      "country": "United States",
      "city": "Brea",
      "address_1": "250 S Kraemer Blvd",
      "reason_for_recall": "The ABL90 FLEX can aspirate Cal 2 solution instead of rinse solution, when the following take place:  ¿ The analyzer is in ready mode  ¿ Operator lifts the inlet and then closes it again, thereby initiating a rinse  ¿ Immediately after lifts the inlet and quickly closes it again",
      "address_2": "",
      "product_quantity": "12,102",
      "code_info": "ABL90 FLEX analyzers with software versions 3.3MR1 or below.",
      "center_classification_date": "20181001",
      "distribution_pattern": "US Distribution and Internationally to Canada.",
      "state": "CA",
      "product_description": "ABL90 FLEX analyzers with software versions 3.3MR1 or below. The ABL90 FLEX Analyzer is a portable, automated Analyzer that measures pH, blood gases, electrolytes, glucose, lactate, bilirubin and oximetry in whole blood.    It is intended for use in a laboratory environment, near patient or point-of-care setting.",
      "report_date": "20181010",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Radiometer America Inc",
      "recall_number": "Z-0010-2019",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "80718",
      "termination_date": "20201106",
      "more_code_info": "",
      "recall_initiation_date": "20180525",
      "postal_code": "92821-6232",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}