  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Irwindale",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "72167",
      "recalling_firm": "Biosense Webster, Inc.",
      "address_1": "15715 Arrow Hwy",
      "address_2": "N/A",
      "postal_code": "91706-2006",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution.  US nationwide, Italy, Canada, Japan, China, Singapore, Colombia, England, Russia, Spain, France, Germany, Finland, Ireland, Australia, Sweden, Hungary, India, Brazil, Slovenia, Netherlands, Mexico, and South Africa.",
      "recall_number": "Z-0003-2016",
      "product_description": "SOUNDSTAR eco 8F and 10F Diagnostic Ultrasound Catheters.  For intra-cardiac and intra-luminal visualization of cardiac and great vessel anatomy and physiology as well as visualization of other devices in the heart.",
      "product_quantity": "1035 (U.S.)",
      "reason_for_recall": "Image disappeared from the cardiac ultrasound system when the CARTO 3 EP Navigation System needed restarting while the patient was experiencing pericardial effusion.  Affects the CARTOSOUND Module of the CARTO 3 EP Navigation System when used with the SOUNDSTAR eco 8F and 10F Diagnostic Ultrasound Catheters. New precautions added.",
      "recall_initiation_date": "20150910",
      "center_classification_date": "20151001",
      "termination_date": "20160513",
      "report_date": "20151007",
      "code_info": "Catalog #: 10439236, 10439072, 10439011, 10438577"
    }
  ]
}