  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Cleveland",
      "state": "OH",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "66312",
      "recalling_firm": "Philips Medical Systems (Cleveland) Inc",
      "address_1": "595 Miner Rd",
      "address_2": "N/A",
      "postal_code": "44143-2131",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US (nationwide) including FL, CO, KY, MS, PA, NC, NJ, TX and Puerto Rico and internationally to Italy.",
      "recall_number": "Z-0033-2014",
      "product_description": "Philips MR Systems: Asset 0.5T, Apollo 0.5T, Infinion 1.5T, Eclipse/Polaris 1.5T, Panorama 0.6T     1.0T, Magnetic Resonance Imaging Systems intended for clinical/radiologic diagnostic imaging applications.",
      "product_quantity": "Total 511 (Asset 0.5T,: 35 units Apollo 0.5T: 15 units, Infinion 1.5T: 116 units, Eclipse/Polaris 1.5T: 290 units, Panorama 0.6T: 55 units)",
      "reason_for_recall": "There is a potential for water to collect in the vent pipe elbow in the magnet venting system.  Water may freeze, blocking the venting system.",
      "recall_initiation_date": "20130807",
      "center_classification_date": "20131021",
      "termination_date": "20140326",
      "report_date": "20131030",
      "code_info": "M/N: 781220, 781230, 781301, 78129 and 78127.  Codes: AM128, 1510, 1518, BA318, AM120, ZA178, ZA196, ZA107, ZA133, ZA172, ZA158, BA149 and ZA208."
    }
  ]
}