      "status": "Terminated",
      "city": "Waukesha",
      "state": "WI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "65958",
      "recalling_firm": "GE Healthcare, LLC",
      "address_1": "3000 N Grandview Blvd",
      "address_2": "",
      "postal_code": "53188-1615",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution-USA (nationwide) including the states of AZ, CA, CO, FL, GA, IL, IA, IS, LA, MA, MO, NH, NM, NY, OH, OK PA, SC, TX, UT, VA, WA, and WI, and the countries of UNITED KINGDOM, DENMARK, CHILE, ALGERIA, ARGENTINA, AUSTRALIA, AUSTRIA, AZERBAIJAN, BAHRAIN, BELGIUM, CANADA, CZECH REPUBLIC, ECUADOR, EGYPT, ETHIOPIA, FINLAND, FRANCE, FRANCE GUIANA, GERMANY, GUADELOUPE, GUATEMALA, GUYANA, INDIA, INDONESIA, IRAQ, ITALY, JAPAN, JORDAN, KAZAHSTAN, KOREA, KUWAIT, LATVIA, MYANMA, NETHERLANDS, NEW ZEALAND, NORWAY, PAKISTAN, POLAND, ROMANIA, RUSSIA, SAUDI ARABIA, SOUTH AFRICA, SPAIN, SWEDEN SWITZERLAND, TAIWAN, THAILAND, TURKEY, UKRAINE, UZBEKISTAN, and VIETNAM.",
      "recall_number": "Z-0023-2014",
      "product_description": "GE, Optima CT660.    Intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data.",
      "product_quantity": "380 (52 US & 328 OUS)",
      "reason_for_recall": "GE Healthcare has recently become aware of a potential safety issue related with the DMPR function and Screen Saved Images on your scanner which involves Optima CT660, Brivo CT385, Discovery PET/CT 610, Discovery PET/CT 710. The DMPR and Screen Save functionality may possibly generate the same  Series UID. It has been seen that some PACS may combine series under a patient using only the Series UID",
      "recall_initiation_date": "20130326",
      "center_classification_date": "20131017",
      "termination_date": "20131205",
      "report_date": "20131023",
      "code_info": "Mfg Lot or Serial #  00000426724CN1  00000429656CN2  00000429809CN7  00000430378CN0  00000430460CN6  00000430480CN4  00000430489CN5  00000430646CN0  00000430650CN2  00000430651CN0  00000430737CN7  00000430795CN5  00000430813CN6  00000430951CN4  00000431016CN5  00000431035CN5  00000431141CN1  00000431269CN0  00000431307CN8  00000431350CN8  00000431506CN5  00000425485CN0  00000424498CN4  00000427441CN1  00000428100CN2  00000429970CN7  00000430427CN5  00000431140CN3  00000431442CN3  00000425128CN6  00000425032CN0  00000428566CN4  00000431662CN6  00000431604CN8  00000431721CN0  00000431708CN7  00000428762CN9  00000428780CN1  00000429600CN0  00000429658CN8  00000429818CN8  00000429877CN4  00000430132CN1  00000430186CN7  00000430264CN2  00000430282CN4  00000430296CN4  00000043742YC6  00000043902YC6  00000045606YC1  00000046476YC8  00000429917CN8  00000037723YC4  00000043482YC9  00000046473YC5  00000042457YC2  00000326336HM5  00000026193YC3  00000328747HM1  00000304641HM4  00000304751HM1  00000305380HM8  00000305871HM6  00000308319HM3  00000310013HM8  00000313520HM9  00000320004HM5  00000322646HM1  00000297489HM7  00000302022HM9  00000318966HM9  00000322993HM7  00000040368YC3  00000327365HM3  00000040081YC2  00000041147YC0  00000042968YC8  00000044150YC1  00000044861YC3  00000045907YC3  00000047740YC6  00000044399YC4  00000045314YC2  00000328079HM9  00000325768HM0  00000027396YC1  00000033407YC8  00000325766HM4  00000299338HM4  00000295308HM1  00000030752YC0  00000312781HM8  00000319291HM1  00000319292HM9  00000322992HM9  00000325769HM8  00000310566HM5  00000311079HM8  00000042014YC1  00000048323YC0  00000048013YC7  00000048014YC5  00000048315YC6  00000048317YC2  00000039804YC0  00000040083YC8  00000041145YC4  00000041828YC5  00000042806YC0  00000042965YC4  00000043220YC3  00000043904YC2  00000044524YC7  00000044858YC9  00000045315YC9  00000045317YC5  00000045603YC8  00000043483YC7  00000044862YC1  00000045316YC7  00000045607YC9  00000045909YC9  00000046475YC0  00000046726YC6  00000046727YC4  00000047521YC0  00000296966HM5  00000038151YC7  00000311081HM4  00000311879HM1  00000328746HM3  00000310567HM3  00000316768HM1  00000316770HM7  00000318137HM7  00000315021HM6  00000318138HM5  00000318965HM1  00000042451YC5  00000278510HM3  00000322643HM8  00000043480YC3  00000046474YC3  00000047058YC3  00000040616YC5  00000310612HM7  00000311876HM7  00000325426HM5  00000047631YC7  00000047738YC0  00000327465HM1  00000327467HM7  00000037908YC1  00000042011YC7  00000042967YC0  00000042969YC6  00000044857YC1  00000044859YC7  00000046276YC2  00000048318YC0  00000045064YC3  00000045318YC3  00000044398YC6  00000027886YC1  00000034385YC5  00000031032YC6  00000026850YC8  00000037592YC3  00000038154YC1  00000039342YC1  00000039808YC1  00000043222YC9  00000047061YC7  00000047520YC2  00000048010YC3  00000048316YC4  00000048321YC4  00000048526YC8  00000048625YC8  00000048320YC6  00000037593YC1  00000037909YC9  00000038153YC3  00000038739YC9  00000039000YC5  00000039001YC3  00000039002YC1  00000039343YC9  00000039344YC7  00000039601YC0  00000039806YC5  00000039807YC3  00000039809YC9  00000040366YC7  00000040367YC5  00000040819YC5  00000040820YC3  00000040821YC1  00000040822YC9  00000041143YC9  00000041144YC7  00000042010YC9  00000041148YC8  00000042012YC5  00000043479YC5  00000042449YC9  00000042453YC1  00000042454YC9  00000042456YC4  00000042966YC2  00000043223YC7  00000043221YC1  00000044149YC3  00000045312YC6  00000045910YC7  00000045912YC3  00000046275YC4  00000045911YC5  00000046471YC9  00000046472YC7  00000046724YC1  00000047059YC1  00000047327YC2  00000047329YC8  00000047518YC6  00000047519YC4  00000321688HM4  00000321949HM0  00000320786HM7  00000320787HM5  00000321689HM2  00000321950HM8  00000327464HM4  00000307765HM8  00000308946HM3  00000308948HM9  00000311080HM6  00000316555HM2  00000318136HM9  00000296969HM9  00000297490HM5  00000301345HM5  00000302020HM3  00000296967HM3  00000301348HM9  00000302023HM7  00000042455YC6  00000042805YC2  00000042964YC7  00000047739YC8  00000048007YC9  00000048015YC2  00000038737YC3  00000040369YC1  00000040617YC3  00000040817YC9  00000041146YC2  00000042452YC3  00000046725YC8  00000040614YC0  00000040615YC7  00000047326YC4  00000047328YC0  00000048009YC5  00000047323YC1  00000047989YC9  00000047990YC7  00000048006YC1  00000039805YC7  00000042013YC3  00000043739YC2  00000043903YC4  00000043741YC8  00000045605YC3  00000040816YC1  00000042448YC1  00000044151YC9  00000044642YC7  00000045604YC6  00000045908YC1  00000047060YC9  00000316556HM0  00000431438CN1  00000308949HM7  00000321951HM6  00000280831HM9  00000280832HM7    00000327466HM9  00000305381HM6  00000315022HM4  00000301347HM1  00000327366HM1  00000034263YC4  00000307766HM6  00000321952HM4  00000296970HM7  00000301346HM3  00000326337HM3  00000326862HM0  00000327463HM6  00000033194YC2  00000328078HM1  00000041535YC6  00000041536YC4  00000041537YC2  00000304638HM0  00000304639HM8  00000304750HM3  00000305868HM2  00000305870HM8  00000306523HM2  00000306524HM0  00000306945HM7  00000306946HM5  00000306947HM3  00000307767HM4  00000308320HM1  00000308322HM7  00000311078HM0  00000311877HM5  00000311878HM3  00000312780HM0  00000312782HM6  00000298210HM6  00000302021HM1  00000305867HM4  00000304640HM6  00000308947HM1  00000310014HM6  00000317418HM2  00000034384YC8  00000315827HM6  00000045608YC7  00000044640YC1  00000430273CN3  00000305869HM0  00000430875CN5  00000324510HM7  00000329532HM6  00000312779HM2  00000039003YC9  00000038738YC1  00000044152YC7  00000308321HM9  00000310568HM1  00000034790YC6  00000031572YC1  00000293945HM2  00000321953HM2  00000322642HM0  00000325265HM7  00000325266HM5  00000037282YC1  00000038740YC7  00000040818YC7  00000041827YC7  00000042804YC5  00000043481YC1  00000043740YC0  00000044526YC2  00000044638YC5  00000044639YC3  00000044641YC9  00000044643YC5  00000044860YC5  00000045065YC0  00000315828HM4  00000315829HM2  00000316790HM5  00000296971HM5  00000044644YC3  00000045066YC8  00000316769HM9  00000320005HM2  00000296968HM1  00000047630YC9  00000048012YC9  00000041829YC3  00000042450YC7",
      "more_code_info": ""
    }
  ]
}