      "status": "Terminated",
      "city": "Plano",
      "state": "TX",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "78154",
      "recalling_firm": "St. Jude Medical, Inc.",
      "address_1": "6901 Preston Rd",
      "address_2": "",
      "postal_code": "75024-2508",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US (nationwide) (foreign consignee list will be provide by the firm at a later time)",
      "recall_number": "Z-0024-2018",
      "product_description": "Proclaim DRG Implantable Pulse Generator Model 3664",
      "product_quantity": "1744 devices",
      "reason_for_recall": "Analysis of the difference between the actual device longevity and the battery indicator status provided by the Clinician Programmer (CP) or Patient Controller (PC) has revealed errors in the CP or PC longevity estimations that can overestimate or under estimate the battery indicator status in certain situations.",
      "recall_initiation_date": "20170912",
      "center_classification_date": "20171017",
      "termination_date": "20200416",
      "report_date": "20171025",
      "code_info": "Serials implanted prior to September 12, 2017 are considered part of this advisory; Serial Numbers -   AAP114.1  AAP267.1  AAP380.1  AAT623.1  ABM656.1  ABT952.1  ACD631.1  ACE905.1  ADJ588.1  ADJ614.1  ADJ842.1  ADJ852.1  ADJ855.1  ADJ863.1  ADJ872.1  ADJ910.1  ADK057.1  ADK078.1  ADK116.1  ADK186.1  ADK293.1  ADN214.1  ADN311.1  ADN340.1  ADN347.1  ADN376.1  ADN398.1  ADR256.1  ADR258.1  ADR260.1  ADR262.1  ADR289.1  ADR299.1  ADR302.1  ADR313.1  ADR362.1  ADR364.1  ADR365.1  ADR400.1  ADR401.1  ADR405.1  ADR410.1  ADR416.1  ADR421.1  ADR433.1  ADR434.1  ADR453.1  ADR457.1  ADR460.1  ADR461.1  ADR463.1  ADR464.1  ADR466.1  ADR469.1  ADR474.1  ADR475.1  ADR477.1  ADR485.1  ADR489.1  ADR492.1  ADR495.1  ADR497.1  ADR498.1  ADR501.1  ADR505.1  ADR509.1  ADR517.1  ADR519.1  ADR526.1  ADR543.1  ADR552.1  ADR557.1  ADR558.1  ADR563.1  ADR566.1  ADR575.1  ADR581.1  ADR598.1  ADR617.1  ADR630.1  ADR632.1  ADR655.1  ADR975.1  ADR986.1  ADS012.1  ADS039.1  ADW504.1  ADW624.1  ADW635.1  ADW686.1  ADW706.1  ADW707.1  ADW724.1  ADW930.1  AEH786.1  AEH900.1  AEJ018.1  AEJ059.1  AEM902.1  AFB279.1  AFB456.1  AFB710.1  AFB717.1  AFB934.1  AFG268.1  AFG397.1  AFG415.1  AFG464.1  AFG577.1  AFG587.1  AFG606.1  AFG780.1  AFG805.1  AFH511.1  AFH513.1  AFH514.1  AFH516.1  AFH518.1  AFH519.1  AFH521.1  AFH522.1  AFH524.1  AFH526.1  AFH527.1  AFH528.1  AFH530.1  AFH535.1  AFH538.1  AFH539.1  AFH543.1  AFH544.1  AFH545.1  AFH546.1  AFH547.1  AFH552.1  AFH554.1  AFH555.1  AFH556.1  AFH557.1  AFH558.1  AFH560.1  AFH561.1  AFH562.1  AFH563.1  AFH564.1  AFH565.1  AFH566.1  AFH570.1  AFH571.1  AFH572.1  AFH576.1  AFH579.1  AFH580.1  AFH582.1  AFH583.1  AFH584.1  AFH588.1  AFH589.1  AFH590.1  AFH592.1  AFH593.1  AFH594.1  AFH596.1  AFH597.1  AFH598.1  AFH599.1  AFH600.1  AFH601.1  AFH602.1  AFH605.1  AFH606.1  AFH607.1  AFH608.1  AFH609.1  AFH610.1  AFH614.1  AFH615.1  AFH618.1  AFH619.1  AFH620.1  AFH621.1  AFH624.1  AFH625.1  AFH626.1  AFH628.1  AFH629.1  AFH630.1  AFH631.1  AFH632.1  AFH633.1  AFH636.1  AFH641.1  AFH642.1  AFH646.1  AFH647.1  AFH648.1  AFH651.1  AFH654.1  AFH660.1  AFH663.1  AFH664.1  AFH665.1  AFH666.1  AFH668.1  AFH669.1  AFH671.1  AFH673.1  AFH674.1  AFH676.1  AFH684.1  AFH685.1  AFH687.1  AFH689.1  AFH695.1  AFH696.1  AFH697.1  AFH699.1  AFH700.1  AFH701.1  AFH702.1  AFH703.1  AFH706.1  AFH707.1  AFH708.1  AFZ281.1  AFZ365.1  AFZ581.1  AFZ582.1  AFZ592.1  AFZ620.1  AFZ631.1  AFZ637.1  AFZ642.1  AFZ649.1  AFZ661.1  AFZ685.1  AFZ691.1  AFZ695.1  AFZ710.1  AFZ711.1  AFZ728.1  AFZ730.1  AFZ768.1  AFZ780.1  AFZ870.1  AFZ881.1  AFZ893.1  AFZ908.1  AFZ915.1  AFZ918.1  AFZ930.1  AFZ933.1  AFZ970.1  AFZ972.1  AFZ978.1  AFZ979.1  AFZ981.1  AFZ998.1  AGA010.1  AGA012.1  AGA015.1  AGA029.1  AGA031.1  AGA032.1  AGA040.1  AGA044.1  AGA047.1  AGA049.1  AGA053.1  AGA061.1  AGA064.1  AGA067.1  AGA068.1  AGA071.1  AGA078.1  AGA083.1  AGA085.1  AGA087.1  AGA088.1  AGA094.1  AGA104.1  AGA108.1  AGA110.1  AGA113.1  AGM588.1  AGM603.1  AGN178.1  AGN273.1  AGN393.1  AGN394.1  AGN429.1  AGN476.1  AGN514.1  AGU357.1  AHD592.1  AHD733.1  AHD737.1  AHD739.1  AHD742.1  AHD745.1  AHD756.1  AHD759.1  AHD760.1  AHD763.1  AHD766.1  AHD775.1  AHD787.1  AHD795.1  AHD800.1  AHK478.1  AHK479.1  AHK480.1  AHK483.1  AHK485.1  AHK489.1  AHK491.1  AHK493.1  AHK498.1  AHK507.1  AHK516.1  AHK520.1  AHK527.1  AHK529.1  AHK533.1  AHK538.1  AHK540.1  AHK549.1  AHK570.1  AHK574.1  AHK583.1  AHK591.1  AHK601.1  AHK605.1  AHK610.1  AHK615.1  AHK655.1  AHK656.1  AHK688.1  AHK689.1  AHK722.1  AHK733.1  AHK736.1  AHK743.1  AHK763.1  AHK774.1  AHK812.1  AHK826.1  AHK959.1  AHL025.1  AHL058.1  AHL172.1  AHL182.1  AHV709.1  AHV710.1  AHV714.1  AHV715.1  AHV720.1  AHV732.1  AHV742.1  AHV752.1  AHV754.1  AHV757.1  AHV759.1  AHV761.1  AHV765.1  AHV766.1  AHV770.1  AHV777.1  AHV781.1  AHV995.1  AHW119.1  AHW177.1  AHW181.1  AHW187.1  AHW189.1  AHW191.1  AHW205.1  AHW206.1  AHW211.1  AHW213.1  AHW214.1  AHW230.1  AHW232.1  AHW240.1  AHW243.1  AHW251.1  AHW257.1  AHW261.1  AHW265.1  AHW277.1  AHW284.1  AHW310.1  AHW317.1  AHW329.1  AHW337.1  AHW341.1  AHW344.1  AHW345.1  AHW353.1  AHW358.1  AHW373.1  AHW377.1  AHW386.1  AHW389.1  AHW399.1  AHW400.1  AHW411.1  AHW419.1  AHW422.1  AHW423.1  AHW427.1  AHW439.1  AHW444.1  AHW448.1  AHW455.1  AHW458.1  AHW459.1  AHW460.1  AHW468.1  AHW469.1  AHW473.1  AHW478.1  AHW486.1  AHW509.1  AHW511.1  AHW516.1  AHW518.1  AHW520.1  AHW521.1  AHW524.1  AHW530.1  AHW536.1  AHW538.1  AHW539.1  AHW540.1  AHW549.1  AHW560.1  AHW566.1  AHW576.1  AHW597.1  AHW601.1  AHW617.1  AHW620.1  AHW622.1  AHW624.1  AHW634.1  AHW639.1  AHW644.1  AHW645.1  AHW649.1  AHW652.1  AJR080.1  AJW366.1  AJW459.1  AJW507.1  AJW511.1  AJW554.1  AJW558.1  AJW572.1  AJW580.1  AJW596.1  AJW607.1  AJW621.1  AJW623.1  AKB521.1  AKB527.1  AKB531.1  AKB537.1  AKB560.1  AKB565.1  AKB566.1  AKB573.1  AKB584.1  AKB595.1  AKB598.1  AKB612.1  AKB613.1  AKB614.1  AKB621.1  AKB633.1  AKB638.1  AKB640.1  AKB642.1  AKB650.1  AKB654.1  AKB675.1  AKB682.1  AKB683.1  AKB712.1  AKB714.1  AKB719.1  AKB724.1  AKB734.1  AKB736.1  AKB755.1  AKB756.1  AKB759.1  AKB762.1  AKB767.1  AKB776.1  AKB779.1  AKB780.1  AKB783.1  AKB785.1  AKB802.1  AKB806.1  AKB809.1  AKB810.1  AKB814.1  AKB825.1  AKB828.1  AKB832.1  AKB843.1  AKB848.1  AKB856.1  AKB864.1  AKB866.1  AKB869.1  AKB876.1  AKB881.1  AKB896.1  AKB897.1  AKB901.1  AKB903.1  AKB922.1  AKB940.1  AKB942.1  AKB951.1  AKB952.1  AKB956.1  AKB959.1  AKB962.1  AKB974.1  AKB979.1  AKB981.1  AKB985.1  AKB994.1  AKC002.1  AKC007.1  AKC010.1  AKC012.1  AKC022.1  AKC024.1  AKC026.1  AKC028.1  AKC049.1  AKC051.1  AKC055.1  AKC060.1  AKC076.1  AKC081.1  AKC086.1  AKC111.1  AKC112.1  AKC114.1  AKC119.1  AKC123.1  AKC126.1  AKC136.1  AKC144.1  AKC155.1  AKC159.1  AKC163.1  AKU509.1  AKU514.1  AKU522.1  AKU524.1  AKU525.1  AKU527.1  AKU530.1  AKU538.1  AKU550.1  AKU573.1  AKU574.1  AKU577.1  AKU578.1  AKU585.1  AKU593.1  AKU597.1  AKU606.1  AKU622.1  AKU625.1  AKU636.1  AKU638.1  AKU655.1  AKU656.1  AKU662.1  AKU668.1  AKU670.1  AKU680.1  AKU686.1  AKU692.1  AKU693.1  AKU695.1  AKU713.1  AKU714.1  AKU715.1  AKU716.1  AKU718.1  AKU722.1  AKU725.1  AKU731.1  AKU734.1  AKU737.1  AKU739.1  AKU753.1  AKU760.1  AKU763.1  AKU772.1  AKU777.1  AKU791.1  AKU796.1  AKU798.1  AKU801.1  AKU806.1  AKU809.1  AKU810.1  AKU814.1  AKU817.1  AKU836.1  AKU839.1  AKU844.1  AKU848.1  AKU851.1  AKU862.1  AKU887.1  AKU890.1  AKU891.1  AKU897.1  AKU911.1  AKU914.1  AKU915.1  AKU920.1  AKU925.1  AKU931.1  AKU937.1  AKU947.1  AKU964.1  AKU967.1  AKU971.1  AKU977.1  AKU986.1  AKU987.1  AKU991.1  AKU992.1  AKU993.1  AKV001.1  AKV003.1  AKV004.1  AKV009.1  AKV012.1  AKV016.1  AKV022.1  AKV036.1  AKV047.1  AKV050.1  AKV058.1  AKV059.1  AKV067.1  AKV077.1  AKV089.1  AKV090.1  AKV092.1  AKV094.1  AKV095.1  AKV096.1  AKV100.1  AKV101.1  AKV102.1  AKV106.1  AKV114.1  AKV120.1  AKV130.1  AKV135.1  AKV137.1  AKV144.1  AKV147.1  AKV148.1  AKV157.1  AKV163.1  AKV167.1  AKV168.1  AKV169.1  AKV170.1  AKV171.1  AKV173.1  AKV177.1  AKV188.1  AKV189.1  AKV199.1  AKV212.1  AKV220.1  AKV221.1  AKV225.1  AKV226.1  AKV228.1  AKV233.1  AKV240.1  AKV241.1  AKV242.1  AKV252.1  AKV255.1  AKV256.1  AKV261.1  AKV265.1  AKV267.1  AKV269.1  AKV270.1  AKV271.1  AKV274.1  AKV276.1  AKV277.1  AKV281.1  AKV284.1  AKV286.1  AKV287.1  AKV288.1  AKV289.1  AKV299.1  AKV300.1  AKV307.1  AKV311.1  AKV314.1  AKV318.1  AKV320.1  AKV322.1  AKV324.1  AKV330.1  AKV335.1  AKV337.1  AKV338.1  AKV350.1  AKV355.1  AKV357.1  AKV358.1  AKV361.1  AKV363.1  AKV370.1  AKV371.1  AKV372.1  AKV377.1  AKV386.1  AKV387.1  AKV391.1  AKV394.1  AKV396.1  AKV397.1  AKV399.1  AKV401.1  AKV403.1  AKV406.1  AKV407.1  AKV409.1  AKV411.1  AKV417.1  AKV421.1  AKV431.1  AKV440.1  AKV445.1  AKV447.1  AKV451.1  AKV458.1  AKV464.1  AKV466.1  AKV467.1  AKV468.1  AKV475.1  AKV486.1  AKV488.1  AKV489.1  AKV492.1  AKV507.1  AKV611.1  ALE754.1  ALE758.1  ALE766.1  ALE773.1  ALE941.1  ALE978.1  ALE983.1  ALE988.1  ALE994.1  ALF010.1  ALF014.1  ALF049.1  ALF063.1  ALF064.1  ALF077.1  ALF101.1  ALF111.1  ALF142.1  ALF153.1  ALF167.1  ALJ325.1  ALJ331.1  ALJ353.1  ALJ371.1  ALJ373.1  ALJ413.1  ALJ414.1  ALJ416.1  ALJ432.1  ALJ458.1  ALJ795.1  ALJ824.1  ALK002.1  ALK014.1  ALK032.1  ALK038.1  ALK051.1  ALK061.1  ALY954.1  ALY991.1  ALZ109.1  ALZ110.1  ALZ115.1  ALZ117.1  ALZ118.1  ALZ128.1  ALZ131.1  ALZ138.1  ALZ139.1  ALZ222.1  ALZ226.1  ALZ232.1  ALZ233.1  ALZ247.1  ALZ254.1  ALZ265.1  ALZ269.1  ALZ270.1  ALZ276.1  ALZ281.1  ALZ282.1  ALZ287.1  ALZ289.1  ALZ291.1  ALZ302.1  ALZ305.1  ALZ307.1  ALZ360.1  ALZ374.1  ALZ378.1  ALZ379.1  ALZ384.1  ALZ392.1  ALZ393.1  ALZ402.1  ALZ407.1  ALZ412.1  ALZ413.1  ALZ414.1  ALZ427.1  ALZ429.1  ALZ430.1  AMJ773.1  AMJ848.1  AMK014.1  AMN719.1  AMN757.1  AMN973.1  AMN975.1  AMN986.1  AMN988.1  AMN991.1  AMP024.1  AMP228.1  AMP270.1  AMP273.1  AMR918.1  AMR929.1  AMR937.1  AMR939.1  AMR943.1  AMR954.1  AMR956.1  AMR959.1  AMR961.1  AMR963.1  AMR975.1  AMR976.1  AMR979.1  AMR982.1  AMR983.1  AMR987.1  AMR989.1  AMS203.1  AMS257.1  AMS267.1  AMS269.1  AMS289.1  AMS292.1  AMS293.1  AMS296.1  AMS305.1  AMS312.1  AMS324.1  AMS329.1  AMS334.1  AMS339.1  AMS340.1  AMS344.1  AMS361.1  AMS363.1  AMS370.1  AMS371.1  AMS391.1  AMS440.1  AMS709.1  AMS778.1  AMS795.1  AMT002.1  AMT011.1  AMT052.1  AMT057.1  AMT068.1  AMT086.1  AMT087.1  AMT099.1  AMT101.1  AMT105.1  AMT111.1  AMT121.1  AMT124.1  AMT195.1  AMT255.1  AMT310.1  AMT338.1  AMT345.1  AMT348.1  AMT349.1  AMT355.1  AMT365.1  ANG890.1  ANG929.1  ANG967.1  ANG972.1  ANH010.1  ANH023.1  ANH042.1  ANH067.1  ANH095.1  ANH098.1  ANH108.1  ANH111.1  ANH114.1  ANH115.1  ANH121.1  ANH139.1  ANH152.1  ANH155.1  ANH157.1  ANH174.1  ANH188.1  ANH196.1  ANH197.1  ANH226.1  ANH231.1  ANH235.1  ANH239.1  ANH242.1  ANH254.1  ANH262.1  ANH264.1  ANH268.1  ANH270.1  ANH273.1  ANH281.1  ANH286.1  ANH394.1  ANH395.1  ANH396.1  ANH422.1  ANH423.1  ANH425.1  ANH430.1  ANH431.1  ANH437.1  ANH441.1  ANH445.1  ANH447.1  ANH458.1  ANH459.1  ANH463.1  ANH482.1  ANH483.1  ANH484.1  ANH502.1  ANH504.1  ANH506.1  ANH522.1  ANH534.1  ANH535.1  ANH558.1  ANH565.1  ANH579.1  ANH583.1  ANH584.1  ANH587.1  ANH588.1  ANH590.1  ANH591.1  ANH592.1  ANH609.1  ANH616.1  ANH621.1  ANH629.1  ANH640.1  ANH650.1  ANL380.1  ANL382.1  ANL385.1  ANL388.1  ANL389.1  ANL392.1  ANL393.1  ANL395.1  ANL396.1  ANL397.1  ANL398.1  ANL399.1  ANL402.1  ANL403.1  ANL404.1  ANL407.1  ANL408.1  ANL409.1  ANL410.1  ANL412.1  ANL413.1  ANL415.1  ANL418.1  ANL420.1  ANL422.1  ANL423.1  ANL424.1  ANL425.1  ANL426.1  ANL427.1  ANL428.1  ANL430.1  ANL431.1  ANL434.1  ANL435.1  ANL436.1  ANL437.1  ANL438.1  ANL452.1  ANL455.1  ANL460.1  ANL461.1  ANL462.1  ANL464.1  ANL468.1  ANL469.1  ANL474.1  ANL476.1  ANL480.1  ANL482.1  ANL491.1  ANL493.1  ANL497.1  ANL498.1  ANL500.1  ANL501.1  ANL502.1  ANL506.1  ANL510.1  ANL517.1  ANL518.1  ANL519.1  ANL521.1  ANL541.1  ANL558.1  ANL564.1  ANL590.1  ANL684.1  ANL686.1  ANL687.1  ANL691.1  ANL694.1  ANL703.1  ANL706.1  ANL721.1  ANL726.1  ANL732.1  ANL734.1  ANL739.1  ANL744.1  ANL781.1  ANL836.1  ANL842.1  ANL868.1  ANL910.1  ANM069.1  ANT277.1  ANT279.1  ANT283.1  ANT286.1  ANT287.1  ANT296.1  ANT297.1  ANT298.1  ANT300.1  ANT307.1  ANT320.1  ANT342.1  ANT383.1  ANT385.1  ANT404.1  ANT411.1  ANT439.1  ANT441.1  ANT451.1  ANT452.1  ANT461.1  ANT472.1  ANT476.1  ANT477.1  ANT478.1  ANT481.1  ANT484.1  ANT534.1  ANT557.1  ANT574.1  ANT618.1  ANT637.1  ANT666.1  ANT754.1  ANT867.1  ANT888.1  ANT896.1  ANT900.1  ANT908.1  ANT911.1  ANT913.1  ANU020.1  ANU033.1  ANU075.1  ANU077.1  ANU088.1  ANU141.1  ANU172.1  ANU179.1  ANU214.1  ANU217.1  ANU244.1  ANU271.1  ANU298.1  ANU300.1  ANU321.1  ANU329.1  ANU333.1  ANU334.1  ANU387.1  ANU391.1  ANU398.1  ANU401.1  APA828.1  APA829.1  APA830.1  APA833.1  APA843.1  APA852.1  APA863.1  APA872.1  APA874.1  APA876.1  APA879.1  APA888.1  APA890.1  APA905.1  APA906.1  APA910.1  APA915.1  APA922.1  APA927.1  APA937.1  APA944.1  APA949.1  APA966.1  APB004.1  APB031.1  APB038.1  APB040.1  APB045.1  APB046.1  APB056.1  APB068.1  APB070.1  APB071.1  APB080.1  APB081.1  APB088.1  APB091.1  APB096.1  APB104.1  APB106.1  APB109.1  APB111.1  APB112.1  APB116.1  APB117.1  APB153.1  APB164.1  APB199.1  APB203.1  APB206.1  APB208.1  APB210.1  APB219.1  APB224.1  APB232.1  APB240.1  APB242.1  APB260.1  APB276.1  APB297.1  APB304.1  APB321.1  APB323.1  APB324.1  APB334.1  APB335.1  APB344.1  APB350.1  APB355.1  APB359.1  APB364.1  APB375.1  APB383.1  APB388.1  APB390.1  APB392.1  APB398.1  APB403.1  APB412.1  APB415.1  APB441.1  APB446.1  APB476.1  APB478.1  APB482.1  APB499.1  APB525.1  APB528.1  APB532.1  APB541.1  APB546.1  APB547.1  APB555.1  APB556.1  APB565.1  APB595.1  APB603.1  APB607.1  APB608.1  APB618.1  APB619.1  APB620.1  APB634.1  APB640.1  APB642.1  APB658.1  APB677.1  APB678.1  APB685.1  APB694.1  APB695.1  APB708.1  APB709.1  APB713.1  APB723.1  APB726.1  APB751.1  APB753.1  APB764.1  APB765.1  APB766.1  APB768.1  APB780.1  APB801.1  APB802.1  APB805.1  APB827.1  APB836.1  APB840.1  APB844.1  APB852.1  APB862.1  APB870.1  APB873.1  APB877.1  APB893.1  APB894.1  APB902.1  APB911.1  APB916.1  APB928.1  APB929.1  APB930.1  APB931.1  APB934.1  APB938.1  APB939.1  APB941.1  APB970.1  APB983.1  APC015.1  APC017.1  APC028.1  APC032.1  APC037.1  APC041.1  APC058.1  APC060.1  APC061.1  APC068.1  APC071.1  APC075.1  APC110.1  APE452.1  APE492.1  APE504.1  APE516.1  APE517.1  APE522.1  APE538.1  APE547.1  APE550.1  APE551.1  APE567.1  APE569.1  APE575.1  APE577.1  APE583.1  APE584.1  APE590.1  APE591.1  APE630.1  APE640.1  APE642.1  APE650.1  APE663.1  APE693.1  APE708.1  APE709.1  APE714.1  APE727.1  APE734.1  APE740.1  APE742.1  APE745.1  APE762.1  APE792.1  APE805.1  APE813.1  APE894.1  APE899.1  APE923.1  APE953.1  APE960.1  APE961.1  APE983.1  APE990.1  APE991.1  APF002.1  APF005.1  APF006.1  APF007.1  APF011.1  APF029.1  APF033.1  APF114.1  APF160.1  APF192.1  APK895.1  APK900.1  APK901.1  APK924.1  APK928.1  APK955.1  APK956.1  APK963.1  APK974.1  APK978.1  APK979.1  APK990.1  APK993.1  APK996.1  APL039.1  APL045.1  APL059.1  APL072.1  APL086.1  APL100.1  APL103.1  APL114.1  APL146.1  APL154.1  APL159.1  APL177.1  APL194.1  APL210.1  APL211.1  APL216.1  APL217.1  APL248.1  APL250.1  APL254.1  APL261.1  APL262.1  APL269.1  APL275.1  APL276.1  APL279.1  APL291.1  APL293.1  APL303.1  APL305.1  APL306.1  APL307.1  APL327.1  APL330.1  APL334.1  APL344.1  APL348.1  APL349.1  APL350.1  APL361.1  APL362.1  APL363.1  APL364.1  APL367.1  APL373.1  APL383.1  APL384.1  APL389.1  APL401.1  APL420.1  APL423.1  APL426.1  APL428.1  APL431.1  APL444.1  APL450.1  APL451.1  APL452.1  APL453.1  APL462.1  APL486.1  APL491.1  APL492.1  APL496.1  APL529.1  APL547.1  APL549.1  APL565.1  APL576.1  APL590.1  APL592.1  APL601.1  APL602.1  APL613.1  APL621.1  APL622.1  APL630.1  APL637.1  APS036.1  APS039.1  APS053.1  APS067.1  APS068.1  APS070.1  APS071.1  APS084.1  APS095.1  APS104.1  APS119.1  APS125.1  APS126.1  APS140.1  APS148.1  APS151.1  APS153.1  APS154.1  APS155.1  APS202.1  APS236.1  APS237.1  APS238.1  APS262.1  APS266.1  APS278.1  APS281.1  APS285.1  APS314.1  APS325.1  APS326.1  APS327.1  APS345.1  APS347.1  APS351.1  APS357.1  APS361.1  APS363.1  APS365.1  APS373.1  APS375.1  APS400.1  APS424.1  APS436.1  APS447.1  APS448.1  APS457.1  APS469.1  APS501.1  APS514.1  APZ186.1  APZ192.1  APZ238.1  APZ251.1  APZ256.1  APZ276.1  APZ277.1  APZ300.1  APZ301.1  APZ323.1  APZ330.1  APZ350.1  APZ412.1  APZ456.1  APZ459.1  APZ560.1  APZ561.1  APZ564.1  APZ587.1  APZ594.1  APZ598.1  APZ608.1  APZ613.1  APZ614.1  APZ618.1  APZ638.1  APZ667.1  APZ670.1  APZ682.1  APZ697.1  APZ710.1  APZ721.1  APZ722.1  APZ758.1  APZ761.1  APZ776.1  APZ819.1  APZ826.1  APZ828.1  ARF828.1  ARF833.1  ARF858.1  ARF859.1  ARF860.1  ARF862.1  ARF867.1  ARF877.1  ARF880.1  ARF885.1  ARF889.1  ARF897.1  ARF949.1  ARF950.1  ARF953.1  ARF961.1  ARF964.1  ARF966.1  ARF970.1  ARF971.1  ARF976.1  ARF982.1  ARF993.1  ARG024.1  ARG028.1  ARG033.1  ARG037.1  ARG039.1  ARG044.1  ARG045.1  ARG047.1  ARG050.1  ARG056.1  ARG059.1  ARG068.1  ARG077.1  ARG123.1  ARG133.1  ARG140.1  ARG146.1  ARG156.1  ARG166.1  ARG168.1  ARG180.1  ARG186.1  ARG196.1  ARG201.1  ARG211.1  ARG214.1  ARG216.1  ARG222.1  ARG226.1  ARG227.1  ARG239.1  ARG254.1  ARG260.1  ARG264.1  ARG265.1  ARG266.1  ARG271.1  ARG273.1  ARG274.1  ARG276.1  ARG280.1  ARG282.1  ARG285.1  ARG297.1  ARG317.1  ARG321.1  ARG340.1  ARG344.1  ARG350.1  ARG368.1  ARG369.1  ARG373.1  ARG379.1  ARG382.1  ARG438.1  ARG442.1  ARG446.1  ARG447.1  ARG462.1  ARG477.1  ARG488.1  ARG490.1  ARG496.1  ARG501.1  ARG502.1  ARG503.1  ARG510.1  ARG512.1  ARG521.1  ARG522.1  ARG525.1  ARG526.1  ARG529.1  ARG535.1  ARG541.1  ARG542.1  ARG550.1  ARG634.1  ARG635.1  ARG639.1  ARG643.1  ARG645.1  ARG649.1  ARG650.1  ARG661.1  ARG662.1  ARJ277.1  ARK058.1  ARK064.1  ARK065.1  ARK092.1  ARK111.1  ARK144.1  ARK149.1  ARK166.1  ARK172.1  ARK188.1  ARK206.1  ARK218.1  ARK219.1  ARK222.1  ARK223.1  ARK224.1  ARK228.1  ARK242.1  ARK244.1  ARK247.1  ARK258.1  ARK261.1  ARK267.1  ARK277.1  ARK288.1  ARK301.1  ARK325.1  ARK335.1  ARK337.1  ARK347.1  ARK356.1  ARK367.1  ARK403.1  ARK404.1  ARK409.1  ARK411.1  ARK414.1  ARK416.1  ARK417.1  ARK435.1  ASJ075.1  ASM524.1  ASM531.1  ASM665.1  ASM687.1  ASM692.1  ASM765.1  ASR938.1  ASV349.1  ASV352.1  ASV357.1  ASV360.1  ASV370.1  ASV374.1  ASV412.1  ASV433.1  ASV462.1  ASV463.1  ASV603.1  ASV691.1  ASV693.1  ASV739.1  ASV746.1  ASV761.1  ASV844.1  ASV854.1  ASV875.1  ASV895.1  ASV942.1  ASV957.1  ASV959.1  ASW057.1  ASW084.1  YW0086.1  ZS0578.1  ZX5726.1  ZX5848.1",
      "more_code_info": ""
    }
  ]
}