      "status": "Terminated",
      "city": "Tarrytown",
      "state": "NY",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "65898",
      "recalling_firm": "Siemens Healthcare Diagnostics",
      "address_1": "511 Benedict Ave",
      "address_2": "",
      "postal_code": "10591-5005",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - USA (nationwide) and the countries of Australia, Austria, Belgium, Canada, Croatia, Denmark, France, Germany, Greece, Italy, Japan, Malaysia, Netherlands, Norway, Philippines, Portugal, South Korea, Singapore, Slovenia, Spain, Switzerland, Thailand, Turkey, United Kingdom and Vietnam.",
      "recall_number": "Z-0011-2014",
      "product_description": "ADVIA 2120/2120i Hematology Systems with ADVIA Autoslide connected to it, SMN 10361162 and SMN 10361798 --- CLASSIFICATION NAME: Counter, Differential Cell, Automated (Particle Counter)    The ADVIA 2120/2120i are hematology systems that utilizes the principles of flow cytometry in order to provide complete blood counts.",
      "product_quantity": "311 units total (US: 51 units; Foreign: 260 units)",
      "reason_for_recall": "Siemens determined that if an autosampler rack jam error occurs during operation of an ADVIA 2120/2120i connected to an ADVIA Autoslide, it is possible that the next slide processed by the Autoslide could be labeled with the wrong sample identification information.",
      "recall_initiation_date": "20130812",
      "center_classification_date": "20131017",
      "termination_date": "20151118",
      "report_date": "20131023",
      "code_info": "All serial numbers connected to the ADVIA 2120/2120i Hematology Systems with ADVIA Autoslide.",
      "more_code_info": ""
    }
  ]
}