  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Lulea",
      "state": "",
      "country": "Sweden",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "78130",
      "recalling_firm": "Handicare AB",
      "address_1": "Maskinvagen 17",
      "address_2": "",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "OH, PA, MD",
      "recall_number": "Z-0014-2018",
      "product_description": "Carina Mobile Lifts, Models:    a) Carina350EM, REF 6060011  b) Carina350EML, REF 60600012  c) Carina350EE, REF 60600009",
      "product_quantity": "9 units",
      "reason_for_recall": "A limited amount of Carina mobile lifts have been delivered with a too small locking pin. If a locking pin  smaller than the correct size is mounted on the Carina mobile lift, the locking pin is likely to cause  increased wear on the plastic bushings. In case of continued use, these components may cause wear on the metal of the piston rod eye.",
      "recall_initiation_date": "20160728",
      "center_classification_date": "20171011",
      "termination_date": "20181217",
      "report_date": "20171018",
      "code_info": "a) Carina350EM, REF 6060011, Serial numbers:  661102000 - 661102160  b) Carina350EML, REF 60600012, Serial numbers:  661200500 - 661200584  c) Carina350EE, REF 60600009, Serial numbers:  660900500 -660900571",
      "more_code_info": ""
    }
  ]
}