  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "North Haven",
      "state": "CT",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "86214",
      "recalling_firm": "Covidien, LP",
      "address_1": "60 Middletown Ave",
      "address_2": "N/A",
      "postal_code": "06473-3908",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide and Global Distribution.",
      "recall_number": "Z-0003-2021",
      "product_description": "Endo GIA Auto Suture Universal Articulating Loading Unit 30 mm - 3.5 mm, Item Code 030452 - Product Usage: The Endo GIA  universal staplers have applications in abdominal, gynecologic, pediatric and thoracic surgery for resection, transection and creation of anastomosis.",
      "product_quantity": "900,007 total",
      "reason_for_recall": "The device staples may not properly form upon application preventing adequate hemostasis.  Use of a product with this assembly error may result in incomplete staple formation.",
      "recall_initiation_date": "20200812",
      "center_classification_date": "20201001",
      "report_date": "20201007",
      "code_info": "Lot Numbers:   T0C085X  T8B048X  T8K049X  T8L087X  T9J072X  T7H002X  T8B049X  T8K095X  T9G027X  T9J099X  T7L026X  T8E063X  T8L084X  T9H043X  T9L073X  T7L027X  T8E064X  T8L085X  T9H115X  T9M016X  T8B014X  T8E075X  T8L086X  T9H116X  T9M111X  T8B015X  T8F074X"
    }
  ]
}