      "status": "Terminated",
      "city": "Waukesha",
      "state": "WI",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "66250",
      "recalling_firm": "GE Healthcare, LLC",
      "address_1": "3000 N Grandview Blvd",
      "address_2": "",
      "postal_code": "53188-1615",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution-USA (nationwide) including the states of US: AK, AZ, CA, CO, DE, FL, GA, IL, KY, LA, MD, MI, MN, MS, MT, NV, NY, NC, OH, OR, PA, SC, TX, WY, and WI and the countries of AUSTRALIA, AUSTRIA, BANGLADESH, BELGIUM, BOLIVIA, BRAZIL, CANADA, CHILE, CHINA, COST ARICA, CROATIA, CZECH REPUBLIC, EGYPT, FINLAND, FRANCE, GERMANY, GREECE, INDIA, INDONESIA, IRAQ, ITALY, JAPAN, KAZAHHSTAN, REPUBLIC OF KOREA, KUWAIT, LEBANON, LITHUANIA, MAURITITUS, MEXICO, MOROCCO, NETHERLANDS, NEW ZEALAND, NIGERIA, PAKISTAN, PERU, PHILIPPINES, POLAND, RUSSIAN FEDERATION, SAUDI ARABIA, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, THAILAND, TURKEY, UNITED ARAB EMIRATES, UNITED KINGDOM, VIETNAM, and VENEZUELA.",
      "recall_number": "Z-0010-2014",
      "product_description": "GE Healthcare, Avance, Avance CS2, Aisys, Aespire View, and Engstrom.    Intended for volume or pressure control ventilation.",
      "product_quantity": "911 (252 US, 659 OUS)",
      "reason_for_recall": "GE Healthcare has recently become aware of a potential safety issue involving unresponsive buttons on the display of the Engstr¿m ventilator and the Aespire View, Aisys, and Avance anesthesia machines.    As a result of a manufacturing issue, the buttons on the left, right, and bottom keypads may not always detect button presses. This may result in the inability to access certain menu functions wh",
      "recall_initiation_date": "20130828",
      "center_classification_date": "20131015",
      "termination_date": "20150130",
      "report_date": "20131023",
      "code_info": "Mfg Lot or Serial #  APHS00701  APHS00802  APHS00729  APHS00730  APHS00700  APHS00692  APHS00932  APHS00886  APHS00861  APHS00865  APHS00866  APHS00867  APHS00868  APHS00887  APHS00888  APHS00860  APHS00881  70138420 - 1  APHS00844  APHS00711  APHS00712  5324664 - 1  APHS00703  APHS00704  5328496 - 1  APHS00690  APHS00719  APHS00720  APHS00722  APHS00723  APHS00828  APHS00834  APHS00835  APHS00836  APHS00843  APHS00809  APHS00854  APHS00855  APHS00856  APHS00857  APHS00858  APHS00691  APHS00702  APHS00705  APHS00721  APHS00807  APHS00808  APHS00811  APHS00685  APHS00695  APHS00696  APHS00697  APHS00698  APHS00699  APHS00805  APHS00806  APHS00812  APHS00810  APHS00864  APHS00643  APHS00644  APHS00645  APHS00646  APHS00647  APHS00648  APHS00649  APHS00650  APHS00651  APHS00652  APHS00653  APHS00654  APHS00655  APHS00656  APHS00657  APHS00666  APHS00667  APHS00731  APHS00732  APHS00733  APHS00734  APHS00735  APHS00736  APHS00737  APHS00738  APHS00739  APHS00740  APHS00741  APHS00742  APHS00743  APHS00744  APHS00745  APHS00748  APHS00749  APHS00750  APHS00751  APHS00752  APHS00753  APHS00754  APHS00755  APHS00756  APHS00757  APHS00758  APHS00759  APHS00760  APHS00761  APHS00762  APHS00763  APHS00764  APHS00765  APHS00766  APHS00767  APHS00768  APHS00769  APHS00770  APHS00771  APHS00772  APHS00773  APHS00774  APHS00775  APHS00776  APHS00777  APHS00778  APHS00779  APHS00780  APHS00781  APHS00782  APHS00783  APHS00784  APHS00785  APHS00786  APHS00787  APHS00788  APHS00789  APHS00790  APHS00933  APHS00686  APHS00687  APHS00688  APHS00689  APHS00706  APHS00707  APHS00710  APHS00726  APHS00826  APHS00827  APHS00838  APHS00846  APHS00847  APHS00869  APHS00871  APHS00874  APHS00875  APHS00876  APHS00882  APHS00883  APHS00884  APHS00885  APHS00872  APHS00873  APHS00693  APHS00725  APHS00801  APHS00879  APHS00880  APHS00727  APHS00728  APHS00682  APHS00877  APHS00878  APHS00849  APHS00850  APHS00852  APHS00853  APHS00839  APHS00840  APHS00841  APHS00842  APHS00677  APHS00859  APHS00253  APHS00694  APHS00804  APHS00845  APHS00862  APHS00863  ANAS01058  ANAS01059  ANAS00814  ANAS00265  ANAS01006  ANAS01007  ANAS01009  ANAS01010  ANAS01011  ANAS01012  ANAS01013  ANAS01014  ANAS01015  ANAS01026  ANAS01027  ANAS01028  ANAS01029  ANAS01030  ANAS01031  ANAS01032  ANAS01033  ANAS01034  ANAS01035  ANAS01036  ANAS01037  ANAS01038  ANAS01039  ANAS01040  ANAS01041  ANAS01042  ANAS01043  ANAS01045  ANAS01046  ANAS01047  ANAS01048  ANAS01049  ANAS01050  ANAS01051  ANAS01052  ANAS00536  ANAS01001  ANAS01002  ANAS01005  ANAS00475  ANAS00821  ANAS01067  ANAS01068  ANAS01069  ANAS01070  ANAS01071  ANAS00999  ANAS01000  ANAS00788  ANAS00789  ANAS00790  ANAS00793  ANAS00794  ANAS00795  ANAS00796  ANAS00797  ANAS00798  ANAS00799  ANAS00913  ANAS00914  ANAS00915  5323093 - 1  5324910 - 1  5324910 - 2  5324910 - 3  5324910 - 4  5324910 - 5  5324911 - 1  5325677 - 1  5325677 - 10  5325677 - 11  5325677 - 12  5325677 - 13  5325677 - 14  5325677 - 15  5325677 - 16  5325677 - 17  5325677 - 2  5325677 - 3  5325677 - 4  5325677 - 5  5325677 - 6  5325677 - 7  5325677 - 8  5325677 - 9  5326146 - 1  5326146 - 2  5326146 - 3  5326146 - 4  5326146 - 5  5326146 - 6  5326146 - 7  5326146 - 8  5327169 - 1  5327169 - 2  5327169 - 3  5327169 - 4  5327169 - 5  5327169 - 6  5327288 - 1  5327288 - 2  5327288 - 3  5329262 - 1  5329262 - 10  5329262 - 11  5329262 - 12  5329262 - 13  5329262 - 14  5329262 - 15  5329262 - 16  5329262 - 17  5329262 - 18  5329262 - 19  5329262 - 2  5329262 - 3  5329262 - 4  5329262 - 5  5329262 - 6  5329262 - 7  5329262 - 8  5329262 - 9  ANAS01053  ANAS01054  5325959 - 1  5325959 - 10  5325959 - 11  5325959 - 12  5325959 - 13  5325959 - 2  5325959 - 3  5325959 - 4  5325959 - 5  5325959 - 6  5325959 - 7  5325959 - 8  5325959 - 9  5326782 - 1  5326782 - 2  ANAS00922  ANAS00932  ANAS00923  ANAS00924  ANAS00900  ANAS00905  ANAS00935  ANAS00936  ANAS00937  ANAS00938  ANAS00939  ANAS00940  ANAS00941  ANAS00942  ANAS00943  ANAS00944  ANAS00945  5323856 - 1  5323856 - 2  5323856 - 3  5323856 - 4  5323856 - 5  5323856 - 6  5327897 - 1  5327897 - 2  ANAS00967  ANAS00968  ANAS00969  ANAS00970  ANAS00972  ANAS00973  ANAS00974  ANAS00975  ANAS00976  ANAS00977  ANAS00978  ANAS00979  ANAS00981  ANAS00982  ANAS00983  ANAS00984  ANAS00985  ANAS00997  ANAS00998  ANAS00906  ANAS00993  ANAS00909  5324418 - 1  5324418 - 2  5324418 - 3  5324492 - 1  5325318 - 1  5325318 - 2  ANAS00964  ANAS00965  5324012 - 1  5324012 - 2  5324012 - 3  5324672 - 1  ANAS00925  ANAS00926  ANAS00950  ANAS00951  ANAS00952  ANAS00953  ANAS00966  ANAS00971  ANAS00991  5326891 - 1  ANAS00800  ANAS00989  ANAS00990  5324604 - 1  5327482 - 1  5327482 - 2  5328933 - 1  ANAS00907  ANAS00934  ANAS01055  ANAS00276  ANAS00787  ANAS00933  ANAS00333  ANAS00386  ANAS01004  ANAS00994  ANAS00995  ANAS00996  ANAS00946  ANAS00947  ANAS00948  ANAS00949  ANAS00954  ANAS00955  ANAS00956  ANAS00957  ANAS00992  ANBS00603  ANBS00604  ANBS00800  ANBS00602  ANBS00370  70130790 - 1  70130790 - 2  70131952 - 1  70131952 - 2  70131952 - 3  70131952 - 4  70131952 - 5  70131952 - 6  70131952 - 7  70131952 - 8  70132184 - 1  70132184 - 2  70132184 - 3  70132184 - 4  70132184 - 5  70132537 - 1  70132537 - 1  70132537 - 2  70132537 - 2  70132537 - 3  70132619 - 1  70132619 - 2  70132619 - 3  70132619 - 4  70132619 - 5  70132619 - 6  70133381 - 1  70133381 - 2  70133381 - 3  70133381 - 4  70133381 - 5  70133381 - 6  70133381 - 7  70134330 - 1  70134330 - 2  70134330 - 3  ANBS00606  ANBS00687  ANBS00326  5321204 - 1  5321204 - 2  5321204 - 3  5322929 - 1  5322929 - 10  5322929 - 11  5322929 - 12  5322929 - 13  5322929 - 14  5322929 - 15  5322929 - 16  5322929 - 17  5322929 - 18  5322929 - 19  5322929 - 2  5322929 - 20  5322929 - 21  5322929 - 3  5322929 - 4  5322929 - 5  5322929 - 6  5322929 - 7  5322929 - 8  5322929 - 9  5324558 - 1  5326594 - 1  5326594 - 2  5328365 - 1  5328365 - 2  5328365 - 3  5328480 - 1  5328480 - 2  5329188 - 1  5329245 - 1  5329245 - 2  5329437 - 1  5329437 - 10  5329437 - 11  5329437 - 12  5329437 - 13  5329437 - 2  5329437 - 3  5329437 - 4  5329437 - 5  5329437 - 6  5329437 - 7  5329437 - 8  5329437 - 9  ANBS00748  ANBS00697  ANBS00698  ANBS00699  ANBS00780  ANBS00781  ANBS00782  ANBS00783  5325323 - 1  5325323 - 1  5325323 - 2  5325323 - 2  5328054 - 1  5328054 - 2  5328054 - 3  5328054 - 4  5328054 - 5  5328054 - 6  ANBS00754  ANBS00693  ANBS00694  ANBS00695  ANBS00696  ANBS00701  ANBS00702  ANBS00703  ANBS00704  ANBS00713  ANBS00714  ANBS00715  ANBS00716  ANBS00717  ANBS00718  ANBS00719  ANBS00720  ANBS00721  ANBS00761  ANBS00762  ANBS00798  5329715 - 1  5329651 - 1  ANBS00681  ANBS00686  ANBS00710  5329692 - 1  5329692 - 2  5329692 - 3  ANBS00755  ANBS00757  ANBS00758  ANBS00759  ANBS00733  ANBS00734  ANBS00735  ANBS00617  ANBS00770  ANBS00771  ANBS00779  ANBS00769  ANBS00737  ANBS00738  ANBS00739  ANBS00740  ANBS00741  ANBS00742  ANBS00743  ANBS00744  ANBS00745  ANBS00746  ANBS00672  ANBS00705  ANBS00706  ANBS00707  ANBS00708  ANBS00709  ANBS00722  ANBS00768  ANBS00801  ANBS00684  ANBS00685  ANBS00711  ANBS00712  ANBS00723  ANBS00724  ANBS00725  ANBS00726  ANBS00727  ANBS00728  ANBS00729  ANBS00730  ANBS00731  ANBS00732  ANBS00772  ANBS00773  ANBS00774  ANBS00775  ANBS00776  ANBS00777  ANBS00778  ANBS00784  ANBS00785  ANBS00786  ANBS00787  ANBS00788  ANBS00789  ANBS00790  ANBS00736  CBCS01139  CBCS01142  CBCS01143  CBCS01146  CBCS01147  CBCS01150  CBCS00975  CBCS00980  CBCS01215  CBCS01232  CBCS01174  5297602 - 1  5297658 - 1  5297658 - 2  5324971 - 1  5324971 - 10  5324971 - 2  5324971 - 3  5324971 - 4  5324971 - 5  5324971 - 6  5324971 - 7  5324971 - 8  5324971 - 9  CBCS00965  CBCS00968  CBCS00969  CBCS00970  CBCS00971  CBCS00972  CBCS00973  CBCS00974  CBCS00976  CBCS00977  CBCS00979  CBCS00981  CBCS00982  CBCS00983  CBCS01177  5326925 - 1  CBCS00963  CBCS01100  5324655 - 1  5324727 - 1  5325591 - 1  5325591 - 2  5325591 - 3  5326544 - 1  5326544 - 2  5326544 - 3  5326544 - 4  5326547 - 1  5326547 - 2  5326652 - 1  5326742 - 1  5326852 - 1  5326852 - 10  5326852 - 11  5326852 - 12  5326852 - 13  5326852 - 14  5326852 - 15  5326852 - 16  5326852 - 17  5326852 - 18  5326852 - 19  5326852 - 2  5326852 - 20  5326852 - 21  5326852 - 22  5326852 - 23  5326852 - 24  5326852 - 25  5326852 - 26  5326852 - 3  5326852 - 4  5326852 - 5  5326852 - 6  5326852 - 7  5326852 - 8  5326852 - 9  5326889 - 1  5326993 - 1  5326993 - 2  5327620 - 1  5327620 - 2  5327620 - 3  5327620 - 4  5327853 - 1  5327853 - 2  5327853 - 3  5328021 - 1  5328021 - 2  5328409 - 1  5328409 - 2  5328627 - 1  5328707 - 1  5328707 - 2  5328707 - 3  CBCS01026  CBCS01027  CBCS01028  CBCS01029  CBCS01030  CBCS01031  CBCS01032  CBCS01033  CBCS01034  CBCS01035  CBCS01036  CBCS01037  CBCS01038  CBCS01039  CBCS01040  CBCS01041  CBCS01042  CBCS01043  CBCS01044  CBCS01045  CBCS01046  CBCS01047  CBCS01048  CBCS01050  CBCS01052  CBCS01053  CBCS01054  CBCS01056  CBCS01057  CBCS01058  CBCS01062  CBCS01063  CBCS01064  CBCS01065  CBCS01066  CBCS01067  CBCS01068  CBCS01070  CBCS01071  CBCS01123  CBCS01124  CBCS01133  CBCS01136  CBCS01138  5323487 - 1  5324756 - 1  5324756 - 2  5325311 - 1  5326690 - 1  5326690 - 2  5326692 - 1  5326693 - 1  5326693 - 2  5326698 - 1  5326698 - 2  5326700 - 1  5320342 - 1  5320342 - 2  5320342 - 3  5320342 - 4  5320342 - 5  5325257 - 1  5325257 - 2  5325257 - 3  5325257 - 4  5325257 - 5  5325257 - 6  CBCS00964  CBCS00967  CBCS01181  CBCS01196  CBCS01197  CBCS01198  CBCS01199  CBCS01200  CBCS01201  CBCS01202  CBCS01203  CBCS01132  CBCS00957  CBCS00958  CBCS01089  CBCS01090  CBCS01148  CBCS01149  CBCS01151  CBCS01152  CBCS01153  CBCS01154  CBCS01155  CBCS01156  CBCS01157  CBCS01158  CBCS01160  CBCS01161  CBCS01162  CBCS01163  CBCS01164  CBCS01168  CBCS01135  CBCS01165  CBCS01166  CBCS01167  CBCS01169  CBCS01091  CBCS01092  CBCS01093  CBCS01094  CBCS01095  CBCS01096  CBCS01097  CBCS01098  CBCS01099  5329011 - 1  CBCS00953  CBCS00955  CBCS00959  CBCS00960  CBCS00961  CBCS00962  CBCS00966  CBCS00984  CBCS00985  CBCS01055  CBCS01059  CBCS01073  CBCS01074  CBCS01075  CBCS01076  CBCS01077  CBCS01078  CBCS01103  CBCS01109  CBCS01110  CBCS01111  CBCS01112  CBCS01113  CBCS01114  CBCS01115  CBCS01116  CBCS01117  CBCS01118  CBCS01119  CBCS01120  CBCS01125  CBCS01126  CBCS01127  CBCS01129  CBCS01130  CBCS01233  CBCS01236  CBCS01252  CBCS01253  CBCS01254  CBCS01069  CBCS01101  CBCS01102  CBCS01121  CBCS01122  5327572 - 1  CBCS01061  5325609 - 1  5325609 - 2  5325609 - 3  CBCS01049  CBCS01051  CBCS01060  CBCS01084  CBCS01104  CBCS01105  CBCS01106  CBCS01107  CBCS01108  5326361 - 1  5326361 - 2",
      "more_code_info": ""
    }
  ]
}