  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Mentor",
      "state": "OH",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "63196",
      "recalling_firm": "US Endoscopy Group Inc",
      "address_1": "5976 Heisley Rd",
      "address_2": "N/A",
      "postal_code": "44060-1873",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution -- USA (nationwide) including the states of  AL, AR, AZ, CA, FL, GA, HI, IA, IL, KY, LA, MD, ME, MI, MO, MS, MT, NC, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, VT, WA, and WI; and to the following countries: Argentina, Australia, Austria, Bahrain, Belgium, Brazil, Caribbean, Chile, Columbia, Croatia, Czech Republic, Denmark, Ecuador, El Salvador, Estonia, Finland, France, Germany, Great Britain, Greece, India, Ireland, Israel, Italy, Jordan, Lithuania, Malta, Mexico, Netherlands, New Zealand, Norway, Panama, Poland, Portugal, Romania, Saudi Arabia, Slovenia, South Africa, Spain, Switzerland, Turkey, and United Arab Emirates.",
      "recall_number": "Z-0006-2013",
      "product_description": "Endoscopic retrieval device.    The disposable Raptor grasping device in intended to retrieve foreign bodies or resected tissue specimens in endoscopic procedures of the upper and lower gastrointestinal tract or biliary tract.",
      "product_quantity": "1153",
      "reason_for_recall": "When the outer catheter is either coiled or in a contorted configuration outside of the endoscope and the patient, and excessive force is applied to the handle, the jaws may become inoperable and may become stuck in the open condition.",
      "recall_initiation_date": "20111114",
      "center_classification_date": "20121003",
      "termination_date": "20130702",
      "report_date": "20121010",
      "code_info": "Model #00711177; S/N: 52979, 52981, 52982, 53643, 53644, 53891, 53892, 54238"
    }
  ]
}