{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Andover",
      "address_1": "3000 Minuteman Rd",
      "reason_for_recall": "During a bolus tracking procedure, no images were generated when the scan was completed, and the raw data file was not available for offline reconstruction. There is a discrepancy between the calculated reconstruction length and the actual scan length that results in the inability to reconstruct raw data. The operator may choose to rescan the patient.",
      "address_2": "",
      "product_quantity": "US-  606",
      "code_info": "Software version 4.16",
      "center_classification_date": "20171013",
      "distribution_pattern": "USA (nationwide) Distribution",
      "state": "MA",
      "product_description": "Philips Healthcare Brilliance iCT   X-Ray, Tomography, Computed      These computed tomography x-ray systems are intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes.",
      "report_date": "20171025",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Philips Electronics North  America Corporation",
      "recall_number": "Z-0019-2018",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "78144",
      "termination_date": "20200226",
      "more_code_info": "",
      "recall_initiation_date": "20170118",
      "postal_code": "01810-1032",
      "voluntary_mandated": "FDA Mandated",
      "status": "Terminated"
    }
  ]
}