  "results": [
    {
      "status": "Terminated",
      "city": "Waukesha",
      "state": "WI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "66277",
      "recalling_firm": "GE Healthcare, LLC",
      "address_1": "3000 N Grandview Blvd",
      "address_2": "",
      "postal_code": "53188-1615",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - USA (nationwided) including DC., and the countries of  AUSTRALIA, CANADA, CHILE, GERMANY, CHANA, GUATEMALA, ICELAND, ISRAEL, ITALY, JAMAICA, JAPAN, RUSIAN FEDERATION, UNITED ARAB EMIRATES and UNITED KINGDOM.",
      "recall_number": "Z-0012-2014",
      "product_description": "GE Healthcare, Precision 500D, Legacy, RFX, and SFX X.     This table is intended for use in a diagnostic X-ray system to support patients during general purpose radiological procedures in the horizontal, vertical and trendelenburg positions. - The Expedio 500D is designed to perform radiographic and fluoroscopic x-ray examinations. - The Precision 500D R&F X-ray System is intended to be used as a universal diagnostic imaging system for radiographic and fluoroscopic examinations, including general R&F and pediatric examinations. The Precision 500D R&F X-ray Systems are intended for use by a qualified/trained doctor or technician on both adult and pediatric subjects for taking diagnostic radiographic and fluoroscopic exposures of the whole body, skull, spinal column, chest, abdomen, extremities, and other body parts. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position.",
      "product_quantity": "810 (783 US, 27 OUS)",
      "reason_for_recall": "GE Healthcare has recently become aware of a potential safety issue involving the spotfilm device power assist handle.    There is a potential for the internal bolts on the power assist handle to be loose or to loosen and/or fall out. This issue, if present, can result in a loss of calibration and offset of the power assist handle and/or, under extreme condition, sticking of the handle in the ON p",
      "recall_initiation_date": "20130830",
      "center_classification_date": "20131017",
      "termination_date": "20140522",
      "report_date": "20131023",
      "code_info": "Serial #                           System ID   00000976624WK7 202884RF2410  00000977748WK3 202884RF2411  00000955594WK7 205877LRF2  00000965983WK0 207373RF1  00000969026WK4 215748LEG1  00000960066WK9 219416ADV  00000975683WK4 228809SR2  00000966338WK6 256235ORF1  00000967770WK9 301896RF  00000961102WK1 302654RF  00000972383WK4 303440BCHRM4  00000964398WK2 303730R3  00000958345WK1 304256RX1A  00000978812WK6 361552LEGD1  00000970377WK8 404321LEG2  00000973016WK9 408259RF1  00000963305WK8 408977RF5  00000963524WK4 414266RM03  00000963118WK5 417533BR2  00000969537WK0 507372LEG  00000959491WK2 518382SRM2  00000976668WK4 540862ARF6  00000966773WK4 561468RM4  00000972380WK0 606285LEGACY  00000966622WK3 619421RF1  00000957841WK0 708216GIRF1  00000983335WK1 713527PRF1  00000964482WK4 775738RM2  00000971676WK2 804289HDF4  00000978181WK6 804747DHP1  00000994573WK4 815740RF1  00000961772WK1 817472LEG1  00000963293WK6 850747ER1  00000975653WK7 856778RF2  00000965747WK9 864295GRRM4  00000973832WK9 904508LEGACY  00000962333WK1 904819RF1  00000970009WK7 907276RM3  00000953103WK9 912633MR3  00000978182WK4 915685ADV3  00000978314WK3 916784LEG  00000977043WK9 936321LEG1  00000963738WK0 941798RM2  00000982588WK6 956388RF1  00000966348WK5 972566RM11  00000962460WK2 RADNETRF130  00000963491WK6 RADNETRF137  00000978497WK6 651232RF4  00000983536WK4 703689RM03  00000011284GP7 210297BMCRF11  00000011281GP3 210297SEBRF3  00000985785WK5 281348P5001  00000014340GP4 308398RF  00000014337GP0 412641RF1  00000984897WK9 435637RM4  00000013713GP3 561863RF1  00000013860GP2 661633XR2  00000014347GP9 713636LBJ7  00000988380WK2 717272RF2  00000011676GP4 720854RF1  00000012513GP8 727823RM3  00000013045GP0 770793P500D  00000012418GP0 781744RF3  00000012521GP1 816348BRF1  00000009196GP7 817255NHPD1  00000013285GP2 828298RF4  00000011737GP4 850325RF4  00000013257GP1 850494WF4  00000014350GP3 856325P500  00000989504WK6 913676SRM4  00000013271GP2 956350VRF3  00000990179WK4 PC7479XR03  00000012826GP4 00105FLU01  00001009797WK0 303825P500  00000009180M31 202745PRE2  00000022650M36 207498P500  00000001835M38 208227ROOM6  00000005813M31 208455RF1  00000002639M33 208678500D  00000022309M39 209536P500  00000015302M33 215710AURORA2  00000002983M35 228377RF10  00000003052M38 228377RF9  00000021223M33 239343RF4  00000021093M30 239432RF11  00000020302M36 240566DF1  00000007194M34 252522LRF1  00000005274M36 256494P5R1  00000992217WK0 260373RF4  00000022744M37 270756P500  00000019347M34 281477MWBP500D  00000008963M31 281637SLSP500D  00000997107WK8 304647PRE  00000022017M38 314251MR2  00000007242M31 314996BW2  00000004648M32 318330RF2  00000008860M39 318473RD4  00000004529M34 337521RF2  00000990537WK3 404325EFL1  00000008599M33 404712GIPR  00000004863M37 404785WBP500D  00000002984M33 406863P500  00000017129M38 412623P500D1  00001001061WK9 414649P500  00000995345WK6 423926RF1  00000017643M38 434654RM4  00000011245M38 469624P500D1  00000019086M38 469764P5001  00000008136M34 484884P500  00000993344WK1 505521RF1  00000020464M34 505609P500  00001004238WK0 509473VR2  00000014498M30 512341RRRF1  00000021219M31 517543P500D  00001003383WK5 540829RF1  00000004233M33 561791RM3  00000011944M36 570824P500  00000004229M31 573468MBS1  00000993783WK0 573756P500  00000005270M34 574523EG2  00000006560M37 574523EG4  00000998427WK9 601288PRF6  00000014272M39 605867P500  00000990547WK2 606678RF2  00001003367WK8 607432GE1  00001003379WK3 607547RF1  00001004242WK2 608745P500  00000021027M38 608775RF500  00000995340WK7 609599RFP4  00000012810M38 609835LMCP500  00001000389WK5 615342P500D3  00000999391WK6 615342P500D4  00000005650M37 617499RF4  00000995344WK9 617732ENDO  00000018235M32 623433RF  00000002217M38 678454P500D1  00000008661M31 702653RF03  00000997661WK4 713359RF3  00000013216M37 713441MOP500DP  00000999916WK0 715847P500  00000018057M30 717270RF2  00000003982M36 717544HCRF6  00001000995WK9 719560RF1  00000002577M35 732840P5D  00000022957M35 740633P500  00000021220M39 757398RM4  00000006859M33 770228RM3  00000021094M38 770400P500D  00000020193M39 785458P500  00000006330M35 802524500D  00000019490M32 812450DRF10  00000998422WK0 812858PRF  00000014501M31 813571RM4  00000020937M39 814723P500  00000001967M39 815741P500  00001004951WK8 815756RM2  00000990278WK4 817255NHRM1  00000020187M31 817782P500  00001007260WK1 864366ARF2  00001003364WK5 864675SFWR2  00000014567M32 910671SERF5  00000008311M33 910907P500RM3  00000021357M39 910938CHP500  00000008197M36 928763WRF1  00001005392WK4 931766RF1  00000023455M39 941629P500  00000010128M37 951CD500  00000022652M32 978741SRF3  00000005653M31 978774RF1  00000991650WK3 JAC3E63  00000020191M33 RADNETRF1284  00000020576M35 RADNETRF1286  00000018006M37 RADNETRF182  00000011818M32 RM7GEP500  00000005493M32 SA2135RX02  00001002438WK8 416813PR500FL1  00000006690M32 IL1000XR06  00000007585M33 00044FLU01  00000007385M38 00065FLU01  00001023876WK4 925947RM2  00001027482WK7 504873TR4P  000000102558K2 337470RF2        00001024517WK3 303RRP5001       00001024518WK1 712362P500       00001024576WK9 815968P500E      00001024577WK7 210617RF1        00001024578WK5 215710AURORA2    00001024725WK2 217443URF5       00001024726WK0 402294EBRF5      00001024741WK9 504837P500       00001024742WK7 501286PRE1       00001024743WK5 225NORF          00001024744WK3 920262P500       00001024874WK8 319268P500       00001024888WK8 937440RM1        00001024889WK6 515295P500       00001025002WK5 541947RF1        00001025003WK3 818364R3         00001025004WK1 606666P500       00001025011WK6 814452P500       00001025012WK4 217732LRF        00001025013WK2 323361R8         00001025055WK3 815431OP500      00001025253WK4 323361RF2        00001025254WK2 216778PREC1      00001025255WK9 509488P500       00001025256WK7 512LRMCRF        00001025366WK4 319272P500       00001025367WK2 ORMCP500         00001025368WK0 501268PRE2       00001025583WK4 337491P500       00001025585WK9 504896P500       00001025586WK7 715532RF         00001025588WK3 812268P500       00001025589WK1 Not Available  00001025590WK9 970870P5001      00001025591WK7 281274WHSLP500D  00001025592WK5 337470RF1        00001025800WK2 630275G04        00001025806WK9 563568P500       00001026022WK2 559459500D       00001026026WK3 970247P500B      00001026240WK0 309691PRF1A      00001026241WK8 631654P500D      00001026242WK6 510814RM4        00001026243WK4 509826MVP500     00001026245WK9 256235P500       00001026246WK7 301765P500       00001026247WK5 508856ORP500     00001026248WK3 918392P500       00001026424WK0 435722P500       00001026425WK7 978368P500       00001026459WK6 806364P500       00001026465WK3 402336P500       00001026466WK1 608364P5002      00001026636WK9 508793MP500      00001026637WK7 318932P500       00001026638WK5 603524500D       00001026639WK3 760873RF2        00001026704WK5 215427DIGRF      00001026705WK2 919RALCPP500M    00001026706WK0 405604P500       00001026970WK2 973579RM2        00001026971WK0 919RALCPP500W    00001026972WK8 417347P500       00001026973WK6 216778PREC2      00001027063WK5 312CMP500D2      00001027079WK1 252847GE3        00001027080WK9 312CMP500D1      00001027081WK7 740845PRE        00001027082WK5 480324P500D      00001027083WK3 918434P500       00001027084WK1 765472P500       00001027085WK8 217333P500D      00001027086WK6 623433RF         00001027087WK4 479549P500       00001027424WK9 402759FCP500     00001027425WK6 305NSHRM5        00001027426WK4 305NSHRM6        00001027483WK5 504896R2P500     00001027484WK3 513870PRE1       00001027485WK0 608258P5002      00001027587WK3 304258RF         00001027588WK1 601798P500       00001027589WK9 859745P500       00001027590WK7 318448RF1        00001027591WK5 913491P500D      00001027592WK3 989246P500       00001027593WK1 614722PRE        00001027594WK9 785243P500       00001027862WK0 715623P500       00001027863WK8 918775P500       00001027864WK6 510372RFP5       00001027866WK1 760957P500D      00001027867WK9 919731WMRF5      00001027868WK7 785458P500       00001028157WK4 816455P500D      00001028158WK2 828213RCRF       00001028159WK0 256486RF1        00001028161WK6 240566DF1        00001028162WK4 FHALTP500        00001028163WK2 828456P500       00001028422WK2 601554P5001      00001028423WK0 336846AMRF       00001028424WK8 404466P500       00001028425WK5 404265P500D1     00001028426WK3 248652P500       00001028632WK6 301295P500A      00001028633WK4 918323P500       00001028634WK2 641228P500       00001028635WK9 775982RF7        00001028822WK3 303673P500D      00001028823WK1 509786P500       00001028824WK9 816554P500D      00001029004WK7 405744OSUVET     00001029005WK4 731644P500       00001029006WK2 573231HCP500     00001029007WK0 217876DRF5A      00001029008WK8 309344MRF1B      00001029134WK2 229353TR3        00001029135WK9 910938CHP500     00001029136WK7 217464P500       00001029137WK5 757398RM4        00001029138WK3 908277P500D      00001029139WK1 240566RHRF1      00001029140WK9 773THOREKP500    00001029226WK6 423877P500       00001029227WK4 916UCD500D       00001029228WK2 865316RF1        00001029425WK4 270725P500       00001029426WK2 314251MR2        00001029427WK0 912681PREC       00001029428WK8 229353TR2        00001029429WK6 434799P500D      00001029430WK4 309467EP500      00001029431WK2 907443P500       00001029432WK0 816795P500D      00001029713WK3 270887RF         00001029714WK1 618798GWP500     00001029715WK8 229391WC1        00001029716WK6 516796P500       00001029717WK4 706554RF1        00001029794WK3 WILMVA500D       00001029837WK0 281540NECCRF     00001029838WK8 479684P500       00001029839WK6 702304RF1        00001030119WK0 530749P500D1     00001030120WK8 423553P500       00001030121WK6 270756P500       00001030122WK4 Not Available  00001030123WK2 301891P500       00001030234WK7 217222QP500      00001030299WK0 269945P500       00001030300WK6 910862P500D      00001030301WK4 Not Available  00001030302WK2 918286P500       00001030303WK0 320845P500       00001030458WK2 309944HH500      00001030459WK0 817865FPSP500    00001030460WK8 615329PP500      00001030461WK6 610284P500       00001030462WK4 937641RMF        00001030580WK3 662OBRF          00001030581WK1 402484AP500      00001030582WK9 229276CRF1       00001030583WK7 Not Available  00001030584WK5 601663P500       00001030634WK8 276988P500       00001030696WK7 361580P500       00001030697WK5 504903P500       00001030699WK1 4083516  00001030700WK7 909796P500       00001030701WK5 269637P500       00001030895WK5 801657P500       00001030896WK3 Not Available  00001030897WK1 808YMP500D       00001030898WK9 303DCIP500       00001030981WK3 Not Available  00001030982WK1 309672P500D2     00001030983WK9 MIMCP500D        00001031089WK4 Not Available  00001031090WK2 Not Available  00001031091WK0 Not Available  00001031092WK8 Not Available  00001031257WK7 Not Available  00001031442WK5 Not Available  00001025587WK5 CANFPRECRX1  00001027088WK2 PC0362XR12       00001026426WK5 GH1042RX02  00001027865WK3 XR1027865WK3  00001030698WK3 XR23354M34  00001028160WK8 YX1285           00001025369WK8 AE1021RX04       00001027096WK5 412372P500D  00001025507WK3 561955RF  00000102474WK4 804287P500       00001024519WK9 203739P500       00001024520WK7 717272RF3        00001024572WK8 405271CCP500     00001024573WK6 650493VARF2      00001024574WK4 281MWP500        00001024575WK1 504842RM3        00001024676WK7 707464RF2        00001024677WK5 Not Available  00001024723WK7 Not Available  00001024724WK5 503623RF1        00001024745WK0 401782SCP500     00001024746WK8 516663RMP5       00001024747WK6 Not Available  00001024748WK4 Not Available  00001024878WK9 605997P500       00001024879WK7 Not Available  00001024880WK5 605747P500       00001024881WK3 605867P500       00001024882WK1 Not Available  00001025005WK8 607547RF2        00001025006WK6 RADNETRF1268     00001025007WK4 757312RM4        00001025008WK2 RADNETRF1269     00001025009WK0 813571RM4        00001025010WK8 RADNETRF1270     00001025049WK6 LLPM500D         00001025050WK4 207564P500       00001025051WK2 315338RF         00001025052WK0 931520P500       00001025054WK6 630275G05        00001025258WK3 985646RF4        00001025260WK9 912367AURORA     00001025261WK7 907PIC2P500      00001025371WK4 918224P500D1     00001025372WK2 818409RF2        00001025373WK0 954JDCHRF1       00001025374WK8 210BMCRF1        00001025375WK5 270251JPRF2      00001025508WK1 508862CCR3       00001025509WK9 212420ENDO1      00001025510WK7 859498P500       00001025511WK5 954JDCHRF2       00001025573WK5 918744P5RM3      00001025574WK3 205408SVRF       00001025575WK0 412623P500D1     00001025576WK8 Not Available  00001025577WK6 407518PRE        00001025578WK4 Not Available  00001025579WK2 719285P500       00001025582WK6 434447CMRM4      00001025876WK2 229P500RF1       00001025877WK0 918744P5RM4      00001025878WK8 606439P500       00001025879WK6 954265ME1        00001025880WK4 518926FLRM4      00001025886WK1 606633P500       00001025887WK9 606237PRE        00001025888WK7 203739RM4        00001025889WK5 269789P500       00001025890WK3 229P500RF2       00001025977WK8 Not Available  00001025978WK6 210MTRF1         00001025988WK5 701252P500       00001026027WK1 434654RM3        00001026028WK9 903CSMP500       00001026029WK7 Not Available  00001026030WK5 559459500D       00001026031WK3 978683LGRF5      00001026146WK9 434654RM4        00001026147WK7 907442KP500      00001026148WK5 704829P500       00001026250WK9 504632P500       00001026251WK7 585596P500       00001026252WK5 208463P500D      00001026253WK3 520GV500D        00001026254WK1 713873P500D      00001026255WK8 727848P5D2       00001026256WK6 727848P5D1       00001026257WK4 936539P500D2     00001026258WK2 Not Available  00001026259WK0 Not Available  00001026260WK8 RADNETRF182      00001026264WK0 781624HSSHRF     00001026467WK9 202269P5003      00001026468WK7 717270RF2        00001026469WK5 409938P500       00001026470WK3 360754ICRF       00001026471WK1 COR481360        00001026472WK9 903GS503         00001026473WK7 406293P500       00001026474WK5 SLRP500FP1       00001026640WK1 541382RM3        00001026641WK9 540SHENP500      00001026642WK7 Not Available  00001026643WK5 561745RM4        00001026709WK4 412623P500D2     00001026710WK2 276228P500       00001026711WK0 Not Available  00001026712WK8 229924P500       00001026713WK6 405752RF2        00001026786WK2 301609P500A      00001026787WK0 631376RM3        00001026788WK8 804893RF1        00001026858WK9 205877BPRF1      00001026859WK7 516256P500       00001026860WK5 850747GC3        00001026861WK3 714827P500       00001026862WK1 509575YMRF2      00001026899WK3 478327P500       00001026900WK9 802885P500D      00001026901WK7 618443SP500      00001026974WK4 912754VIDAR1     00001026975WK1 706651D3         00001026976WK9 501202P500SH     00001026977WK7 631376RM2        00001026978WK5 540CULRF2        00001026980WK1 720SAMOBP500     00001027089WK0 276666P500       00001027090WK8 314577RM7        00001027091WK6 307864P500D      00001027092WK4 307684P500       00001027093WK2 716845DRF1       00001027094WK0 360413P500       00001027095WK7 803751MRF5       00001027097WK3 520CICP500       00001027098WK1 608356P500       00001027237WK5 469764P5001      00001027428WK0 469764P5002      00001027430WK6 949452RM1        00001027431WK4 928348RF2        00001027477WK7 229312FP1        00001027478WK5 520CICP500       00001027615WK2 315425A320       00001027616WK0 812450DRF10      00001027617WK8 281477MWBP500D   00001027618WK6 419933RF         00001027619WK4 704834P500       00001027620WK2 912384RM5        00001027622WK8 MERCYP500R1      00001027854WK7 281274P5DGR      00001027855WK4 503717FP1        00001027856WK2 206598UW213      00001027857WK0 205939SVRF       00001027858WK8 805497RM3        00001027872WK9 937382PRE        00001027873WK7 RADNETRF1284     00001027874WK5 817782P500       00001027875WK2 706356TRF        00001028165WK7 724983500D       00001028166WK5 214820RF3        00001028167WK3 Not Available  00001028168WK1 315363P500       00001028169WK9 505609P500       00001028170WK7 713SCP500D       00001028171WK5 541382CRM1       00001028172WK3 713SCP500D2      00001028430WK5 804SMHOP1        00001028431WK3 270932P500       00001028432WK1 RADNETRF1286     00001028433WK9 718226VZRM3      00001028434WK7 412367P500D2     00001028435WK4 949489RM1        00001028436WK2 740383PRE2       00001028437WK0 SLWP500          00001028601WK1 814723P500       00001028602WK9 RADNETRF183      00001028636WK7 2972752  00001028637WK5 510231P500D      00001028638WK3 608775RF500      00001028639WK1 503VAPRF1        00001028817WK3 RADNETRF1283     00001028818WK1 702383RF6        00001028819WK9 718579RM508      00001028820WK7 239432RF11       00001028821WK5 770400P500D      00001028911WK4 561671RF1        00001028999WK9 517543P500D      00001029001WK3 COMFORTR2        00001029002WK1 RADNETRF1287     00001029003WK9 COR481341        00001029144WK1 814624RF2        00001029145WK8 718226PRM10      00001029146WK6 530244P500D      00001029147WK4 313343ERCP       00001029148WK2 404501P500       00001029152WK4 415DAVP500D      00001029153WK2 718226PRM1       00001029231WK6 856355P500       00001029232WK4 856247HWP500     00001029233WK2 631726P500D      00001029234WK0 404686EMRF7      00001029235WK7 Not Available  00001029319WK9 COR481342        00001029320WK7 325670PR7        00001029433WK8 603356P500D      00001029434WK6 215487P500       00001029435WK3 804254RMC        00001029436WK1 541472P500       00001029437WK9 205333FL2        00001029438WK7 COMFORTR3        00001029439WK5 239343RF4        00001029440WK3 MERCYP500R2      00001029558WK2 239432RF14       00001029559WK0 419238P500       00001029560WK8 479587P59        00001029561WK6 441236P500       00001029562WK4 479587P52        00001029708WK3 580379P500       00001029710WK9 205592PRF1       00001029711WK7 209754P500       00001029712WK5 606833P500D      00001029829WK7 937208PRE        00001029830WK5 419228P500       00001029831WK3 205592PRF3       00001029832WK1 620332P500       00001029833WK9 229312MRF1       00001029836WK2 904VACRF         00001030124WK0 229312MRF2       00001030125WK7 231225P500       00001030126WK5 PMCP500          00001030127WK3 RADNETRF1290     00001030128WK1 956632RF2        00001030184WK4 904215RF1        00001030185WK1 408885FL2        00001030186WK9 252847VCH500D    00001030187WK7 816347SLERF      00001030188WK5 209536P500       00001030305WK5 863314P500       00001030306WK3 954587RM4        00001030307WK1 207498P500       00001030308WK9 732294P5D        00001030309WK7 207795P500       00001030310WK5 305HIAP500       00001030415WK2 0004014412A      00001030416WK0 608356P5002      00001030417WK8 270417RF2        00001030463WK2 RADNETRF190      00001030464WK0 830703VVRF1      00001030466WK5 Not Available  00001030467WK3 740633P500       00001030577WK9 618288RM2        00001030578WK7 978741SRF3       00001030579WK5 724772P500D1     00001030704WK9 217383P500A      00001030705WK6 704667RF1        00001030707WK2 501202P500RM2    00001030708WK0 UICGERF1         00001030709WK8 863763RM1        00001030710WK6 478PCHP500       00001030711WK4 Not Available  00001030891WK4 718226VZRM2      00001030892WK2 Not Available  00001030893WK0 505WHRM1         00001030894WK8 408885FL1        00001030985WK4 570621SEPREC     00001030986WK2 941629P500       00001031093WK6 757889RM4        00001031094WK4 937734PRE        00001031095WK1 281420SJM500D1   00001031096WK9 937641RMA        00001031196WK7 910FHHOKE+P500   00001031197WK5 Not Available  00001031198WK3 Not Available  00001031259WK3 Not Available  00001031260WK1 Not Available  00001031312WK0 Not Available  00001031313WK8 Not Available  00001031314WK6 Not Available  00001031315WK3 Not Available  00001031316WK1 Not Available  00010283448WK9 781306RF1        00001027876WK0 910072221  00001027429WK8 752026XR05       00001025262WK5 IE1545RF01  00001028083WK2 A5745526  00001026979WK3 RU3638RX02       00001029000WK5 RU7316RX01  00001029709WK1 RU1279RX03       00001025259WK1 10448FLU03       00001025262WK5 IE1545RF01       00000350013WK9 207647XR1  00000319904WK9 215377RF3  00000330000WK1 580336PRF  00000473802WK7 208227RM2  00000467274WK7 210632RRF  00000440241WK8 316665HR3  00000507540WK3 334277SFX  00000437356WK9 402486MPEX1  00000399552WK9 404589GSF  00000486316WK3 407335RM2  00000459485WK9 410224RF  00000442423WK0 415641ADV  00000420775WK9 601872OR02  00000501994WK8 703776RM2  00000453196WK8 713799TRF  00000481391WK1 804392SCARF1  00000450288WK6 813932SPEC  00000432606WK2 814946RF1  00000516246WK6 919443NDRF1  00000508469WK4 216971R8  00000516440WK5 304257LEG  00000518141WK7 510370RFDG2  00000510617WK4 512476SRM6  00000514169WK2 516968LEG1  00000516439WK7 813681RM3  00000506733WK5 817526MLEG  00000499668WK2 818331SFX  00000512452WK4 847297HF3  00000518883WK4 910249LRF1  00000516878WK6 912633MR1  00000514167WK6 912633MR2  Not Available 519254RM6A  00000530446WK4 209225LEG2  00000521503WK3 315393RF1  00000519118WK4 412578RF9  00000517706WK8 510370RFDG1  00000530677WK4 618253HMRF1  00000523399WK4 718430RM2  00000521501WK7 718430RM3  00000519515WK1 813681SPEC1  00000523405WK9 912751LD  00000551900WK4 228575XR1  00000549174WK1 303778RM6A  00000951422WK5 352351RM2  00000549578WK3 407830RF4  00000546368WK2 414785RM3  00000547501WK7 415923PR5  00000554359WK0 480860RM7  00000554358WK2 480860RM8  00000554360WK8 480860RM9  00000549122WK0 602879RM1  00000552616WK5 603778RM1  00000951884WK6 615769SKYRF3  00000541723WK3 615792RM1  00000553951WK5 618544CMRF1  00000552314WK7 706651D4  00000552929WK2 804541JR2A  00000952157WK6 865693RF3  00000534305WK8 904825LEGACY  00000534304WK1 915595RF4  00000541893WK4 918294SCRF2  00000951254WK2 956632RM1  000000000BJ880 410014XR06  N/A 206386RF1  N/A 212646PREC  N/A 281351TLEG  N/A 281890MWRF  N/A 318448COP2  N/A 404571DL1  N/A 408885LEG1  N/A 512341RRF1  N/A 605665FL  N/A 606723XR2  N/A 606836BX3A  N/A 610284RF1  N/A 612758RF  N/A 616781RF2  N/A 618288ARF2  N/A 702255RF2  N/A 717765RF1  N/A 718667PRM3  N/A 813725RF1  N/A 813432RF1  N/A 815786D5K  N/A 906226DMR2  N/A 912273CR1  N/A 931879JRF2  N/A 956721RF1  Not Available 0002872736  Not Available 202745PRE  Not Available 203688P500  Not Available 205286BRF  Not Available 214345PRF  Not Available 254562RF1  Not Available 262OLCGOC1A  Not Available 270417RF1  Not Available 407649LG2  Not Available 414777MP2  Not Available 504652RP2  Not Available 573760PR1  Not Available 580251ROOM1  Not Available 601268W5  Not Available 609835P500  Not Available 610284PRO2  Not Available 610284RF3  Not Available 617726RF4E2  Not Available 706754HR2  Not Available 713770THUNDER  Not Available 717267RF6  Not Available 815740XRDER  Not Available 909GERF1  Not Available 912466P1207  Not Available 928537RM1  Not Available 956632RCR  Not Available 973972C308  Not Available DCOSDI000000011  Not Available DCOSDI000000018  Not Available DCOSDI000000222  Not Available DCOSDI000000703  Not Available DCOSDI00001717  Not Available DCOSDI00001740  Not Available DCOSDI00001752  Not Available DCOSDI100000008  Not Available DCOSDI100000043  Not Available DCOSDI100000152  Not Available DCOSDI100000272  Not Available DCOSDI100000412  Not Available DCOSDI100000700  Not Available DCOSDI100000820  Not Available DCOSDI110000289  Not Available DCOSDI140000071  Not Available DCOSDI200000107  Not Available DCOSDI200001358  Not Available DCOSDI300000408  Not Available DCOSDI300000436  Not Available DCOSDI300000440  Not Available DCOSDI300000920  Not Available DCOSDI400000063  Not Available DCOSDI400000167  Not Available DCOSDI50000419  Not Available DCOSDI50000425  Not Available DCOSDI50000426  Not Available DCOSDI600000212  Not Available KPICT  Not Available LBDEMO2  Not Available RADNETRF1206  Not Available RADNETRF1213  Not Available RADNETRF1225  Not Available RADNETRF1240  Not Available RADNETRF1254  194130WK1 RADNETRF186  Not Available 416284LS16  Not Available 506648L9A  Not Available 514934BRAD3",
      "more_code_info": ""
    }
  ]
}