  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-10-28",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Fremont",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "99870",
      "recalling_firm": "Think Surgical, Inc.",
      "address_1": "47201 Lakeview Blvd",
      "address_2": "",
      "postal_code": "94538-6530",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "US Nationwide distribution in the states of NY and DE.",
      "recall_number": "Z-0192-2027",
      "product_description": "Think Surgical Square Driver  REF: 109126",
      "product_quantity": "15 units",
      "reason_for_recall": "Robotic system square drive failed to meet design specifications potentially resulting in pin becoming stuck in square driver or the pin wobbling during drilling.",
      "recall_initiation_date": "20260820",
      "center_classification_date": "20261016",
      "report_date": "20261028",
      "code_info": "Lot: 25356/Primary DI: M9461091260",
      "more_code_info": ""
    }
  ]
}