{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "North Haven",
      "state": "CT",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "86214",
      "recalling_firm": "Covidien, LP",
      "address_1": "60 Middletown Ave",
      "address_2": "N/A",
      "postal_code": "06473-3908",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide and Global Distribution.",
      "recall_number": "Z-0001-2021",
      "product_description": "Endo GIA Auto Suture Gray Universal Articulating Loading Unit 30 mm - 2.0 mm, Item Code 030450 - Product Usage: The Endo GIA  universal staplers have applications in abdominal, gynecologic, pediatric and thoracic surgery for resection, transection and creation of anastomosis.",
      "product_quantity": "900,007 total",
      "reason_for_recall": "The device staples may not properly form upon application preventing adequate hemostasis.  Use of a product with this assembly error may result in incomplete staple formation.",
      "recall_initiation_date": "20200812",
      "center_classification_date": "20201001",
      "report_date": "20201007",
      "code_info": "Lot Numbers:   T0A108X  T7H003X  T8D008X  T8K108X  T9G060X  T0E051X  T7L023X  T8H047X   T8M079X  T7G003X  T8D007X  T8K047X  T9F132X"
    }
  ]
}