{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Caesarea",
      "state": "",
      "country": "Israel",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "88576",
      "recalling_firm": "Paltop Advanced Dental Solutions Ltd",
      "address_1": "5, Ha-Shita",
      "address_2": "",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution. No foreign distribution from US.",
      "recall_number": "Z-0016-2022",
      "product_description": "Scan Abutment SU, Cat. No. 30-70104  - Product Usage: used to take an impression with an intra oral scanner or optical scanner to provide a digital registration of an implant or abutment platform. The Paltop Advanced Dental Solutions Implant System is indicated for use in surgical and restorative applications for placement in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, in order to restore the patient s chewing function.",
      "product_quantity": "16",
      "reason_for_recall": "Some devices are outside tolerances and showed a maximal deviation of 2 degrees in the orientation between the peek head and titanium base, and an incongruence between the size engraved on the peek has been identified. Deviations may delay the final restoration.",
      "recall_initiation_date": "20210726",
      "center_classification_date": "20211004",
      "termination_date": "20230921",
      "report_date": "20211013",
      "code_info": "All lots manufactured prior to July 21, 2021.",
      "more_code_info": ""
    }
  ]
}