  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Jerusalem",
      "state": "",
      "country": "Israel",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "80895",
      "recalling_firm": "Oridion Medical 1987 Ltd.",
      "address_1": "Har Hotzvim Industrial Park Box 45025",
      "address_2": "Har Hotzvim Industrial Park",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "The products were distributed US nationwide and Puerto Rico.    The products were distributed to the following foreign countries: Argentina, Australia, Austria, Bahrain, Belgium, Canada, Columbia, Costa Rica, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Hungary, Ireland, Israel, Italy, Japan, Kazakhstan, Kuwait, Mexico, Netherlands, Norway, Oman, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Thailand, Tunisia, Turkey, UAE, UK, and Vietnam.",
      "recall_number": "Z-0033-2019",
      "product_description": "Capnostream20 (US)M REFURBISHED, Part Number CS78653",
      "product_quantity": "43",
      "reason_for_recall": "The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory default settings when the monitor is powered off.",
      "recall_initiation_date": "20180731",
      "center_classification_date": "20181003",
      "termination_date": "20220920",
      "report_date": "20181010",
      "code_info": "B300001025  B300001053  B300001106  B300001109  B300001140  B300001178  B300001219  B300001282  B300001304  B300001324  B300001262  B300001094  B300001177  B300001208  B300001242  B300001246  B300001313  B500002196  B300001188  B300001236  B300002173  B500002408  B300001071  B300001075  B300001124  B300001204  B300001248  B300001443  B500001655  B500001659  B500001660  B500001666  B500001670  B500001683  B500001701  B500001806  B500001817  B500002130  B500002189  B500002231  B300001066  B300001244  B300001207",
      "more_code_info": ""
    }
  ]
}