  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Birmingham",
      "state": "N/A",
      "country": "United Kingdom",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "78047",
      "recalling_firm": "The Binding Site Group, Ltd.",
      "address_1": "8 Calthorpe Road",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "Worldwide Distribution - US Distribution to the state of CA., and to the countries of : United Kingdom, Germany, France, Czech Republic, Belgium, Panama, Bulgaria, Iran, Argentina, Turkey Romania, Colombia, and Saudi Arabia.",
      "recall_number": "Z-0002-2018",
      "product_description": "Human Alpha-1-Antitrypsin Kit for use on the SPAPlus analyser, Product #NK034.S.  The firm name on the label is The Binding Site Group Ltd., Birmingham, UK.    Product is designed for the quantitative in-vitro determination of Alpha-1-antitrypsin in human serum using the SPAPLUS turbidimetric analyser. The measurement of Alpha-1-antitrypsin aids in the diagnosis of several conditions including adult cirrhosis of the liver. In addition, Alpha-1-antitrypsin deficiency has been associated with pulmonary emphysema. This test should be used in conjunction with other laboratory and clinical findings.",
      "product_quantity": "361 kits",
      "reason_for_recall": "The high control for the kit is recovering low and may be out of range for assigned values.",
      "recall_initiation_date": "20170823",
      "center_classification_date": "20171004",
      "termination_date": "20190221",
      "report_date": "20171011",
      "code_info": "Lot #400900, Exp. 12/31/2017, UDI #05051700016421",
      "more_code_info": ""
    }
  ]
}