      "status": "Terminated",
      "city": "Racine",
      "state": "WI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "83435",
      "recalling_firm": "Inter-Med Llc",
      "address_1": "2200 Northwestern Ave",
      "address_2": "N/A",
      "postal_code": "53404-2500",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US, South Africa, Switzerland, Poland, Italy, Singapore, Australia",
      "recall_number": "Z-0013-2020",
      "product_description": "Vista Dental Products Sodium Hypochlorite - 3%, 10 - 12cc Pre-filled Syringes, REF 502315",
      "product_quantity": "582 units",
      "reason_for_recall": "The affected lots may be packaged in syringes which include an incorrect bung.  The incorrect bung is made of a latex-free isoprene rubber material.  After prolonged storage and exposure to contained solutions, the rubber bung may degrade due to chemical interaction with the bleach solution.",
      "recall_initiation_date": "20190627",
      "center_classification_date": "20191001",
      "termination_date": "20210810",
      "report_date": "20191009",
      "code_info": "Lot/Work Order Numbers:  2018-1349  2018-1587  2018-1853  20182107, UDI (01)10818207020427  20182303, UDI (01)10818207020427  20182584, UDI (01)10818207020427  20182666, UDI (01)10818207020427  20190199, UDI (01)10818207020427  20190433, UDI (01)10818207020427"
    }
  ]
}