  "results": [
    {
      "status": "Terminated",
      "city": "Malvern",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "72091",
      "recalling_firm": "Siemens Medical Solutions USA, Inc",
      "address_1": "40 Liberty Blvd",
      "address_2": "N/A",
      "postal_code": "19355-1418",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide Distribution.",
      "recall_number": "Z-0020-2016",
      "product_description": "SOMATOM Definition Flash; intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from either the same axial plane taken at different angles or spiral planes taken at different angles.",
      "product_quantity": "113 total",
      "reason_for_recall": "software bug issues for SW-Version VA48A_SP0. The following safety issues were resolved: 1)  Correction to improve visual warning and error indication son the gantry display. 2) Correction to improve acquisition data in order to optimize image quality. 3) Correction to improve robustness and general system behavior in some exception handling procedures. 4) Correction to improve auto post processin",
      "recall_initiation_date": "20150814",
      "center_classification_date": "20151005",
      "termination_date": "20161205",
      "report_date": "20151014",
      "code_info": "Model#  10430603  with Serial numbers  73476  73586  73544  73648  73636  73621  73533  73423  73424  73520  73493  73459  74191  73487  73492  73494  73497  73013  73657  73535  73667  73490  73513  74192  73697  73635  73471  73405  73600  73558  73591  73546  73478  73504  73413  73674  73445  73547  73420  73430"
    }
  ]
}