      "country": "United States",
      "city": "Palo Alto",
      "address_1": "3100 Hansen Way",
      "reason_for_recall": "Reports have been received of an anomaly that can result in a treatment without intended gating (respiratory tracking/monitoring). This issue occurs when a patient planned with gating is treated on more than one system.",
      "address_2": "",
      "product_quantity": "4",
      "code_info": "VitalBeam Version 2.5MR1 Systems (software 02.05.13.05/02.05.13.07):  H192482 VITAL BEAM - 321274492  H192521 VITAL BEAM  H192743 VITAL BEAM  H192816 VITAL BEAM    Varian Product Code family H19, which includes TrueBeam, TrueBeam STx, VitalBeam and EDGE.",
      "center_classification_date": "20181001",
      "distribution_pattern": "Worldwide Distribution - US nationwide in the states of AL, AR, AZ, CA, CO, FL, Guan, IL IN, KS, KY, LA, MA, MD, MI, MO, NH, NJ, NY, OH, OK, PA, Puerto Rico, TN, TX, VA, WI and the countries of :  Algeria, Australia, Azerbaijan, Belgium, Brazil, Bulgaria, Canada, China, Croatia, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Georgia, Germany, India, Ireland, Israel, Italy, Japan, Kazakhstan, Lebanon, Morocco, Nepal, The Netherlands, New Zealand, Norway, Panama, Poland, Reunion Russia, Saudi Arabia, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey and United Kingdom",
      "state": "CA",
      "product_description": "VitalBeam Radiotherapy Delivery System Version 2.5    Product Usage:  The VitalBeam delivery systems are intended to provide precision radiotherapy and stereotactic radiosurgery for lesions, tumors, and conditions anywhere in the body where radiation treatment is indicated.",
      "report_date": "20181010",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Varian Medical Systems",
      "recall_number": "Z-0005-2019",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "81010",
      "termination_date": "20200828",
      "more_code_info": "",
      "recall_initiation_date": "20180820",
      "postal_code": "94304-1030",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}