  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Racine",
      "state": "WI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "83435",
      "recalling_firm": "Inter-Med Llc",
      "address_1": "2200 Northwestern Ave",
      "address_2": "N/A",
      "postal_code": "53404-2500",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US, South Africa, Switzerland, Poland, Italy, Singapore, Australia",
      "recall_number": "Z-0012-2020",
      "product_description": "Vista Dental Products Sodium Hypochlorite - 6%, 10 - 12cc Pre-filled Syringes, REF 502300",
      "product_quantity": "1992 units",
      "reason_for_recall": "The affected lots may be packaged in syringes which include an incorrect bung.  The incorrect bung is made of a latex-free isoprene rubber material.  After prolonged storage and exposure to contained solutions, the rubber bung may degrade due to chemical interaction with the bleach solution.",
      "recall_initiation_date": "20190627",
      "center_classification_date": "20191001",
      "termination_date": "20210810",
      "report_date": "20191009",
      "code_info": "Lot/Work Order Numbers:  2018-1348  2018-1585  2018-1586  2018-1852  20182106, UDI (01)10818207020465  20182302, UDI (01)10818207020465  20182458, UDI (01)10818207020465  20190725, UDI (01)10818207020465  20190835, UDI (01)10818207020465"
    }
  ]
}