{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-29",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Carlsbad",
      "address_1": "2271 Cosmos Ct",
      "reason_for_recall": "A failure of a specific component may prevent the ventilators from operating on AC power or transitioning back to AC power when operating on battery.  Additionally, if a battery is not present or is depleted the ventilators will not operate.  This failure will cause the loss of ventilator support, which may result in hypercarbia or hypoxemia.",
      "address_2": "",
      "product_quantity": "2,206 units total (665 units in US)",
      "code_info": "Serial Numbers:    VS0002577  VS0002579  VS0002580  VS0002581  VS0002583  VS0002584  VS0002585  VS0002586  VS0002587  VS0002588  VS0002589  VS0002590  VS0002593  VS0002595  VS0002596  VS0002599  VS0002602  VS0002603  VS0002604  VS0002605  VS0002606  VS0002609  VS0002610  VS0002611  VS0002613  VS0002615  VS0002616  VS0002625  VS0002630  VS0002746  VS0002750  VS0002752  VS0002753  VS0002759  VS0002728  VS0002773  VS0002765  VS0002772  VS0002828  VS0002842  VS0002830  VS0002834  VS0002859  VS0002873  VS0002874  VS0002877  VS0002879  VS0002884  VS0002844  VS0002846  VS0002847  VS0002851  VS0002856  VS0002891  VS0002815  VS0002816  VS0002924  VS0002927  VS0002929  VS0002932  VS0002933  VS0002934  VS0002936  VS0002938  VS0002939  VS0002943  VS0002942  VS0002947  VS0002946  VS0002960  VS0002994  VS0002998  VS0002708  VS0002817  VS0002940  VS0002958  VS0002959  VS0002963  VS0003023  VS0003029  VS0003074  VS0003079  VS0003080  VS0003092  VS0003093  VS0003141  VS0003095  VS0003114  VS0003176  VS0003260  VS0003139  VS0003172  VS0003259  VS0003286  VS0003303  VS0003343  VS0003357  VS0003361  VS0003430  VS0003440  VS0003444  VS0003446  VS0003448  VS0003451  VS0003494  VS0003500  VS0003503  VS0003504  VS0003505  VS0003506  VS0003507  VS0003509  VS0003511  VS0003513  VS0003514  VS0003518  VS0003545  VS0003546  VS0003554  VS0003549  VS0003557  VS0003607  VS0003601  VS0003704  VS0003684  VS0003690  VS0003694  VS0003698  VS0003701  VS0003702  VS0003705  VS0003707  VS0003709  VS0003713  VS0003657  VS0003733  VS0003600  VS0003757  VS0003743  VS0003744  VS0003750  VS0003751  VS0003755  VS0003756  VS0003818  VS0003819  VS0003820  VS0003822  VS0003823  VS0003825  VS0003826  VS0003828  VS0003829  VS0003830  VS0003831  VS0003876  VS0003886  VS0003890  VS0003891  VS0003892  VS0003922  VS0003907  VS0003921  VS0003923  VS0003924  VS0003925  VS0003896  VS0003913  VS0003939  VS0003940  VS0003942  VS0003897  VS0003906  VS0003944  VS0003945  VS0003960  VS0003971  VS0003975  VS0003973  VS0004010  VS0004016  VS0004017  VS0003791  VS0003792  VS0003993  VS0004004  VS0004023  VS0004024  VS0004056  VS0004062  VS0004068  VS0004069  VS0004072  VS0004077  VS0004079  VS0004087  VS0004095  VS0004060  VS0004065  VS0004083  VS0004097  VS0004098  VS0004099  VS0004101  VS0004104  VS0004105  VS0004121  VS0002980  VS0002983  VS3000564  VS3002057  VS3002123  VS3004918  VS3005977  VS3006242  VS3006568  VS3006632  VS3007083  VS3007093  VS3007326  VS3007399  VS3008641  VS3008943  VS3008981  VS3009885  VS3010067  VS3010449  VS3010988  VS3012196  123600002  123600005  123600011  123600018  123600023  123600025  123600029  123600030  123600036  123600039  123600044  123600053  123600056  123600059  123600063  123600077  123600078  123600080  123600081  123600082  123600095  124600102  124600115  124600118  124600157  124600158  124600172  124600179  124600180  124600182  124600183  124600186  124600200  124600201  124600217  124600224  124600225  124600228  124600242  124600247  124600248  124600256  124600257  124600261  124600264  124600266  124600269  124600270  124600278  124600293  124600295  124600296  124600298  124600299  124600301  124600303  124600304  124600306  124600309  124600310  124600312  124600315  124600317  124600321  124600326  124600327  124600329  124600334  124600339  124600343  124600346  124600348  124600354  124600355  124600356  124600358  124600360  124600364  124600373  124600375  124600379  124600381  124600382  124600384  124600386  124600389  124600401  124600410  124600413  124600414  124600430  124600431  124600434  124600440  124600443  124600444  124600446  124600453  124600455  124600457  124600459  124600460  124600463  124600465  124600471  124600476  124600477  124600481  124600483  124600488  124600490  124600491  124600494  124600501  124600502  124600508  124600510  124600512  124600519  124600523  124600525  124600526  124600528  124600529  124600530  124600534  124600535  124600538  124600545  124600546  124600550  124600553  124600555  124600558  124600559  124600563  124600568  124600571  124600573  124600577  124600581  124600583  124600585  124600593  124600598  131200641  131200665  131200673  131200676  131200677  131200680  131200682  131200683  131200686  131200697  131200699  131200705  131200706  131200713  131200720  131200722  131200724  131200729  131200732  131200741  131200742  131200743  131200744  131200747  131200748  131200752  131200753  131200757  131200760  131200764  131200765  131200771  131200773  131200779  131200782  131200785  131200787  131200788  131200789  131200791  131200796  131200798  131200799  131200804  131200805  131200806  131200807  131200813  131200816  131200817  131200818  131200822  131200825  131200826  131200829  131200834  131200835  131200844  131200847  131200848  131200854  131200859  131200867  131200869  131200878  131200883  131200884  131200913  131200918  131200921  131200923  131200925  131200931  131901041  131901053  131901072  131901087  131901102  131901109  131901137  131901143  131901157  131901167  131901193  131901194  131901206  131901225  132201293  132201333  132201346  132201348  132201350  132201355  132201359  132201368  132201370  132201387  132201395  132201403  132201406  132201426  132201434  132201438  132201456  132201465  132201469  132201471  132201479  132201482  132201488  132201491  132201497  132201500  132201509  132201510  132201514  132201528  132201537  133201558  133201566  133201580  133201583  133201585  133201598  133201604  133201616  133201617  133201618  133201631  133201639  133201641  133201645  133201647  133201650  133201653  133201658  133201662  133201667  133201669  133201683  133201691  133201693  133201694  133201697  133701702  133701709  133701710  133701715  133701716  133701718  133701721  133701722  133701723  133701735  133701737  133701745  133701747  133701755  133701756  133701758  133701762  133701763  133701765  133701768  133701785  133701792  133701793  133701794  133701796  133701799  133701801  133701807  133701835  133701841  133701842  133701844  133701846  133701847  133701849  133701864  133701876  133701877  133701880  133701881  133701884  133701892  133701896  133701900  133701902  133701904  133701905  133701906  133701908  133701909  133701913  133701922  133701934  133701944  133701949  135001963  135001972  135001973  135001979  135001984  135001987  135001995  135001996  135002011  135002012  135002019  135002024  135002027  135002030  135002042  135002045  135002049  135002051  135002052  135002058  135002062  135002065  135002089  135002093  135002098  135002099  141602105  141602108  141602109  141602110  141602111  141602118  141602120  141602122  141602124  141602127  141602139  141602142  141602146  141602149  141602152  141602154  141602161  141602162  141602167  141602170  141602174  141602178  141602183  141602193  141602198  141602199  141602203  141602204  141602205  141602206  141602209  141602211  141602213  141602217  141602218  141602220  141602224  141602231  141602233  141602234  141602237  141602240  141602245  141602249  141602252  141602254  141602259  141602260  141602263  141602267  141602268  141602273  141602278  141602284  141602301  141602305  141602309  141602315  141602320  141602324  141602326  141602332  141602344  141602447  141602451  141602463    International Serial Numbers:    VS0002612  VS0002600  VS0002601  VS0002628  VS0002635  VS0002637  VS0002638  VS0002696  VS0002697  VS0002699  VS0002701  VS0002702  VS0002619  VS0002620  VS0002641  VS0002695  VS0002703  VS0002704  VS0002706  VS0002707  VS0002700  VS0002709  VS0002647  VS0002713  VS0002714  VS0002715  VS0002718  VS0002717  VS0002724  VS0002730  VS0002732  VS0002745  VS0002747  VS0002748  VS0002743  VS0002716  VS0002723  VS0002731  VS0002754  VS0002755  VS0002758  VS0002760  VS0002761  VS0002762  VS0002763  VS0002767  VS0002757  VS0002764  VS0002766  VS0002780  VS0002784  VS0002785  VS0002786  VS0002788  VS0002790  VS0002779  VS0002792  VS0002793  VS0002794  VS0002798  VS0002799  VS0002803  VS0002806  VS0002807  VS0002808  VS0002810  VS0002811  VS0002813  VS0002801  VS0002805  VS0002819  VS0002827  VS0002831  VS0002832  VS0002833  VS0002839  VS0002841  VS0002698  VS0002820  VS0002825  VS0002826  VS0002829  VS0002838  VS0002864  VS0002911  VS0002857  VS0002896  VS0002893  VS0002909  VS0002912  VS0002915  VS0002917  VS0002919  VS0002920  VS0002921  VS0002923  VS0002926  VS0002949  VS0002987  VS0002989  VS0002990  VS0002991  VS0002992  VS0002995  VS0002997  VS0002999  VS0003000  VS0003001  VS0003002  VS0003004  VS0002633  VS0002710  VS0002878  VS0002996  VS0003005  VS0003041  VS0003043  VS0002622  VS0002636  VS0002644  VS0002865  VS0003026  VS0003031  VS0003036  VS0003037  VS0003038  VS0003039  VS0003042  VS0003047  VS0003003  VS0003049  VS0002618  VS0002646  VS0003011  VS0003030  VS0003035  VS0003044  VS0003048  VS0003059  VS0003061  VS0003069  VS0002592  VS0003028  VS0003032  VS0003033  VS0003034  VS0003057  VS0003058  VS0003062  VS0003063  VS0003065  VS0003066  VS0003068  VS0003070  VS0003076  VS0003050  VS0003073  VS0003083  VS0003084  VS0003086  VS0003021  VS0003046  VS0003087  VS0003088  VS0003025  VS0003064  VS0003075  VS0003091  VS0003097  VS0003102  VS0003105  VS0003018  VS0003098  VS0003106  VS0003113  VS0003118  VS0003119  VS0003122  VS0003234  VS0003186  VS0003193  VS0003194  VS0003208  VS0003104  VS0003117  VS0003161  VS0003163  VS0003173  VS0003174  VS0003182  VS0003188  VS0003190  VS0003254  VS0003082  VS0003227  VS0003235  VS0003301  VS0003373  VS0003374  VS0003375  VS0003378  VS0003379  VS0003382  VS0003383  VS0003384  VS0003385  VS0003386  VS0003387  VS0003388  VS0003389  VS0003390  VS0003392  VS0003393  VS0003394  VS0003396  VS0003398  VS0003400  VS0003401  VS0003402  VS0003403  VS0003406  VS0003391  VS0003404  VS0003405  VS0003407  VS0003409  VS0003410  VS0003411  VS0003397  VS0003399  VS0003408  VS0003412  VS0003413  VS0003414  VS0003415  VS0003416  VS0003417  VS0003419  VS0003476  VS0003533  VS0003536  VS0003540  VS0003542  VS0003543  VS0003555  VS0003450  VS0003453  VS0003454  VS0003456  VS0003457  VS0003458  VS0003459  VS0003460  VS0003461  VS0003462  VS0003463  VS0003464  VS0003465  VS0003466  VS0003468  VS0003469  VS0003470  VS0003471  VS0003472  VS0003475  VS0003480  VS0003531  VS0003541  VS0003560  VS0003563  VS0003571  VS0003573  VS0003574  VS0003576  VS0003577  VS0003578  VS0003579  VS0003581  VS0003582  VS0003583  VS0003584  VS0003585  VS0003586  VS0003587  VS0003588  VS0003589  VS0003590  VS0003591  VS0003592  VS0003593  VS0003594  VS0003559  VS0003562  VS0003567  VS0003568  VS0003569  VS0003570  VS0003572  VS0003575  VS0003595  VS0003596  VS0003598  VS0003599  VS0003602  VS0003604  VS0003605  VS0003608  VS0003618  VS0003613  VS0003615  VS0003620  VS0003621  VS0003622  VS0003623  VS0003625  VS0003626  VS0003627  VS0003628  VS0003632  VS0003634  VS0003635  VS0003603  VS0003617  VS0003630  VS0003631  VS0003638  VS0003639  VS0003597  VS0003606  VS0003629  VS0003644  VS0003646  VS0003649  VS0003660  VS0003611  VS0003612  VS0003614  VS0003642  VS0003643  VS0003645  VS0003648  VS0003650  VS0003651  VS0003652  VS0003653  VS0003655  VS0003656  VS0003654  VS0003663  VS0003664  VS0003666  VS0003670  VS0003674  VS0003665  VS0003671  VS0003679  VS0003609  VS0003662  VS0003669  VS0003673  VS0003675  VS0003676  VS0003677  VS0003678  VS0003682  VS0003683  VS0003619  VS0003659  VS0003745  VS0003753  VS0003758  VS0003759  VS0003760  VS0003773  VS0003776  VS0003777  VS0003778  VS0003779  VS0003780  VS0003781  VS0003782  VS0003783  VS0003784  VS0003785  VS0003788  VS0003802  VS0003807  VS0003808  VS0003809  VS0003811  VS0003799  VS0003810  VS0003812  VS0003813  VS0003814  VS0003815  VS0003816  VS0003817  VS0003821  VS0003834  VS0003835  VS0003827  VS0003832  VS0003833  VS0003836  VS0003837  VS0003838  VS0003841  VS0003866  VS0003878  VS0003879  VS0003881  VS0003882  VS0003883  VS0003888  VS0003920  VS0003926  VS0003927  VS0003928  VS0003929  VS0003930  VS0003934  VS0003862  VS0003869  VS0003874  VS0003900  VS0003902  VS0003910  VS0003914  VS0003937  VS0003963  VS0003965  VS0003969  VS0003977  VS0003980  VS0003989  VS0003990  VS0003997  VS0004000  VS0004020  VS0004018  VS0003985  VS0003999  VS0004003  VS0004005  VS0004001  VS0004002  VS0003986  VS0003996  VS0004011  VS0004014  VS0004015  VS0004019  VS0004021  VS0004022  VS0004025  VS0004026  VS0004027  VS0004028  VS0004029  VS0004030  VS0004031  VS0004032  VS0004033  VS0004034  VS0004035  VS0004036  VS0004038  VS0004039  VS0004040  VS0003961  VS0003983  VS0003991  VS0004037  VS0004041  VS0004042  VS0004043  VS0004044  VS0004045  VS0004046  VS0004047  VS0004048  VS0004049  VS0004050  VS0004051  VS0004052  VS0004053  VS0004054  VS0004055  VS0004100  VS0004102  VS0004103  VS0004106  VS0004107  VS0004108  VS0004109  VS0004110  VS0004111  VS0004112  VS0004113  VS0004114  VS0004117  VS0004118  VS0004119  VS0004120  VS0004122  VS0004123  VS0004124  VS0004125  VS0004126  VS0004127  VS0004115  VS0004116  VS0004128  VS0004129  VS0004130  VS0004131  VS0004132  VS0004133  VS0004134  VS0004135  VS0004136  VS0004137  VS0004138  VS0004140  VS0004141  VS0004142  VS0004143  VS0004144  VS0004145  VS0004146  VS0004147  VS0004148  VS0004150  VS0004152  VS0004153  VS0004154  VS0002705  VS0002711  VS0002712  VS0002719  VS0002721  VS0002726  VS0002727  VS0002729  VS0002733  VS0002734  VS0002735  VS0002736  VS0002737  VS0002738  VS0002742  VS0002849  VS0002858  VS0002863  VS0002867  VS0002875  VS0002876  VS0002882  VS0002908  VS0002910  VS0002922  VS0002937  VS0002944  VS0002945  VS0002948  VS0002950  VS0002951  VS0002952  VS0002953  VS0002954  VS0002955  VS0002956  VS0002957  VS0002961  VS0002962  VS0002965  VS0002966  VS0002967  VS0002968  VS0002969  VS0002970  VS0002971  VS0002972  VS0002973  VS0002974  VS0002975  VS0002976  VS0002977  VS0002978  VS0002981  VS0002982  VS0003017  VS0003089  VS0003116  VS0003121  VS0003134  VS0003432  VS0003435  VS0003439  VS0003441  VS0003442  VS0003443  VS0003445  VS0003972  VS0003981  VS0003988  VS0003998  VS0004008  VS0004009  VS0004012  VS0004013  VS0004057  VS0004058  VS0004059  VS0004061  VS0004064  VS0004066  VS0004067  VS0003107  VS0003112  VS0003123  VS0003126  VS0003128  VS0003137  VS0003138  VS0003140  VS0003145  VS0003150  VS0003153  VS0003154  VS0003155  VS0003164  VS0003167  VS0003168  VS0003169  VS0003171  VS0003191  VS0003200  VS0003213  VS0003215  VS0003216  VS0003219  VS0003221  VS0003224  VS0003225  VS0003226  VS0003231  VS0003275  VS0003094  VS0003125  VS0003131  VS0003142  VS0003146  VS0003147  VS0003151  VS0003158  VS0003160  VS0003165  VS0003166  VS0003170  VS0003175  VS0003179  VS0003181  VS0003183  VS0003189  VS0003198  VS0003205  VS0003207  VS0003214  VS0003217  VS0003218  VS0003222  VS0003223  VS0003229  VS0003230  VS0003232  VS0003236  VS0003237  VS0003238  VS0003240  VS0003241  VS0003243  VS0003245  VS0003247  VS0003248  VS0003251  VS0003252  VS0003261  VS0003262  VS0003263  VS0003265  VS0003267  VS0003268  VS0003270  VS0003271  VS0003272  VS0003274  VS0003276  VS0003277  VS0003278  VS0003279  VS0003280  VS0003281  VS0003282  VS0003284  VS0003285  VS0003287  VS0003288  VS0003103  VS0003111  VS0003148  VS0003157  VS0003162  VS0003177  VS0003178  VS0003184  VS0003187  VS0003192  VS0003196  VS0003199  VS0003201  VS0003204  VS0003212  VS0003228  VS0003242  VS0003246  VS0003250  VS0003253  VS0003255  VS0003266  VS0003289  VS0003294  VS0003295  VS0003297  VS0003299  VS0003302  VS0003305  VS0003306  VS0003180  VS0003197  VS0003209  VS0003211  VS0003220  VS0003233  VS0003244  VS0003256  VS0003257  VS0003258  VS0003264  VS0003273  VS0003283  VS0003291  VS0003292  VS0003293  VS0003296  VS0003298  VS0003300  VS0003304  VS0003307  VS0003308  VS0003309  VS0003310  VS0003311  VS0003312  VS0003313  VS0003314  VS0003315  VS0003316  VS0003317  VS0003318  VS0003319  VS0003320  VS0003321  VS0003322  VS0003323  VS0003324  VS0003325  VS0003326  VS0003327  VS0003328  VS0003329  VS0003330  VS0003331  VS0003332  VS0003333  VS0003334  VS0003335  VS0003336  VS0003337  VS0003338  VS0003339  VS0003340  VS0003342  VS0003344  VS0003345  VS0003347  VS0003350  VS0003351  VS0003352  VS0003353  VS0003354  VS0003355  VS0003356  VS0003358  VS0003467  VS0003481  VS0003484  VS0003486  VS0003487  VS0003488  VS0003489  VS0003479  VS0003491  VS0003492  VS0003493  VS0003495  VS0003496  VS0003497  VS0003498  VS0003501  VS0003502  VS0004139  VS0002578  VS0002582  VS0002598  VS0002607  VS0002614  VS0002617  VS0002621  VS0002624  VS0002626  VS0002627  VS0002629  VS0002632  VS0002634  VS0002639  VS0002640  VS0002642  VS0002643  VS0002645  VS0002648  VS0002660  VS0002744  VS0002722  VS0002749  VS0002751  VS0002770  VS0002756  VS0002768  VS0002769  VS0002774  VS0002777  VS0002623  VS0002725  VS0002775  VS0002778  VS0002781  VS0002782  VS0002783  VS0002787  VS0002795  VS0002797  VS0002800  VS0002802  VS0002812  VS0002818  VS0002804  VS0002771  VS0002776  VS0002789  VS0002836  VS0002862  VS0002791  VS0002835  VS0002845  VS0002853  VS0002855  VS0002860  VS0002861  VS0002869  VS0002897  VS0002840  VS0002872  VS0002883  VS0002889  VS0002899  VS0002903  VS0002906  VS0002907  VS0002881  VS0002900  VS0002890  VS0002913  VS0002914  VS0002916  VS0002918  VS0002925  VS0002931  VS0002935  VS0002941  VS0003007  VS0003008  VS0003009  VS0003013  VS0003014  VS0003015  VS0003016  VS0003019  VS0003020  VS0003022  VS0003024  VS0003052  VS0002988  VS0003027  VS0003040  VS0003051  VS0003053  VS0003054  VS0003067  VS0003099  VS0003100  VS0003136  VS0003143  VS0003156  VS0003185  VS0003124  VS0003006  VS0003010  VS0003012  VS0003144  VS0003195  VS0003210  VS0003290  VS0003349  VS0003152  VS0003239  VS0003269  VS0003341  VS0003348  VS0003359  VS0003360  VS0003362  VS0003363  VS0003364  VS0003365  VS0003367  VS0003368  VS0003369  VS0003371  VS0003372  VS0003437  VS0003447  VS0003455  VS0003477  VS0003482  VS0003483  VS0003485  VS0003490  VS0003525  VS0003526  VS0003528  VS0003530  VS0003499  VS0003523  VS0003529  VS0003534  VS0003537  VS0003539  VS0003532  VS0003552  VS0003556  VS0003558  VS0003561  VS0003564  VS0003565  VS0003566  VS0003580  VS0003610  VS0003640  VS0003667  VS0003685  VS0003624  VS0003633  VS0003637  VS0003680  VS0003681  VS0003688  VS0003691  VS0003692  VS0003695  VS0003708  VS0003689  VS0003658  VS0003712  VS0003720  VS0003726  VS0003734  VS0003729  VS0003687  VS0003732  VS0003735  VS0003736  VS0003737  VS0003738  VS0003739  VS0003740  VS0003741  VS0003742  VS0003761  VS0003747  VS0003752  VS0003754  VS0003764  VS0003766  VS0003767  VS0003768  VS0003731  VS0003748  VS0003749  VS0003804  VS0003805  VS0003806  VS0003870  VS0003857  VS0003889  VS0003884  VS0003885  VS0003887  VS0003912  VS0003931  VS0003932  VS0003933  VS0003936  VS0003893  VS0003915  VS0003853  VS0003894  VS0003895  VS0003898  VS0003899  VS0003901  VS0003903  VS0003905  VS0003908  VS0003911  VS0003943  VS0003947  VS0003951  VS0003938  VS0003995  VS0003992  VS0003473  VS0003478  VS0003510  VS0003512  VS0003515  VS0003516  VS0003517  VS0003519  VS0003520  VS0003521  VS0003524  VS0003527  VS0002597  VS0002608  VS0002870  VS0002888  VS0003159  VS0003202  VS0003203  VS0003206  VS0003249  VS0003918  VS0003919  VS0003941  VS0003946  VS0003948  VS0003949  VS0003950  VS0003952  VS0003953  VS0003955  VS0003956  VS0003957  VS0004073  VS0004078  VS0004080  VS0004082  VS0004084  VS0004085  VS0004086  VS0004088  VS0004089  VS0004090  VS0004091  VS0004092  VS0004093  VS0004094  VS0004096  VS0003127  VS0003130  VS0003746  VS0003762  VS0003763  VS0003765  VS0003769  VS0003770  VS0003771  VS0003772  VS0003774  VS0003775  VS0003346  VS0003366  VS0003370  VS0003376  VS0003377  VS0003872  VS0002885  VS0002901  VS0002902  VS0002904  VS0003423  VS0003426  VS0003428  VS0003433  VS0003434  VS0003436  VS0003438  VS0003787  VS0003793  VS0003794  VS0003795  VS0003796  VS0003797  VS0003798  VS0003800  VS0003801  VS0003803  VS0003522  VS0002928  VS0002930  VS0003535  VS0003538  VS0003544  VS0003867  VS0003868  VS0003873  VS0003875  VS0003877  VS0003909  VS0003954  VS0003958  VS0003959  VS0003964  VS0003967  VS0003149  VS0002843  VS0002886  VS0002905  VS0003474  VS0003547  VS0003871  VS0003916  VS0003917  VS0003839  VS0003840  VS0003842  VS0003843  VS0003844  VS0003845  VS0003846  VS0003847  VS0003848  VS0003849  VS0003850  VS0003851  VS0003852  VS0003854  VS0003855  VS0003856  VS0003858  VS0003859  VS0003860  VS0003861  VS0003863  VS0003864  VS0003865  VS0003904  VS0003935  VS0003994  VS0004007  VS0002854  VS0002868  VS0002871  VS0003085  VS0003090  VS0003101  VS0003115  VS0003129  VS0003132  VS0003133  VS0003135  VS0003641  VS0003636  VS0003647  VS0003699  VS0003706  VS0003710  VS0003711  VS0003714  VS0003715  VS0003716  VS0003717  VS0003718  VS0003719  VS0003721  VS0003722  VS0003723  VS0003724  VS0003725  VS0003786  VS0003789  VS0003790  VS0003968  VS0003974  VS0002796  VS0002809  VS0002814  VS0002822  VS0002964  VS0002986  VS0003380  VS0003381  VS0003697  VS0003449  VS0002821  VS0002824  VS0002837  VS0002852  VS0003727  VS0003730  VS0002880  VS0002887  VS0002892  VS0002894  VS0002895  VS0002898  VS0002720  VS0002739  VS0002740  VS0002741  VS0002979  VS0002984  VS0002985  VS0002993  VS0003056  VS0003060  VS0003071  VS0003077  VS0003078  VS0003081  VS0003072  VS0003096  VS0003108  VS0003109  VS0003110  VS0003120  VS0003395  VS0003418  VS0003420  VS0003421  VS0003422  VS0003424  VS0003425  VS0003427  VS0003429  VS0003431  VS0003962  VS0003966  VS0003970  VS0003976  VS0003978  VS0003979  VS0003982  VS0003984  VS0003987  VS0004063  VS0004070  VS0004071  VS0004074  VS0004075  VS0004076  VS0004081  VS0002848  VS0002850  VS0003548  VS0003550  VS0003551  VS0003553  VS0003616  VS0003661  VS0003668  VS0003672  VS0003686  VS0003693  VS0003696  VS0003700  VS0003703  VS0002631  VS0003055    133701729  135002060  131200654  131200658  131200671  131200715  131200737  131901116  132201431  133701733  133701751  133701819  133701833  133701915  141602235  141602236  123600015  123600038  123600040  123600042  123600045  123600085  124600123  124600280  124600286  124600357  124600441  124600447  124600452  124600475  124600584  124600587  131200649  131200666  131200668  131200678  131200696  131200702  131200730  131200790  131200811  132201258  132201303  132201360  132201392  132201423  132201433  132201480  133201568  133201574  133201577  133201590  133201607  133201635  133201644  133201672  133201677  133201689  133701742  133701784  133701788  133701797  133701798  133701821  133701837  133701838  133701918  133701943  135002006  123600035  123600049  123600098  124600159  124600185  124600196  124600213  124600218  124600220  124600221  124600259  124600331  124600342  124600365  124600395  124600404  124600408  124600416  124600418  124600468  124600489  124600499  124600504  131200652  131200663  131200675  131200679  131200689  131200694  131200695  131200745  131200749  131200781  131200784  131200800  131200808  131200810  131200831  131200837  131200866  131200933  131200935  131900945  131901054  131901156  131901186  131901212  131901213  132201275  132201299  132201322  132201400  132201440  132201477  132201506  132201516  133201584  133201630  133201637  133201656  133201657  133201670  133201687  133701712  133701719  133701741  133701749  133701752  133701760  133701772  133701783  133701804  133701843  133701861  133701872  133701875  133701920  133701932  135001959  135002014  135002023  135002029  135002039  135002072  135002075  135002076  135002087  135002088  141602117  141602143  141602148  141602185  141602191  141602225  141602226  141602227  141602256  141602257  141602286  141602288  141602401  141602413  141602415  131200721  141602270  123600086  131200823  132201306  133701916  123600093  141602349  141602369  141602395  141602397  141602399  124600205  124600552  131200692  131200809  131200812  131901183  131901235  133701736  133701836  132201441  131200750  133201559  135002000  135002002  141602112  141602115  141602116  141602166  141602180  141602184  141602201  141602242  141602244  124600448  124600449  131200661  131200704  131200717  131200735  131200767  132201300  132201451  132201467  131200670  131200734  131200852  133701818  131200725  131200758  131200778  135002034  141602126  141602265  141602238  141602269  141602280",
      "center_classification_date": "20141021",
      "distribution_pattern": "Worldwide Distribution - USA, Argentina, China, Egypt, Ethiopia, Honduras, Hong Kong, India, Indonesia, Iraq, Israel, Italy, Jordan,   Libyan Arab Yamahiriya, Malaysia, Mali, Mexico, Myanmar (Burma). Nepal, Nigeria. Pakistan, Peru, Philippines, Serbia, Singapore, South Africa, Sri Lanka,   United Republic of Tanzania, Thailand, Vietnam, Zimbabwe,   Republic of Korea, and Netherlands.",
      "state": "CA",
      "product_description": "Esprit V1000 and V200 Ventilators, Model No. V1000 and V200; and the 3rd Generation Power Supply Repair Part (RP)Kits    Microprocessor-controlled, electrically powered mechanical ventilators intended for use by qualified medical personnel to provide continuous or intermittent ventilator support for adult, pediatric and neonatal patients as prescribed by a physician.",
      "report_date": "20141029",
      "classification": "Class I",
      "openfda": {},
      "recalling_firm": "Respironics California Inc",
      "recall_number": "Z-0008-2015",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "69313",
      "termination_date": "20160928",
      "more_code_info": "",
      "recall_initiation_date": "20140917",
      "postal_code": "92011-1517",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}