  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Louvain La Neuve",
      "state": "N/A",
      "country": "Belgium",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "68846",
      "recalling_firm": "Ion Beam Applications S.A.",
      "address_1": "Chemin du Cyclotron, 3",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution-USA (nationwide) including the states of PA, NJ, IL, WA, and TN, and the countries of Germany, Italy, and Czech Republic.",
      "recall_number": "Z-0020-2015",
      "product_description": "Proteus 235     Proton Therapy System for cancer treatment.",
      "product_quantity": "8",
      "reason_for_recall": "While the Proteus 235 is in the pencil beam scanning (PBS) treatment mode, on rare occasions, Low Level Radio Frequency (LLRF) pulsed dee voltage (i.e., voltage applied between two electrodes in the cyclotron to accelerate protons) may generate a \u001chot spot\u001d on a pencil beam position within a tumor, generating a dose that might exceed the prescribed dose on a single voxel.",
      "recall_initiation_date": "20140327",
      "center_classification_date": "20141029",
      "termination_date": "20141210",
      "report_date": "20141105",
      "code_info": "PAT.107, PAT.109, PAT.112, PAT.113, PAT.114,  PAT.115, PAT.116, PAT.119.",
      "more_code_info": ""
    }
  ]
}