{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Orlando",
      "state": "FL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "69192",
      "recalling_firm": "Invivo Corporation",
      "address_1": "12151 Research Pkwy",
      "address_2": "Suite 200",
      "postal_code": "32826-3222",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "Worldwide Distribution: US distribution in state of: CA and countries of: Poland, Singapore, and Sweden.",
      "recall_number": "Z-0001-2015",
      "product_description": "Philips Expression MR200 MRI Patient Monitoring System Model 866120",
      "product_quantity": "10 units",
      "reason_for_recall": "During production, excess metal (aluminum) debris was found stuck to the cling plastic that covers the device's pole assembly. The metallic debris could affect system performance due to the potential for a failure to power up or loss of power. This could result in the loss of monitoring and a delay in diagnosis of the patient.",
      "recall_initiation_date": "20140908",
      "center_classification_date": "20141001",
      "termination_date": "20170109",
      "report_date": "20141008",
      "code_info": "Model 866120 with the following lots #'s:  US33600006, US33600007, US33600011, US33600012 and US33600013"
    }
  ]
}