  "results": [
    {
      "status": "Terminated",
      "city": "Morrisville",
      "state": "NC",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "83119",
      "recalling_firm": "TELEFLEX-MORRISVILLE",
      "address_1": "3015 Carrington Mill Blvd",
      "address_2": "N/A",
      "postal_code": "27560-5437",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US nationwide distribution.  AL, CA, CO, CT, FL, GA, HI, IL, IN, KS, KY, LA, MA, MD, ME, MI, MO, MS, NC, NE, NJ, NM, NY, OH, OK, OR, PA, SC, TX, VA, WA and WY",
      "recall_number": "Z-0009-2020",
      "product_description": "Babi.Plus 12.5 cm H2O Pressure Relief Manifold Catalog # 2691  for use with ventilators.",
      "product_quantity": "9,312 each",
      "reason_for_recall": "Device vented gas below the stated pressure.",
      "recall_initiation_date": "20190614",
      "center_classification_date": "20191106",
      "termination_date": "20240612",
      "report_date": "20191113",
      "code_info": "Catalog # 2691 SKU #: XTF0046  Lot #'s: 180806, 180910, 181029, 181105, 181204, 190225 and 190327"
    }
  ]
}