      "status": "Terminated",
      "city": "Irwindale",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "72167",
      "recalling_firm": "Biosense Webster, Inc.",
      "address_1": "15715 Arrow Hwy",
      "address_2": "N/A",
      "postal_code": "91706-2006",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution.  US nationwide, Italy, Canada, Japan, China, Singapore, Colombia, England, Russia, Spain, France, Germany, Finland, Ireland, Australia, Sweden, Hungary, India, Brazil, Slovenia, Netherlands, Mexico, and South Africa.",
      "recall_number": "Z-0002-2016",
      "product_description": "Carto 3 EP Navigation System. Electro physiology system which views of the electrical activity of the heart through real-time data on 3-D, color-coded cardiac maps.",
      "product_quantity": "1035 (U.S.)",
      "reason_for_recall": "Image disappeared from the cardiac ultrasound system when the CARTO 3 EP Navigation System needed restarting while the patient was experiencing pericardial effusion.  Affects the CARTOSOUND Module of the CARTO 3 EP Navigation System when used with the SOUNDSTAR eco 8F and 10F Diagnostic Ultrasound Catheters. New precautions added.",
      "recall_initiation_date": "20150910",
      "center_classification_date": "20151001",
      "termination_date": "20160513",
      "report_date": "20151007",
      "code_info": "Catalog #: C3SOUND, 10439236, 10439072, 10439011, 10438577"
    }
  ]
}