  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Malvern",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "65161",
      "recalling_firm": "Siemens Medical Solutions USA, Inc",
      "address_1": "51 Valley Stream Pkwy",
      "address_2": "N/A",
      "postal_code": "19355",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide Distribution",
      "recall_number": "Z-0016-2014",
      "product_description": "Siemens RAD Fluoro Uro Systems - Luminos dRF, Ysio and Uroskop Omnia with software versions VB10D, VB10F, VB10C and VB10E.     Fluroscopic X-Ray Systems",
      "product_quantity": "23",
      "reason_for_recall": "Siemens has discovered through product monitoring a potential malfunction when using system operating with software versions VB10D, VB10F, VB10C and VB10.  During a RAD examination the possibility exists that radiation may continue longer than expected and may result in unnecessary dose applied during an acquisition.  Additionally, it is not possible for the user to stop the radiation manually; ho",
      "recall_initiation_date": "20130506",
      "center_classification_date": "20131101",
      "termination_date": "20141106",
      "report_date": "20131113",
      "code_info": "Material 10094910, with serial numbers 1036, 1045, 1025, 1040, 1041, 1032, 1047 and 1035; Material 10281013, with serial numbers 21221, 21223, 21213, 21089, 21098, 21163, 21171, 21194, 21212, 21186, 21188, 21167, 21187 and 21189; Material 10281163, with serial number 31114."
    }
  ]
}