  2847204  4466278  4490996  2963298  2963300  2963305  2963301  2963295  2868532  2847203  2868536  2818713  2843912  2847209  2847212  2847206  2847210  2963302  2963292  2843914  2843916  2868531  2847207  2868530  2818701  2843911  4472023  4487862  2818719  2963304  2963303  2963294  2847205  2818705  4472025  4482417  2847208  2818720  2868529  2843917  2843909  2847198  4482412  2868528  4482418  4482415  2843910  2843915  2843913  2843908  4482416  2669477  2669483  2669512  2669478  2669486  2669492  2669518  2669491  2669496  2669493  2669509  2669508  2669514  2669479  2669533  2669538  2669535  2669519  2669513  2669510  2669506  2669488  2669484  2669481  2669480  2669532  2669489  2669497  2669505  2696338  2696337  2696334  2696333  2696332  2696344  2669537  2696331  2696342  2669516  2669502  2669511  2669507  2669498  2669528  2696343  2700407  2669500  2669525  2700406  2718573  2718567  2700410  2696340  2700408  2696341  2700411  2700413  2700412  2700405  2696336  2696339  2696330  2700409  2669526  2669499  2669501  2669504  2669487  2669476  2669475  2669490  2669534  2669527  2669482  2669523  2669521  2669530  2669529  2669515  2696335  2669495  2669494  2669503  2669520  2669536  2669524  2669522  2669531  2700414  2669485  2669517  2669539  3002748  2718565  2718563  2718562  2821471  2821469  2821468  2821472  2821466  2821467  2941127  3002749  2821470  2941121  2723386  2941130  2941124  2941129  2941125  2941123  3004785  3004784  3050039  3050038  2723388  2276860  2941126",
      "more_code_info": ""
    }
  ]
}