  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Oberkochen",
      "state": "N/A",
      "country": "Germany",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "78082",
      "recalling_firm": "Carl Zeiss Meditec AG",
      "address_1": "Rudolf-Eber-Str. 11",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide Distribution including AZ, CA, CT, DC, FL, GA, OH, and PA.",
      "recall_number": "Z-0001-2018",
      "product_description": "Carl Zeiss INTRABEAM PRS 500, product #304534-0000-500.  The firm name on the label is Carl Zeiss Meditec AG, Jena, Germany.  The INTRABEAM System is indicated for radiation therapy treatments",
      "product_quantity": "9 devices",
      "reason_for_recall": "The device had an incorrect printed calibration value on the calibration certificate.",
      "recall_initiation_date": "20170830",
      "center_classification_date": "20171002",
      "termination_date": "20180124",
      "report_date": "20171011",
      "code_info": "Intrabeam ser. #6402501246, Ion chamber ser. #SN001254; Intrabeam ser. #6402501241, Ion chamber ser. #SN001745; Intrabeam ser. #6402100932, Ion chamber ser. #SN001994; Intrabeam ser. #6402100958, Ion chamber ser. #SN002096; Intrabeam ser. #6402101042, Ion chamber ser. #SN002265; Intrabeam ser. #6402101029, Ion chamber ser. #SN002270; Intrabeam ser. #6402101034, Ion chamber ser. #SN002302; Intrabeam ser. #6402101105, Ion chamber ser. #SN002651; and Intrabeam ser. #6402101130, Ion chamber ser. #SN002684.",
      "more_code_info": ""
    }
  ]
}